EJCTS Click here for details of sales representative
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shin, H.
Right arrow Articles by Kawada, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shin, H.
Right arrow Articles by Kawada, S.

Eur J Cardiothorac Surg 2000;17:492-494
© 2000 Elsevier Science NL


Case report

Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition

Hankei Shin, Ryohei Yozu, Tomohiko Sumida, Shiaki Kawada

Division of Cardiovascular Surgery, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo 160-8582, Japan

Corresponding author. Tel.: +81-3-5363-3804; fax: +81-3-5379-3034
e-mail: h-shin{at}med.keio.ac.jp

We describe a rare intraaortic balloon pump (IABP) vascular complication as a result of malpositioning of the IABP. A 61-year-old man with unstable angina underwent emergency coronary artery bypass grafting soon after the insertion of an IABP. Postoperative hemodynamics were stable, but acute hepatic dysfunction occurred on the second postoperative day. Doppler echography revealed the absence of hepatic arterial flow. The IABP was removed, and arterial flow was immediately restored. Thereafter, the hepatic function recovered rapidly. This is a rare case that demonstrates how IABP can cause mechanical abdominal arterial branch obstruction. Evaluations using Doppler echography are useful in detecting such IABP complications.

Key Words: Intraaortic balloon pump • Acute hepatic failure • Doppler echography




This article has been cited by other articles:


Home page
ANGIOLOGYHome page
J. Mand'ak, V. Lonsky, J. Dominik, and P. Zacek
Vascular Complications of the Intra-aortic Balloon Counterpulsation
Angiology, January 1, 2005; 56(1): 69 - 74.
[Abstract] [PDF]


Home page
Ann. Thorac. Surg.Home page
R. Pandey, G. Ciotti, and M. Pozzi
Anomalous origin of the left coronary artery from the pulmonary artery: results of surgical correction in five infants
Ann. Thorac. Surg., November 1, 2002; 74(5): 1625 - 1630.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2000 European Association for Cardio-Thoracic Surgery. Published by Elsevier. All rights reserved.