European Journal of Cardio-Thoracic Surgery, Vol 9, 259-267, Copyright © 1995 by European Association for Cardio-thoracic Surgery
DJ Wheatley, MM Tolland, V Pathi, GM Bernacca, GD Murray, KG Davidson, P Breeze and SM Cobbe
Between 2 February 1987 and 20 March 1990, 170 patients were randomly
allocated to receive a new pericardial heart valve (the Bioflo) or the
Carpentier-Edwards supra-annular porcine bioprosthesis. Eighty-five
patients (mean age 61 years, range 38-77) received 93 Bioflo valves, 46
having aortic valve replacement (AVR), 31 mitral valve replacement (MVR)
and 8 aortic and mitral valve replacement (A+MVR); 85 patients (mean age
62.1 years, range 41-77) received 99 Carpentier-Edwards porcine valves (48
AVR, 23 MVR and 14 A+MVR). Cumulative follow-up totals 926 patient-years
(mean 5.45 +/- 1.93, median 6.03, maximum 7.47 years). The overall
operative mortality rate for the Bioflo group was 2.4%, and 5.9% for the
Carpentier-Edwards group. At 7 years, there was no statistically
significant difference in survival or in any prosthesis-related
complication between the pericardial and porcine valve recipients overall,
or when the data were analysed according to valve implant position. The
actuarial survival rate at 7 years for the Bioflo group was 80.1 +/- 5.1%
and 72.3 +/- 5.9% for the Carpentier- Edwards group. Freedom from
structural valve deterioration (SVD) at 7 years for Bioflo patients was
98.5 +/- 1.5% and for Carpentier-Edwards patients 91.2 +/- 4.1%. No patient
in the Bioflo group has required reoperation for SVD. The randomised
prospective trial has proven practical and effective and has shown the
pericardial valve to perform at least as well as the porcine valve up to 7
years when all of the standard outcome measures of valve performance are
assessed.
ARTICLES
Randomised, prospective evaluation of a new pericardial heart valve: outcome after seven years
Department of Cardiac Surgery, Glasgow Royal Infirmary University NHS Trust, Scotland, UK.
This article has been cited by other articles:
![]() |
T. Le Tourneau, A. Vincentelli, G. Fayad, C. Savoye, O. H. Fabre, A. Prat, and H. Warembourg Ten-year echocardiographic and clinical follow-up of aortic Carpentier-Edwards pericardial and supraannular prosthesis: a case-match study Ann. Thorac. Surg., December 1, 2002; 74(6): 2010 - 2015. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Borowiec, T. W. Dubiel, H.-E. Hansson, J. Landelius, S.-O. Nystrom, and J. W. Borowiec Pericarbon Pericardial Valve Prosthesis: Midterm Results of the Aortic Valve Replacement Angiology, January 1, 1998; 49(1): 1 - 11. [Abstract] [PDF] |
||||