European Journal of Cardio-Thoracic Surgery, Vol 6, 267-271, Copyright © 1992 by European Association for Cardio-thoracic Surgery
RT Johnston, NA Weerasena, M Butterfield, J Fisher and TJ Spyt
The hydrodynamic and haemodynamic function of 29 mm mitral and 23 mm aortic
St Jude Medical (SJM) and CarboMedics (CM) heart valves have been compared
in a pulsatile flow simulator in the laboratory and by doppler ultrasound
in vivo. The laboratory studies showed that there was no significant
difference in the pressure drop across the valves. The valves also had
similar regurgitant volumes. Doppler ultrasound confirmed these results
with a mean pressure drop of 12.9 +/- 5.5 mmHg across the CM aortic valves
and 12.6 +/- 5.5 mmHg for the SJM aortic valves. The mean mitral diastolic
pressure drops were 3.8 +/- 1.8 and 4.6 +/- 1.4 mmHg for the CM and SJM
valves, respectively. Mild aortic regurgitation was detected in 5 of 14
patients in the CM group and 1 out of 6 patients in the SJM group, though
none of the cases were clinically significant. Both the in vitro and in
vivo results show these valves to have acceptable and comparable function.
ARTICLES
Carbomedics and St. Jude Medical bileaflet valves: an in vitro and in vivo comparison
Regional Cardiothoracic Unit, Groby Road Hospital, Leicester, UK.
This article has been cited by other articles:
![]() |
T. Gudbjartsson, T. Absi, and S. Aranki Mitral Valve Replacement Card. Surg. Adult, January 1, 2008; 3(2008): 1031 - 1068. [Full Text] |
||||
![]() |
T. Gudbjartsson, S. Aranki, and L. H. Cohn Mechanical/Bioprosthetic Mitral Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 951 - 986. [Full Text] |
||||
![]() |
J. F. Nistal, A. Hurle, J. M. Revuelta, and M. Gandarillas CLINICAL EXPERIENCE WITH THE CARBOMEDICS VALVE: EARLY RESULTS WITH A NEW BILEAFLET MECHANICAL PROSTHESIS J. Thorac. Cardiovasc. Surg., July 1, 1996; 112(1): 59 - 68. [Abstract] [Full Text] |
||||
![]() |
A. Laske, R. Jenni, M. Maloigne, G. Vassalli, O. Bertel, and M. I. Turina Pressure Gradients Across Bileaflet Aortic Valves by Direct Measurement and Echocardiography Ann. Thorac. Surg., January 1, 1996; 61(1): 48 - 57. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |