Eur J Cardiothorac Surg 2000;17:20-24
© 2000 Elsevier Science NL
a Department of Cardio-thoracic Surgery, Herz-Kreislauf-Klinik Bevensen, Römstedter Str. 25, 29549 Bad Bevensen, Germany
b Department of Cardiovascular Research, University Hospital Skejby Sygehus, Brendstrupg
rdsvej, DK-8200 Aarhus, Denmark
Corresponding author. Tel.: +49-5821-821-702; fax: +49-5821-821-777
e-mail: prof.laas{at}t-online.de
Objective: High intensity transient signals (HITS) representing microembolization to the brain have been found to contribute to cognitive impairment and psychoneurological dysfunction in patients carrying a mechanical aortic valve. It is unknown, whether HITS represent gaseous or solid emboli. This animal study evaluates the impact of valve orientation on HITS for two different mechanical valves with both valves implanted in their best and worst orientation, which has been defined in previous studies with respect to downstream turbulence. Methods: In four pigs a rotation device carrying either a Medtronic Hall (MH) or St. Jude Medical (SJM) valve size 23 mm was implanted. The device allowed rotation of the implanted valves without reopening of the aorta. Approximately 30 min after weaning from extracorporeal circulation, a Doppler probe was placed on both common carotid arteries. In different orientations of the implanted valves (best and worst position), HITS were detected by the Doppler probe and recorded for ten min by a transcranial Doppler sonography device (Medilab Inc., Estenfeld, Germany). Results: HITS showed significant change with rotation for both valve designs. With the major orifice of the MH oriented towards the non-coronary leaflet (optimum position) very low HITS-counts (0.81.7/min) were observed. In the worst orientation HITS rose to 4366/min. For the SJM the HITS count in the optimum position was 23.424/min and in the worst orientation 3848/min. Conclusions: Valve orientation has an important impact on microembolization to the brain. In the optimum orientation (large orifice facing the non-coronary leaflet) the Medtronic Hall valve showed neglectible incidence of HITS. The St. Jude Medical bileaflet valve showed less variation but demonstrated significant HITS counts at any orientation. As the MH in the worst position shows significantly higher turbulent stresses than the SJM but no higher incidence of HITS, a strong correlation between turbulence and HITS was not demonstrated by this study.
Key Words: High intensity transient signals Prosthetic heart valves Medtronic St. Jude Medical
This article has been cited by other articles:
![]() |
L. Guerrieri Wolf, B. P. Choudhary, Y. Abu-Omar, and D. P. Taggart Solid and gaseous cerebral microembolization after biologic and mechanical aortic valve replacement: Investigation with multirange and multifrequency transcranial Doppler ultrasound J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 512 - 520. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Machler, G. Reiter, M. Perthel, U. Reiter, P. Bergmann, M. Zink, R. Rienmuller, and J. Laas Influence of a tilting prosthetic mitral valve orientation on the left ventricular flow -- an experimental in vivo magnetic resonance imaging study Eur. J. Cardiothorac. Surg., July 1, 2007; 32(1): 102 - 107. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Hassanein, A. Albert, I. Florath, Y. Y. Hegazy, U. Rosendahl, S. Bauer, and J. Ennker Concomitant aortic valve replacement and coronary bypass: the effect of valve type on the blood flow in bypass grafts Eur. J. Cardiothorac. Surg., March 1, 2007; 31(3): 391 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
N C Radu, E W M Kirsch, M-L Hillion, F Lagneau, L Drouet, and D Loisance Embolic and bleeding events after modified Bentall procedure in selected patients Heart, January 1, 2007; 93(1): 107 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Nyboe, J. A. Funder, M. H. Smerup, H. Nygaard, and J. M. Hasenkam Turbulent stress measurements downstream of three bileaflet heart valve designs in pigs. Eur. J. Cardiothorac. Surg., June 1, 2006; 29(6): 1008 - 1013. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Machler, M. Perthel, G. Reiter, U. Reiter, M. Zink, P. Bergmann, A. Waltensdorfer, and J. Laas Influence of bileaflet prosthetic mitral valve orientation on left ventricular flow--an experimental in vivo magnetic resonance imaging study Eur. J. Cardiothorac. Surg., October 1, 2004; 26(4): 747 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Kleine, M. Scherer, U. Abdel-Rahman, A. A. Klesius, H. Ackermann, and A. Moritz Effect of mechanical aortic valve orientation on coronary artery flow: Comparison of tilting disc versus bileaflet prostheses in pigs J. Thorac. Cardiovasc. Surg., November 1, 2002; 124(5): 925 - 932. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kofidis, S. Fischer, R. Leyh, H. Mair, M. Deckert, R. Haberl, A. Haverich, and B. Reichart Clinical relevance of intracranial high intensity transient signals in patients following prosthetic aortic valve replacement Eur. J. Cardiothorac. Surg., January 1, 2002; 21(1): 22 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Butchart, H.-H. Li, N. Payne, K. Buchan, and G. L. Grunkemeier Twenty years' experience with the Medtronic Hall valve J. Thorac. Cardiovasc. Surg., June 1, 2001; 121(6): 1090 - 1100. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |