a Department of Thoracic and Cardiovascular Surgery, Cardiovascular Hospital (Louis Pradel), 69394 Lyon cedex 03, France
b CREATIS, UMR 5515, U630 INSERM, Department of Cardiovascular Radiology, Cardiovascular Hospital (Louis Pradel), 69394 Lyon cedex 03, France
c SEGULA TECHNOLOGIES, 25 rue du Lyonnais, 69800 St Priest, France
Received 16 August 2007; received in revised form 7 October 2007; accepted 15 October 2007.
* Corresponding author. Address: 28, Avenue du Doyen Lépine, 69500 Bron, France. Tel.: +33 472119014; fax: +33 472357291. (Email: sulaiman{at}creatis.univ-lyon1.fr).
Purpose: It is essential to evaluate new stent designs before in vivo testing. The purpose of this study was to develop and validate a controlled and reproducible patient-derived process to produce a life-size in vitro model of aortic arch aneurysm for endovascular procedure simulation. Methods: A three-dimensional magnetic resonance angiography (3D MRA) image derived from a 60-year-old patient with aortic arch aneurysm was segmented using a home-made software package which allows one-click automatic segmentation of the aorta, meshing, and conversion to standard tessellation language (STL) format. A rapid prototyping technique established a stereolithographic model to produce a replica of the whole aorta, including the arch aneurysm and supra-aortic arteries. Results: The final model was made by pouring silicone rubber to obtain a sturdy, life-size, soft, transparent, plastic cast, accurately reproducing both the internal and external anatomy of the aortic aneurysm. This model was used under perfusion by an extracorporeal circulation pump, to test ex vivo stent deployment. Conclusion: The combination of easy segmentation and conversion to the STL format with industrial stereolithography techniques enabled a realistic silicon vascular phantom to be created for endovascular procedure simulation, image modality calibration, and new stent design.
Key Words: Stereolithography Aortic aneurysm Interventional radiology
This article has been cited by other articles:
![]() |
R. Sodian, D. Schmauss, C. Schmitz, A. Bigdeli, S. Haeberle, M. Schmoeckel, M. Markert, T. Lueth, F. Freudenthal, B. Reichart, et al. 3-dimensional printing of models to create custom-made devices for coil embolization of an anastomotic leak after aortic arch replacement. Ann. Thorac. Surg., September 1, 2009; 88(3): 974 - 978. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. M. Carter and M. B. Marshall Open lobectomy simulator is an effective tool for teaching thoracic surgical skills. Ann. Thorac. Surg., May 1, 2009; 87(5): 1546 - 1551. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kalejs and L. K. v. Segesser Rapid prototyping of compliant human aortic roots for assessment of valved stents Interactive CardioVascular and Thoracic Surgery, February 1, 2009; 8(2): 182 - 186. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |