The hydrodynamic function of six size 19 mm bileaflet mechanical heart
valves was investigated in a pulsatile flow simulator. The St. Jude
Standard, St. Jude Haemodynamic Plus, CarboMedics, Sorin Biomedica,
Duromedics and ATS valves were studied. The mean pressure drop and
effective orifice area during forward flow, regurgitant volumes and total
energy loss were analysed. The St Jude Haemodynamic Plus and Sorin
Biomedica valves had significantly lower forward flow pressure drops and
total energy losses than the other four valves. Assuming that all six
valves can be implanted in the equivalent size of aortic root, this should
lead to improved in vivo performance.
ARTICLES
Comparative study of the hydrodynamic function of six size 19 mm bileaflet heart valves
Department of Mechanical Engineering, University of Leeds, UK.
This article has been cited by other articles:
![]() |
F. Bakhtiary, O. Dzemali, U. Steinseiffer, C. Schmitz, B. Glasmacher, A. Moritz, and P. Kleine Opening and closing kinematics of fresh and calcified aortic valve prostheses: An in vitro study J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 657 - 662. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1058 - 1064. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Otero, J. L. Pomar, J. M. Revuelta, and J. J. Rufilanchas Comparative Evaluation of Small-Size Sorin Slimline and St. Jude HP Heart Valve Prostheses Ann. Thorac. Surg., April 1, 2005; 79(4): 1284 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio, L. Caprili, D. Casarotto, and G. Gerosa Small aortic annulus: The hydrodynamic performances of 5 commercially available bileaflet mechanical valves J. Thorac. Cardiovasc. Surg., September 1, 2004; 128(3): 457 - 462. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Albes, M. Hartrumpf, V. Rudolph, T. Krempl, E. Huttemann, R. Vollandt, and T. Wahlers Are mechanical valves with enhanced inner diameter advantageous in the small sized aortic annulus? Ann. Thorac. Surg., November 1, 2003; 76(5): 1564 - 1570. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund, S. L. Nielsen, H. Arildsen, L. B. Ilkjaer, and H. K. Pilegaard Standard aortic St. Jude valve at 18 years: performance profile and determinants of outcome Ann. Thorac. Surg., May 1, 2000; 69(5): 1459 - 1465. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund, K. Emmertsen, T. T. Nielsen, F. T. Jensen, C. Flo, H. K. Pilegaard, B. S. Rasmussen, O. K. Hansen, and L. H. Kristensen Impact of Size Mismatch and Left Ventricular Function on Performance of the St. Jude Disc Valve After Aortic Valve Replacement Ann. Thorac. Surg., May 1, 1997; 63(5): 1227 - 1234. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |