Eur J Cardiothorac Surg 1999;15:353-358
© 1999 Elsevier Science NL
University of Ottawa Heart Institute, Ottawa, Ontario, Canada
Received 29 September 1998; received in revised form 21 December 1998; accepted 8 January 1999.
Corresponding author. Tel.: +1-613-761-4720; fax: +1-613-761-4338; e-mail: frubens@heartinst.on.ca
| Abstract |
|---|
|
|
|---|
-agonist and inotrope use. Terminal complement complex (SC5b-9) was measured using an ELISA. Results: Upon initiation of bypass, there was a decrease in mean arterial pressure (MAP) in the control group, not seen in the test group (P=0.0005, ANOVA). There was a decrease in the total volume of replacement fluid given intraoperatively in the test group as compared with the control group (total plus prime; control 5.3±1.2 L, test 4.4±1.9 L, P=0.03, MannWhitney test). There was a trend to decreased need for inotrope infusion in the test group after CPB (test 1/17, control 6/17, Fisher exact test; P=0.085). No difference was seen in the generation of terminal complement complex between the groups either during or after CPB. Conclusions: The decrease in blood pressure in the control group, upon the initiation of CPB, did not occur in patients undergoing CPB with the circuit prepared with the surface modifying additive. The decrease in blood pressure was likely associated with the increase in total administered fluids intraoperatively (approximately 1 l/patient) and perhaps the trend towards higher use of inotropes in the control patients as opposed to the test patients. These haemodynamic changes did not appear to be related to complement activation early in CPB.
Key Words: Cardiopulmonary bypass Complement activation Blood pressure response Biocompatibility
| Introduction |
|---|
|
|
|---|
A novel biomaterial approach to limit systemic events has involved the addition of a polysiloxane-containing co-polymer to the base polymer resin during manufacture of the CPB circuit. The former co-polymer migrates to the surface during fabrication, creating regions of alternating hydrophilic and hydrophobic domains. In in vitro and in vivo studies, there has been promising evidence that the detrimental effects of the blood-biomaterial interaction may be minimized [1] [2]. We have recently completed a clinical trial to evaluate a variety of laboratory and clinical parameters involving this new biomaterial in CPB [3]. Haematological biocompatibility was significantly improved with this modified surface as compared with control circuits. The objective of the current trial was to determine if the dramatic blood changes seen in these patients were paralleled by systemic changes in haemodynamics and complement activation, as determined by generation of the terminal complement complex SC5b-9.
| 2. Materials and methods |
|---|
|
|
|---|
The patients were randomly assigned to surgery with either the test circuit (COBE SMARxTTM Biocompatible Extracorporeal Circuit, COBE Cardiovascular, Arvada, CO) or a standard untreated circuit. The perfusionist assigned the treatment immediately preoperatively, by opening a sealed, numbered envelope prior to set-up of the extracorporeal circuit. The randomization was accomplished using a random number table [4]. The cannulae and tubing used in these two circuits were identical in appearance, so that all members of the surgical and anaesthesia teams, excluding the perfusionist, were blinded to the patient designation.
Anaesthetic management
All cardiac medications excluding aspirin, were continued up to the day of surgery. The premedication included diazepam and morphine. The patients were monitored with a radial arterial pressure line and a SwanGanz catheter, inserted prior to intubation. The anaesthetic induction included midazolam 0.04 mg/kg, sufentanil 0.5 µg/kg, rocuronium 1 mg/kg and ketamine 1.5 mg/kg. Anaesthetic was maintained with midazolam 0.5 µg/kg per min, sufentanil 0.5 µg/kg per h, rocuronium 10 µg/kg per min and isoflurane as required to maintain haemodynamic stability.
Conduct of CPB
Prior to bypass, all patients were anticoagulated with porcine heparin (Organon Teknika, Toronto, Canada) to achieve a kaolin driven ACT>480 s. Initial dosing was prescribed by a heparin dose response with Hepcon instrumentation (Hemotec, Medtronic, Parker, CO). Heparin (4500 units) was also added to the 1.5 l pump prime to establish a 3.0 units ml-1 concentration. Upon initiation of bypass, the heparin serum concentration was monitored every 20 min by way of Hepcon automated protamine assay and subsequent doses of heparin were administered as needed to sustain a heparin level of 3.0 units ml-1
[5]. Following the termination of CPB, heparin reversal was accomplished with a 1.1 protamine/heparin ratio (mg/mg) dose of protamine sulphate (Fujisawa Canada, Ontario, Canada) based on the final heparin serum concentration assayed.
CPB was conducted using a roller pump, a flat sheet polypropylene 1.3 m2 membrane oxygenator (COBE CML DuoTM, COBE Cardiovascular Arvada, CO), a 43 µ arterial filter (COBE SentryTM), a closed venous reservoir bag, ascending aortic cannula, and a two-stage single venous cannula return. As noted above, all of the surfaces in the test group, including the cannulae and the filters, were prepared with the surface modifying additive co-polymer. Bypass flows were maintained at 2.43.2 l/m2 per min. The body temperature was initially reduced to a systemic temperature of 32°C.
Hypotensive episodes, where the mean arterial pressure (MAP) was less than 50 mmHg were treated with 100 µg boluses of phenylephrine until the MAP exceeded 55 mmHg, and blood flows were raised but not to exceed 3.2 l/min per m2. Hypertensive episodes, where the MAP exceeded>90 mmHg, were treated with Isoflurane vapour (0.52.0% of membrane ventilation gas). In addition, minute blood flows were sustained to obtain a minimum venous haemoglobin saturation of 65 and 75% at nasopharyngeal temperatures of 36 and 32°C, respectively. Fluids (Ringers Lactate, pentastarch) were given by the perfusionist as needed to treat hypotension and low filling pressures. Fluids were also given by the anaesthetist after CPB as necessary, to maintain preload up to a pulmonary capillary wedge pressure of 18 mmHg.
Cardiac arrest was achieved using topical pericardial saline irrigation as well as antegrade cold crystalloid cardioplegia via the aortic root or the bypass grafts, at 20-min intervals. As there was no cardiotomy reservoir (integrated or otherwise), cardiac venting was accomplished via the aortic root by gravity drainage directly into the venous CPB line. All CPB solutions were filtered by a 40 µm transfusion filter (SQ40S Pall Biomedical Products, NY). At the completion of the procedure, the patient was rewarmed to a nasopharyngeal temperature of 37°C and then weaned from CPB after mechanical ventilation had been re-commenced. All scavenged blood from the mediastinum, collected intraoperatively and up to 4 h postoperatively, was processed by filtration (30 µm), centrifugation and washing (BRATTM, COBE Cardiovascular) prior to re-infusion.
Collection of blood specimens and measurement of plasma haemoglobin and terminal complement complex
Blood specimens were obtained from the side-port of the central line cordis after removal of six dead space volumes of blood. The sampling was carried out at the following time points.
Blood samples were taken into citrate (3.8%) on ice, then centrifuged to produce platelet poor plasma (PPP).
Routine haematology parameters were then determined including the WBC count (Serono, Model 9000, Baker Diagnostic, Allentown, PA). The Quidel SC5b-9 Enzyme Immunoassay (Quidel, San Diego, CA) was used for the quantitation of the terminal complement complex SC5b-9 in samples of blood.
Management of inotropic agents and fluids after CPB
The anaesthesia staff, during withdrawal from CPB, administered ephedrine hydrochloride or calcium chloride, in the presence of hypotension, if the cardiac contraction, as determined by transesophageal echo or direct inspection was adequate. Intravenous inotrope infusions were administered to patients separating from CPB if the cardiac index was <2.2 l/min per m2 in the presence of an adequate preload. The choice of inotrope (dopamine, dobutamine, noradrenaline or milrinone) was left to the discretion of the anaesthetist. Post-CPB, all patients were given an infusion of D5W:NS at 25 ml/h for the first 24 h. Boluses of fluid (Ringers Lactate or pentastarch) were administered by the intensivist for hypovolemia (PCWP<12 mmHg) in the presence of decreased CI, low urine output (<0.5 ml/kg per h) or hypotension.
Data analysis and statistics
Continuous variables were analyzed using unpaired t-tests and categorical variables were analyzed using a chi-squared test or Fisher's exact test as appropriate. Repeated measures were analyzed using repeated measures ANOVA. If the ANOVA was significant (P<0.05), then unpaired t-tests were done at individual time points using the Bonferroni correction for multiple testing. Data that was not normally distributed were analyzed using the MannWhitney tests.
| Results |
|---|
|
|
|---|
|
|
Upon initiation of CPB, there was notable decrease in the MAP in the control group, not seen in the test group ( Fig. 1 , P=0.0005, ANOVA). Within 5 min of the initiation of CPB however, the pressure increased to match the pressure in the test group, and no further differences in pressure occurred. There was a trend to a decrease in the total administered neosynephrine but this was not significant (test group median 0 mg, interquartile range 1200 mg, control group median 300 mg, interquartile range 1100 mg, P=0.231, MannWhitney test). There was a significant decrease in the total volume of replacement fluid given intraoperatively in the test group as compared to the control group ( Fig. 2 , total plus prime; control 5.3±1.2 l, test 4.4±1.9 l, P=0.03, MannWhitney test). There was no difference in fluid administration in the first 24 h postoperatively.
|
|
|
|
| Discussion |
|---|
|
|
|---|
The systemic and clinical changes in this same group of patients were evaluated in the current study. Here, a marked difference was identified in the blood pressure response of the test and control groups upon the initiation of CPB. The decrease in blood pressure in the control patients was large enough to require an intervention in most cases. This hypotension was likely responsible for the significant increase in the total administered fluids (approximately 1 l per patient). It was also likely related to a trend towards a greater mean dose of neosynephrine during CPB and a trend to the more frequent use of inotropic infusions after CPB in the control patients.
The immediacy of the hypotensive response suggests that a soluble agent is released in the control group, mediating peripheral vasodilation during CPB. By some unknown mechanism, the trigger for this release is inhibited during bypass with the modified surface. One of the principal possibilities that was considered, involved activation of the inflammatory cascade with production of bradykinin and activation of complement. To support this, it has been previously demonstrated that complement activation is an immediate and significant consequence of the blood-biomaterial interaction [8] [9]. This is a result of factor XII activation and the subsequent activation of the kallikrein pathway [8]. Further, terminal complement complex (SC5b-9) is a sensitive indicator of complement activation [2] [10]. However, as there was no difference in this marker between the two groups, we cannot explain the haemodynamic differences on the basis of this portion of the inflammatory cascade.
Another potential mechanism for the decrease in blood pressure, which was not evaluated in detail in this trial, may relate to polymorphonuclear leucocytes (PMN) activation and accumulation within the oxygenator. In support of this, Eberhart et al. have previously demonstrated in a porcine model of CPB, that this surface modifying additive significantly decreases PMN deposition in the oxygenator as compared to bypass with a standard untreated surface [1]. A white blood cells (WBC) differential was not prepared in the present trial so we could not confirm if PMN accumulation in the oxygenator occurred with the same pattern. Leukocytes were not specifically identified on scanning electron micrographs in the arterial filters collected from this trial [3] yet, there was definitely a marked increase in debris on the control surfaces which may have consisted of cellular elements. The nature of this debris will need to be clarified in further studies. In a similar clinical trial assessing circuits prepared with surface modifying additives, Gu et al. [11] did not show any difference in total elastase release between patients undergoing bypass with this test surface, and a control surface. In both groups, elastase did increase, however, the changes were not immediate and did not seem to parallel the rapid hypotensive response seen in our trial. There are many other products that may be released from activated PMNs which should be re-assessed in further trials. Moreover, the time course of their release and any relationship to platelet-neutrophil aggregate formation should be evaluated [12] [13].
Finally, we have considered that the immediate decrease in blood pressure may be related to endothelial-derived relaxing factor (EDRF) release from endothelial cells, triggered by the onset of CPB. This endothelial product is present within immediately-available intracellular stores [14] and is well-known to be the most potent vasodilator in vivo. However, because of its extremely short half-life, it is difficult to identify its presence as a causative agent and levels have not been measured during the course of bypass. Alternatively, but less likely, exposure of blood to the test surface results in release of the counterpart vasoconstrictive agent endothelin-1. However, there are virtually no intracellular stores of this agent, and its production is regulated at the transcriptional level, suggesting a time response for its release that would correlate poorly with the blood pressure changes seen in the present trial [15].
In conclusion, the use of a surface modifying additive prevented the initial hypotension related to the initiation of CPB, in patients undergoing elective coronary artery bypass grafting. The hypotension correlated with increased intraoperative fluid requirements in these patients. The causative mechanism for the protective influence of the test circuit was not related to complement activation. Further trials will be necessary to clarify the inter-relationship of PMN activation and platelet preservation during CPB with this surface, and to evaluate the role of EDRF release in the early stages of CPB in this phenomenon.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. M. Kirshbom, B. E. Miller, K. Spitzer, K. A. Easley, C. E. Spainhour, B. E. Kogon, and K. R. Kanter Failure of surface-modified bypass circuits to improve platelet function during pediatric cardiac surgery. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 675 - 680. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G Raja and G. D Dreyfus Modulation of Systemic Inflammatory Response after Cardiac Surgery Asian Cardiovasc Thorac Ann, December 1, 2005; 13(4): 382 - 395. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Allen, W. T McBride, I. S Young, S. W MacGowan, T. J McMurray, S. Prabhu, S P. Penugonda, and M. A Armstrong A clinical, renal and immunological assessment of Surface Modifying Additive Treated (SMARTTM) cardiopulmonary bypass circuits Perfusion, September 1, 2005; 20(5): 255 - 262. [Abstract] [PDF] |
||||
![]() |
F. D. Rubens, H. Nathan, R. Labow, K. S. Williams, D. Wozny, J. Karsh, M. Ruel, and T. Mesana Effects of Methylprednisolone and a Biocompatible Copolymer Circuit on Blood Activation During Cardiopulmonary Bypass Ann. Thorac. Surg., February 1, 2005; 79(2): 655 - 665. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||