EJCTS Click here for details of sales representative
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mächler, H.
Right arrow Articles by Laas, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mächler, H.
Right arrow Articles by Laas, J.
Related Collections
Right arrow Valve disease

Eur J Cardiothorac Surg 2004;26:747-753
© 2004 Elsevier Science NL


Influence of bileaflet prosthetic mitral valve orientation on left ventricular flow—an experimental in vivo magnetic resonance imaging study

H. Mächlera,*, M. Perthelb, G. Reiterc, U. Reiterc, M. Zinkd, P. Bergmanna, A. Waltensdorferd, J. Laasb

a Division of Cardiac Surgery, University Medical Center, Medical University, A-8036 Graz, Austria
b Division of Cardiac Surgery, Bad Bevensen, Germany
c Division of Radiology, University Medical Center, Medical University, A-8036 Graz, Austria
d Division of Anesthesiology, University Medical Center, Medical University, A-8036 Graz, Austria

Received 13 January 2004; received in revised form 7 June 2004; accepted 7 June 2004.

* Corresponding author. Tel.: +43-316-385-4671; fax: +43-316-385-4672. (E-mail: heinrich.maechler{at}uni-graz.at).

Objective: Orientation-related bileaflet mechanical valve flow and velocity studies in the downstream area are limited in mitral valve replacement studies. Methods: In five sheep, ventricular blood flow was visualized prior to the implantation of a mitral Edwards Mira Bileaflet Mechanical Valve Model 9600. The implant orientation was either anatomic, with a 45° rotation, or anti-anatomic, with a 90° rotation. Sheep were positioned within an 1.5T field strength MR scanner (Magnetom Sonata; Siemens) to assess time-dependent three-dimensional blood flow velocities displayed as color-encoded vectors. Results: The preoperative ventricular velocity profiles presented negligible individual variances. Streamlines passed homogeneously without any spatial differences into the left ventricle. Starting from the anatomical position, the areas with inhomogeneous and accelerated local blood velocities increased in comparison to the preoperative status. Rotating the prosthesis until it was in a 45° position caused a significant increase in turbulence immediately downstream; fluids stagnated longer at the apex. In the anti-anatomic orientation, mean velocities decreased. In all three positions, but less so in the anatomical position, the flow pattern of the blood helix at the apex was disturbed. The intraventricular flow patterns between protheses in the three orientations were, however, not significant when compared to the differences between physiologic intraventricular flow and any of the postoperative measurements. Conclusions: To achieve optimal hemodynamics, rotation of the mitral valve has to be considered carefully, as has long been known from aortic valve replacement studies. To this end, a method for qualitative assessment of left ventricular blood flow patterns was developed.




This article has been cited by other articles:


Home page
Am. J. Physiol. Heart Circ. Physiol.Home page
H. Watanabe, S. Sugiura, and T. Hisada
The looped heart does not save energy by maintaining the momentum of blood flowing in the ventricle
Am J Physiol Heart Circ Physiol, May 1, 2008; 294(5): H2191 - H2196.
[Abstract] [Full Text] [PDF]


Home page
Eur. J. Cardiothorac. Surg.Home page
H. Machler, G. Reiter, M. Perthel, U. Reiter, P. Bergmann, M. Zink, R. Rienmuller, and J. Laas
Influence of a tilting prosthetic mitral valve orientation on the left ventricular flow -- an experimental in vivo magnetic resonance imaging study
Eur. J. Cardiothorac. Surg., July 1, 2007; 32(1): 102 - 107.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2004 European Association for Cardio-Thoracic Surgery. Published by Elsevier. All rights reserved.