Objective: Advancing life expectancy with the increased prevalence of aortic valve degenerative disease brings the need for an aortic bioprosthesis with excellent haemodynamic performance and comparable durability. The Mitroflow bioprosthesis has been on the worldwide market, except in the United States, since 1982, while the current model (1991) has only recently gained regulatory approval in the latter country. This study was primarily performed to determine the durability of the current Mitroflow bioprosthesis. Methods: The contemporary Mitroflow bioprosthesis was implanted in 381 patients in three centres. The mean age was 76.4 years (range 53–91 years) and the mean follow-up period was 5.4 ± 3.4 years, a total of 2048.7 years of evaluation. Prosthesis–patient mismatch (PPM) was classified by reference effective orifice area index categories: normal
0.85 cm2
m–2 (53.9%), mild 0.84–0.76 cm2
m–2 (33.9%), moderate
0.75–0.66 cm2
m–2 (11.7%) and severe
0.65 cm2
m–2 (0.5%). Results: The survival, at 10 years, was 39.9 ± 7.9% for 50–69 years, 27.0 ± 3.7% for 70–79 years and 16.6 ± 4.4% for
80 years (p
= 0.011). There was a trend (p = 0.063) influencing survival for moderate-to-severe PPM. Of the independent predictors influencing survival – moderate-to-severe projected effective orifice area index (pEOAI) (Hazard Ratio (HR) 1.6, p
= 0.0142) and left ventricular dysfunction (ejection fraction <35%) (HR 1.9, p
= 0.0193) were included. The 10-year freedom from structural valve deterioration (SVD) at explant assessing the same age groups as survival was not different (p
= 0.081). The 10-year actual/actuarial freedom from SVD, at explant was for
60 years – 94.4 ± 1.4% (85.2 ± 3.9%), for
65 years – 94.2 ± 1.4% (85.0 ± 4.0%), for 61–70 years – 97.4 ± 2.6% (95.7 ± 4.3%) and for >70 years – 94.0 ± 1.5% (83.2 ± 4.6%). Conclusions: The Mitroflow external mounted, pericardial aortic bioprosthesis with documented excellent haemodynamics (especially for the small aortic root), demonstrates that prosthesis–patient mismatch in moderate and severe categories can essentially be eliminated, with durability performance comparable to other heterograft (porcine and pericardial) bioprostheses.
Key Words: Heart valve Bioprosthesis Aortic valve replacement Prosthesis