Eur J Cardiothorac Surg 2006;29:93-99
© 2006 Elsevier Science NL
a Alder Hey Royal Children Hospital, Eaton Road, Liverpool, L12 2AP, England, UK
b MOX, Politecnico di Milano, Milano, Italy
c Ecole Polytéchnique Fédérale Lausanne, Lausanne, Switzerland
d Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland
Received 19 June 2005; received in revised form 16 September 2005; accepted 17 October 2005.
* Corresponding author. Tel.: +44 151 2525713; fax: +44 151 2525643. (Email: Antonio.Corno{at}rlc.nhs.uk).
Objective: To evaluate the differences between non-circular shape of FloWatch®-PAB and conventional pulmonary artery (PA) banding. Methods: Geometrical analysis. Conventional banding and FloWatch®-PAB perimeters were plotted against cross-sections. Computational fluid dynamics (CFD) model. CFD compared non-circular FloWatch®-PAB cross-sections with conventional banding regarding pressure gradients. Clinical data. Seven children, median age 2 months (7 days to 3 years), median weight 4.2 kg (3.29.8 kg), with complex congenital heart defects underwent PA banding with FloWatch®-PAB implantation. Results: Geometrical analysis. Conventional banding: progressive reduction of cross-sections was accompanied by progressive reduction of PA perimeters. FloWatch®-PAB: with equal reduction of cross-sections the PA perimeter remained constant. CFD model. Non-circular and circular banding provided same trans-banding pressure gradients for same cross-sections at any given flow. Clinical data. Mean PA internal diameter at banding was 13.3 ± 4.5 mm. After a mean interval of 5.9 ± 3.7 months, all children underwent intra-cardiac repair and simple FloWatch®-PAB removal without PA reconstruction. Mean PA internal diameter with FloWatch®-PAB removal increased from 3.0 ± 0.8 to 12.4 ± 4.5 mm (normal mean internal diameter for the age = 9.9 ± 1.6). No residual pressure gradient was recorded in correspondence of the site of the previous FloWatch®-PAB implantation in 6/7 patients, 10 mmHg peak and 5 mmHg mean gradient in 1/7. Conclusions: The non-circular shape of FloWatch®-PAB can replace conventional circular banding with the following advantages: (a) the pressure gradient will remain essentially the same as for conventional circular banding for any given cross-section, but with significantly smaller reduction of PA perimeter; and (b) PA reconstruction at the time of de-banding for intra-cardiac repair can be avoided.
Key Words: Congenital heart disease Pulmonary arteries Pulmonary artery banding Pulmonary hypertension Surgery Ventricular septal defect
This article has been cited by other articles:
![]() |
A. F. Corno Invited Commentary Ann. Thorac. Surg., February 1, 2008; 85(2): 598 - 598. [Full Text] [PDF] |
||||
![]() |
A. F. Corno and E. S. Mickaily-Huber Comparative computational fluid dynamic study of two distal Contegra conduit anastomoses Interactive CardioVascular and Thoracic Surgery, February 1, 2008; 7(1): 1 - 5. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Assad, M. M. Zamith, M. F. Silva, P. G. Thomaz, L. A. Miana, V. C. Guerra, C. A. Pedra, and M. B. Marcial A Novel Adjustable Pulmonary Artery Banding System for Hypoplastic Left Heart Syndrome Ann. Thorac. Surg., December 1, 2007; 84(6): 2081 - 2084. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, E. J. Ladusans, M. Pozzi, and S. Kerr FloWatch versus conventional pulmonary artery banding. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1413 - 1420. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, M. Pozzi, and L. K. von Segesser FloWatch and pseudoaneurysm: Complication versus coincidence J. Thorac. Cardiovasc. Surg., April 1, 2006; 131(4): 928 - 929. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |