Eur J Cardiothorac Surg 1998;14:64-69
© 1998 Elsevier Science NL
Non-invasive measurement of cardiac output during coronary artery bypass grafting
Amram J. Cohena,c,*,
Dimitri Arnaudovb,c,
Deeb Zabeedab,c,
Lex Schultheisd,
John Lashingere,
Arie Schachnera,c
a Department of Cardiovascular Surgery, The Edith Wolfson Medical Center, Holon, 58100 Israel
b Department of Anesthesiology, The Edith Wolfson Medical Center, Holon, 58100 Israel
c Sackler School of Medicine, Tel Aviv University, Tel Aviv, Israel
d Department of Anesthesiology and Critical Care Medicine, Johns Hopkins Medical Center, Baltimore, MD, USA
e Department of Cardiovascular Surgery, Johns Hopkins Medical Center, Baltimore, MD, USA
Received 18 November 1997;
received in revised form 23 March 1998;
accepted 15 April 1998.
* Corresponding author. Tel.: +972 3 5028723; fax: +972 3 5028735.
 |
Abstract
|
|---|
Objective: A new device, using whole body bioresistance measurements and a new equation for calculating stroke volume has been developed. Using this equation, an attempt was made to correlate whole body bioresistance cardiac output with thermodilution cardiac output in patients undergoing coronary artery bypass grafting. Methods: Thirty-one adults undergoing elective coronary artery bypass grafting were studied prospectively. Simultaneous paired cardiac output measurements by whole body bioresistance and thermodilution were made at five time points during coronary artery bypass grafting: in anesthetized patients before incision (T1), after sternotomy (T2), after opening the pericardium (T3), ten min post bypass (T4), and in the intensive care unit (T5). The patients had a mean of three thermodilution cardiac outputs compared with a mean of three bioimpedance measurements at each time point. The bias and precision between the methods were calculated. Results: There was good correlation between bioresistance cardiac output (nCO) and thermodilution cardiac output (ThCO) measurements in both groups for all recorded times. The patients' mean ThCO and nCO, as well as bias and precision between methods were calculated. Mean ThCO ranged between 4.14 and 5.06 l/min; mean nCO ranged between 4.12 and 4.97 l/min. Bias calculations ranged between -0.072 and 0.104 l/min. Precision (2 SD) calculations ranged between 0.873 and 1.228 l/min for 95% confidence intervals. Pearson's correlation ranged from 0.919 to 0.938. Conclusions: Cardiac output measured with the new device correlates well with the thermodilution measurements of cardiac output during and immediately following coronary artery bypass grafting. The overall agreement between the two methods was good. The new device is an accurate non-invasive method of measuring cardiac output during coronary artery bypass grafting.
Key Words: Cardiac output Cardiac surgery Thermodilution Electrical bioimpedance Bioresistance Hemodynamics
 |
Introduction
|
|---|
Hemodynamic monitoring continues to be an integral part of peri- and postoperative care for patients undergoing cardiac surgery. Cardiac output (CO) is an important parameter in these measurements. To date, the clinical tool used to measure CO is the SwanGanz catheter using a thermodilution technique. However, the procedure is invasive, expensive, and may lead to complications
[1]
[2].
A number of attempts have been made to determine CO in a non-invasive manner
[3]
[4]
[5]
[6]. Using thoracic electrical bioimpedance techniques (TEB), moderate success has been achieved in some clinical settings
[3]
[4]
[6]
[7]
[8]. The technique has been unsuccessful when applied to patients undergoing cardiac surgery
[9]
[10]
[11]. A new non-invasive cardiac system device has been developed to utilize whole body bioresistance in a semi-empirical formula to determine the CO. Accuracy of the CO measurement using this device was established comparing thermodilution CO (ThCO) for patients undergoing right and left heart catheterization
[12]. The purpose of this investigation was to compare the new technology for measuring cardiac output with thermodilution measured CO in patients undergoing coronary artery bypass grafting (CABG).
 |
Patients and methods
|
|---|
Patients
Thirty-one patients undergoing elective, primary CABG were prospectively studied; 15 at Wolfson Medical Center and 16 at Johns Hopkins University School of Medicine. Patients with aortic valve insufficiency, mitral and tricuspid regurgitation or cardiac shunt were not included. The CABG was similar in both hospitals. All patients underwent a median sternotomy. Cardiopulmonary bypass was established using an aortic and single venous cannula. Diastolic arrest was achieved with cold sanguineous potassium cardioplegia in both antegrade and retrograde fashion. The internal mammary artery was used to graft the left anterior coronary artery (LAD) in all cases, and saphenous vein grafts were used to bypass the other obstructed vessels. Proximal anastomoses were performed using a partial cross clamp. During the study, no patient had atrial flutter nor atrial fibrillation.
Protocol
The study protocol was approved by Edith Wolfson and Johns Hopkins Institutional Review Boards. For each patient demographic and clinical data was tabulated. In each patient, CO measurements were taken at five different time intervals during the procedure; after induction of anesthesia but before incision (T1), after sternotomy (T2), after creation of a pericardial pocket (T3), ten min after weaning from bypass (T4), and immediately after arrival in the intensive care unit (T5).
At each time interval, three adequate thermodilution measurements were made. All measurements were made at end expiration during the respiratory cycle. Thirty-one patients underwent 155 thermodilution CO measurements. Nineteen were excluded because of 10% variation between the measurements. Simultaneously, three bioresistance measurements were taken at each interval, and their average was considered the bioresistance CO.
Each bioresistance measurement took 20 s. Since these measurements were continuous, there was no time between measurements such that to achieve an average bioresistance CO measurement took 1 min.
Thermodilution method
At both hospitals seven French true size thermodilution catheters (Baxter Healthcare, Irvine, CA) were inserted in the operating room after induction of anesthesia. The proper location of the thermodilution catheter was confirmed by hemodynamic measurements and by postoperative chest roentgenograms. The pulmonary artery catheter was connected to a thermodilution cardiac output monitor (`Horizon 2000', Mennen Medical, Israel was used at Wolfson Medical Center, and 7010 Series Marquette, Madison, WI at Johns Hopkins Institutions). Ten milliliters of room temperature 5% dextrose solution was injected manually at end-expiration by an experienced anesthesiologist who was unaware of the bioresistance CO results.
Whole body resistance method
NICaSTM 2001, a non-invasive cardiac system device (NICaSTM 2001, Teledyne-NIM, LLC, North Andover, MA, 510K, certificate number K942227), was utilized for measuring bioresistance CO. Two proprietary designed, NICaSTM disposable electrodes (510K, certificate number K972002) were applied to each wrist, and were then connected to the non-invasive cardiac system device. The system passed an AC current 30 kHz and 1.4 mA through the whole body. It then measured the resistive portion of the bioimpedance. Since the electrodes were placed on the distal aspect of the extremities, the current passed through all major arteries and veins of the systemic circulation. As such, the system measured the bioresistance of the systemic circulation and allowed calculation of the stroke volume.
In 1992, Tsoglin and Frinerman and developed a new semi-empirical formula for calculating the stroke volume (SV)
[13].
where Kel is a coefficient related to blood electrolytes, Kweight is a ratio of a patient's weight to ideal weight, IB is the index balance equal to the ratio of the measured extracellular fluid (as measured from the baseline impedance) to the expected extracellular fluid volume, HCTcorr is a hematocrit correction factor proportional to measured hematocrit, Ksex,age is a coefficient depending upon a patient's sex and age, H is a patient's height,
R is the incremental change of the resistive portion of the bioimpedance, R is the baseline whole body bioresistance and
+ß/ß is the ratio of the sum of the systole time plus diastole time divided by the diastole time derived from the fine structure of the varying portion of the bioresistance.
Using whole body resistance measurements and the above formula, SV was calculated and converted to CO.
Statistical analysis
The correlation between methods at each time point was evaluated by Pearson's correlation coefficient and simple linear regression. The degree of agreement between methods at each time point was evaluated by calculation of the bias (mean between-method difference) and precision (mean bias ±2 SD)
[14].
 |
Results
|
|---|
Thirty-one patients were evaluated in the study. Demographic and clinical data are shown in Table 1. The agreement between the average CO measured by bioresistance versus thermodilution is shown in Table 2. The plot of the regression analysis and difference versus mean for all measurements at all times in the 31 patients is shown in
Fig. 1
a,b. There was good correlation between ThCO and nCO ranges of cardiac output during all phases of CABG and the immediate postoperative period.

View larger version (22K):
[in this window]
[in a new window]
|
Fig. 1. (a) Linear regression analysis comparing bioresistance measured cardiac output with thermodilution cardiac output for all averaged measurements in the study. (b) Mean difference between bioresistance measured cardiac output and thermodilution cardiac output for all averaged measurements in the study.
|
|
 |
Discussion
|
|---|
The initial attempt to obtain CO by measuring the stroke volume (SV) through thoracic electrical bioimpedance (TEB) was performed by Kubicek
[15] at the National Aeronautics Space Agency (NASA) where he introduced the equation that became the basis for bioimpedance cardiography:
where SV is related to the resistivity of blood (
), dZ/dt is the first peak of the derivative of the bioimpedance curve, L is the distance between the electrodes, Z0 is the mean time averaged thoracic bioimpedance and T is the ventricular ejection fraction.
The equation was modified by Bernstein
[16] to:
where VEPT is a coefficient that represents the volume of electrically participating tissue. Using this equation, limited clinical success has been achieved in determining CO using bioimpedance techniques. Correlation to thermodilution techniques have been achieved in healthy volunteers, patients undergoing operations without cardiopulmonary bypass, patients undergoing procedures in the cardiac catheterization laboratory, and small numbers of intensive care unit patients
[3]
[4]
[7]
[17]
[18]
[19]
[20].
This equation has been problematic due to difficulties in accurately computing VEPT, and its application becomes impractical in a patient undergoing rapid hemodynamic changes. Furthermore, determination of VEPT is dependent upon the accurate placement of electrodes, which is not always possible during open heart surgery. In addition, this equation still depends upon the first derivative of the bioimpedance curve, which is a rapidly fluctuating factor. As a result, attempts to correlate bioimpedance CO in the patient with thermodilution techniques during CABG have been unsuccessful
[9]
[10].
In addition to practical problems, there is a conceptual problem in applying TEB to measure stroke volume. TEB measurements of the bioimpedance includes both systemic and pulmonary circulations. The two circulations cannot be separated by measurements and variation in proportion of these circulation will lead to inaccurate measurements of SV
[21]
[22]. This will not happen in whole body bioresistance measurements where the systemic circulation dominates the measurement and reflects LV SV.
The physiological and physical basis of bioimpedance has been studied by many authors
[23]
[24]
[25]. However, the variation between different tissues and their complex structures make it difficult to model their electrical behavior. In terms of bioresistance, the body may be divided into the `blood compartment' and `tissue compartment'. In the `blood compartment', conductivity is high and therefore a current introduced into the body will travel primarily in this compartment. The resistance changes in this compartment as the blood volume varies in the great arteries due to pulsatile flow. In the `tissue compartment', there is less significant current flow and constant resistance. In each individual, the resistance in the `tissue compartment' is determined by the patient's height, lean body weight (muscle) to fat ratio, sex, age, body build, extracellular fluid volume, and electrolytes. In the `blood compartment', the resistance is determined by hematocrit, electrolytes and the pulsatile blood volume changes within the system. To measure SV, the pulsatile changes within the blood compartment and baseline bioresistance must be measured accurately and all other factors must be accounted for with appropriate correction factors. The measurement of whole body bioresistance allows for the accurate measurement of pulsatile changes and baseline resistance, and the proposed correction factors allow for the accurate calculation of the systemic bioresistance with which accurate LV SV can be calculated.
The method utilized in this study has been proven accurate in patients undergoing catheterization
[12]. It was also shown to be accurate in a pilot study in patients undergoing CABG
[26]. Compared with previous attempts to correlate bioresistance with SV, the equation has major advantages.
- The equation does not depend on rapidly fluctuating time derivatives of bioimpedance.
- The equation uses empirically derived coefficients that are obtained from laboratory values and simultaneously measured bioimpedance values instead of the artificial and difficult to approximate VEPT.
- The precise placement of the electrodes is not a critical factor.
- The NICaSTM electrode arrangement is optimal to measure the left ventricular SV.
- The respiration has almost no effect on the NICaSTM CO measurements in real time.
The new methodology showed good correlation between thermodilution and bioresistance CO during all phases of CABG and the immediate postoperative period in two independent hospital populations. Correlation was good in all ranges, including low cardiac output. This fact is important since patients undergoing CABG frequently have low cardiac output. This was even true in the immediate post cardiopulmonary bypass period where changes in the patients' volume status, temperature, blood electrolytes and hemodynamics are changing rapidly. Such a correlation would be impossible using previous techniques.
The new device has certain limitations. First, it cannot measure CO while using diathermy. Second, the device is sensitive to movement so that the patient cannot be manipulated while the CO is being measured. These two factors require that the operation stop for the 2030 s required to measure CO. Finally, the device measures SV and multiplies it by heart rate to measure CO. Arrythmias in which the heart rate varies significantly will result in a non-representative measurement of CO.
In summary, a new device using a semi-empirical equation relating SV to changes in the resistive portion of the patients' bioresistance has been developed. Information about specific patient data, blood components and body habitus inserted into the analytical software has significantly improved the ability to calculate SV from the bioresistance data. The device allows accurate, easy to obtain, non-invasive CO during cardiac surgery. Within the above stated limitations, we have confirmed the validity of the new bioresistance methodology in 31 patients who underwent CABG. Bioresistance measured cardiac output correlated well with ThCO during the CABG procedure and the immediate postoperative period.
 |
Acknowledgments
|
|---|
This paper was prepared in consultation with Diklah Geva who prepared the statistical calculations, and with the technical assistance of Sally Esakov.
 |
References
|
|---|
- Mermel L.A., Maki D.G. Infectious complications of SwanGanz pulmonary artery catheters: pathogenesis, epidemiology, prevention, and management. Am J Respir Crit Care Med 1994;149:1020-1036.[Medline]
- Choh J.H., Khazei A.H., Ihm H.J., Thatcher W.C., Batty P.R. Catheter induced pulmonary arterial perforation during open heart surgery. J Cardiovasc Surg (Torino) 1994;35:61-64.[Medline]
- Denniston J.C., Maher J.T., Reeves J.T., Cruz J.C., Cymerman A., Grover R.F. Measurement of cardiac output by electrical impedance at rest and during exercise. J Appl Physiol 1976;42:91-95.
- Smith S.A., Russell A.E., West M.J., Chalmers J. Automated non-invasive measurement of cardiac output: comparison of electrical bioimpedance and carbon dioxide rebreathing techniques. Br Heart J 1988;59:292-298.[Abstract/Free Full Text]
- Reybrouck T., Fagard R. Assessment of cardiac output at rest and during exercise by a carbon dioxide rebreathing method. Eur Heart J 1990;11(suppl I):21-25.
- Belardinelli R., Ciampani N., Costantini C., Blandini A., Purcaro A. Comparison of impedance cardiography with thermodilution and direct Fick methods for non-invasive measurement of stroke volume and cardiac output during incremental exercise in patients with ischemic cardiomyopathy. Am J Cardiol 1996;77:1293-1301.[Medline]
- White S.W., Quail A.W., DeLeeuw P.W., Traugott F.M., Brown W.J., Porges W.L., Cottee D.B. Impedance cardiography for cardiac output measurement: an evaluation of accuracy and limitation. Eur Heart J 1990;11(Suppl):79-92.
