European Journal of Cardio-Thoracic Surgery, Vol 7, 84-90, Copyright © 1993 by European Association for Cardio-thoracic Surgery
A Moritz, W Klepetko, S Rodler, A Foger, W Schreiner, F Grabenwoger and E Wolner
The Duromedics Edwards valve (DE) was designed with a self-irrigating hinge
mechanism to reduce the rate of thromboembolic complications. This report
presents a prospective follow-up of patients after valve replacement with
this prosthesis in a university hospital. Five hundred seven patients had
DE prostheses implanted in the aortic (n = 268), mitral (n = 183) or aortic
and mitral positions (n = 56). The perioperative mortality was 6.9%.
Follow-up was 98% complete, comprising 2009 patient years for a mean
follow-up of 48 months: (range 27-84). The actuarial freedom from
complications is calculated as follows (linearized rates in parentheses):
late mortality 81.0 +/- 2.3% (3.4%/patient year), thromboembolism 93.5 +/-
1.5% (1.1%/patient year), anticoagulation-related hemorrhage 89.5 +/- 2.0%
(1.9%/patient year), prosthetic valve endocarditis 96.0 +/- 1.2%
(0.7%/patient year), valve- related mortality 93 +/- 1.8% (1.5%/patient
year), valve failure 89 +/- 2.0% (2%/patient year), treatment failure 88.0
+/- 2.0% (2.2%/patient year) and all valve-related morbidity and mortality
74.0 +/- 2.8% (5.3%/patient year). Two events of leaflet escape were
observed in the study group (0.09%/patient year). Both patients were
reoperated successfully. Mechanical hemolysis was subclinical in all cases.
The DE shows a complication rate comparable to other modern mechanical
valve prostheses. After solving the problem of durability, reconsideration
of this valve is worthwhile due to its low risk of thromboembolic
complications.
ARTICLES
Six-year follow-up after heart valve replacement with the Edwards Duromedics bileaflet prosthesis
Second Surgical Department, University of Vienna, Wien, Austria.
This article has been cited by other articles:
![]() |
W. B. Hemmer, M. Doss, A. Hannekum, and X. Kapfer Leaflet escape in a TEKNA and an original duromedics bileaflet valve Ann. Thorac. Surg., March 1, 2000; 69(3): 942 - 944. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Podesser, G. Khuenl-Brady, E. Eigenbauer, S. Roedler, A. Schmiedberger, E. Wolner, and A. Moritz Long-term results of heart valve replacement with the Edwards Duromedics bileaflet prosthesis: A prospective ten-year clinical follow-up J. Thorac. Cardiovasc. Surg., May 1, 1998; 115(5): 1121 - 1129. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |