EJCTS Click here to go to Siemens website
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kallis, P.
Right arrow Articles by Treasure, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kallis, P.
Right arrow Articles by Treasure, T.

European Journal of Cardio-Thoracic Surgery, Vol 7, 206-210, Copyright © 1993 by European Association for Cardio-thoracic Surgery


ARTICLES

Early results of intravascular oxygenation

P Kallis, NM al-Saady, ED Bennett and T Treasure
Department of Cardiothoracic Surgery, St. George's Hospital, London, UK.

We have used the intravascular oxygenator (IVOX) in eight patients (14- 76 years) with acute respiratory failure. At the time of referral for IVOX all patients were on inotropic support, five had pneumothoraces and two were on haemofiltration. Ventilatory data before IVOX were: ventilated for 2-14 days (median 6), PEEP 5-12 cm H2O (12), PIP 34-95 cm H2O (49), FiO2 0.7-1 (1), PaO2 5.4-26 KPa (8.5) and PaCO2 3.7-23 KPa (8). The intravascular oxygenator was used for 6 h-12 days (median 4), the oxygen transfer achieved was 68-140 ml/min (85) and the carbon dioxide removed was 40-106 ml/min (59). It was possible to decrease the ventilator settings with a significant increase in the PaO2 by 48 h (P < 0.05) but the fall in PaCO2 did not reach significance. As the IVOX was turned from "off" to "on" the mean PaO2 increase was 0.35 KPa +/- 0.14 SEM (P = 0.04) and the mean PaCO2 decrease 0.7 KPa +/- 0.2 SEM (P = 0.02) without any significant change in cardiac output. Two patients survived and six died from multisystem failure with three patients being hypoxic at the time of death. Postmortem examinations on five patients did not reveal any IVOX-related complications. We conclude that IVOX is safe but it is not as efficient in gas exchange as extracorporeal membrane oxygenation at present. Further improvement in the gas exchange efficiency of this prototype could render IVOX a very useful device.


This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
T. Mihaljevic, L. K. von Segesser, M. Tonz, B. Leskosek, and M. I. Turina
Influence of Hemodynamics on the Performances of Intravascular Gas Exchangers
Ann. Thorac. Surg., December 1, 1995; 60(6): 1665 - 1670.
[Abstract] [Full Text]