- Teo K.K., Hetherington M.D., Haennel R., Greenwood P.V., Rossall R.E., Kappagoda T. Cardiac output measured by impedance cardiography during maximal exercise tests. Cardiovasc Res 1985;19:737-743.[Medline]
- Sageman W.S., Amundson D.E. Thoracic electrical bioimpedance measurement of cardiac output in postaortocoronary bypass patients. Crit Care Med 1993;21(8):1139-1142.[Medline]
- Clarke D.E., Raffin T.A. Thoracic electrical bioimpedance measurement of cardiac output not ready for prime time. Crit Care Med 1993;21(8):1111-1112.[Medline]
- Thomas A.N., Ryan J., Doran B.R., Pollard B.J. Bioimpedance versus thermodilution cardiac output measurement: the Bomed NCCOM3 after coronary bypass surgery. Intensive Care Med 1991;17(7):383-386.[Medline]
- Miller HI, Frinerman E, Tsoglin A, Frinerman A, Rosenschein U, Keren G, Roth A, Laniado S. Non-invasive bioimpedance measurement of cardiac output: a validation study (Abstr). Second Conference of the Israel Heart Society, 1993:58.
- Tsoglin, A., Frinerman, E, N.I. Medical, Ltd. Non-invasive method and device for collecting measurements representing body activity and determining cardiorespiratory parameters of the human body based upon the measurements collected. US Patent No. 5,469,859, Nov. 28 1995.
- Bland, J.M., Altman, D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;Feb 81(8476):307310.
- Kubicek W.G., Patterson R.P., Witsoe D.A. Impedance cardiography as a non-invasive method of monitoring cardiac function and other parameters of the cardiovascular system. Ann NY Acad Sci 1970;170:724-729.
- Bernstein D.P. A new stroke volume equation for thoracic electrical bioimpedance: theory and rationale. Crit Care Med 1986;14:904-909.[Medline]
- Donovan K.D., Dobb G.J., Woods W.P., Hockings B.E. Comparison of transthoracic electrical impedance and thermodilution methods for measuring cardiac output. Crit Care Med 1986;14:1038-1044.[Medline]
- Salandin V., Zussa C., Risica G., Michielon P., Paccagnella A., Cipolotti G., Simini G. Comparison of cardiac output estimates by thoracic electrical bioimpedance, thermodilution and Flick methods. Crit Care Med 1988;16:1157-1158.[Medline]
- Thomas S.H.L. Impedance cardiography using Sramek-Bernstein method: accuracy and variability at rest and during exercise. Br J Clin Pharmacacol 1992;34:467-476.[Medline]
- Wong D.H., Onishi R., Tremper K.K., Reeves C., Zaccari J., Wong A.B., Miller J.B., Cordero V., Davidson J. Thoracic bioimpedance and Doppler cardiac output measurement: learning curve and interobserver reproducibility. Crit Care Med 1989;17:1194-1198.[Medline]
- Fuller H.P. Evaluation of left ventricular function by impedance cardiography: a review. Prog Cardiovasc Dis 1994;36:267-273.[Medline]
- Patterson R.P. Sources of the thoracic cardiogenic electrical impedance signal as determined by a model. Med Biol Eng Comput 1985;23:41-417.
- Karnegis J.N., Kubicek W.G. Physiological correlates of the cardiac thoracic impedance waveform. Am Heart J 1970;79:519-523.[Medline]
- Sakamoto K., Muto K., Kanai H., Iizuka M. Problems of impedance cardiography. Med Biol Eng Comput 1979;17:697-709.[Medline]
- Paul E., Marik M.D., Judy E. A comparison of hemodynamic parameters divided from transthoracic electrical bioimpedance with those parameters obtained by thermodilution and ventricular angiography. Crit Care Med 1997;25:1545-1550.[Medline]
- Cohen A, Frinerman E, Katz M, Ezra S, Dotan A, Weissberg A, Schachner A. Validity of bioimpedance hemodynamic measurements during coronary artery bypass grafting (CABG) (Abstr). J Cardiovasc Diag Proc 1996;13:57.