Eur J Cardiothorac Surg 1999;16:211-217
© 1999 Elsevier Science NL
a Department of Thoracic and Cardiovascular Surgery (Klinik für Thorax-, Herz- und Gefäßchirurgie), Johann Wolfgang Goethe University, Theodor-Stern-Kai 7, D-60590 Frankfurt /Main, Germany
b Central Laboratory, Department of Internal Medicine (Zentrallabor, Zentrum der Inneren Medizin), Johann Wolfgang Goethe University, Frankfurt /Main, Germany
Corresponding author. Tel.: +49-69-63016141; fax: +49-69-63015849
| Abstract |
|---|
|
|
|---|
Key Words: Cardiopulmonary bypass Coronary artery bypass grafting Inflammatory whole body response Heparin coating Synthetic protein treatment
| 1. Introduction |
|---|
|
|
|---|
The whole body response to CPB may lead to organ dysfunction [4], and cause temporary biochemical disorders in every patient. Due to major advances in oxygenator technology [5], current research focuses on the biocompatibility of CPB [6], and specifically on reducing the pathogenicity of CPB surfaces by coating [7] or modification [8]. Additional technologies such as leucocyte or particle filtration are evolving concepts to prevent CPB-related pathogenicity.
The historical perspective for the development of heparin coating technologies is: (1) Possible reduction or avoidance of anticoagulation for CPB [9,10]; and (2) prevention of the interaction between blood and foreign surfaces. The question remains, if the pharmacodynamics of heparin bound to artificial surfaces may serve as the explanation for biochemical benefits as seen by several researchers [5,11,12]. The synthetic protein treatment (SafelineTM; Jostra AG, Hirrlingen, Germany) may be an appropriate tool to answer this question, because it prevents contact of blood with artificial surfaces to the membrane, but does not employ heparin. SafelineTM involves the physically stable binding of a synthetic protein by electrostatic and van der Waal's forces, resulting in a hydrophilic surface.
The current prospective double-blind randomized clinical trial aimed at comparing protein treated oxygenators versus heparin coated (Bioline CoatingTM, Jostra AG Hirrlingen, Germany) oxygenators and circuits, without changing further variables such as ACT management or cardiotomy blood processing in a low risk CABG patient population. The heparin used for the Bioline coatingTM process was LiqueminTM (Roche, Grenzach-Wyhlen, Germany).
| 2. Methods |
|---|
|
|
|---|
Very strict exclusion criteria were followed, to ensure that potential clinical and biochemical effects would be associated with CPB surface technology rather than comorbidity (Table 1).
|
Demographic data of the two patient groups are listed in Table 2.
|
Anesthesia was maintained with sufentanyl, pancuronium or propofol, ventilation was performed with a O2/N2O mixture establishing a FiO2 of 0.5 and aiming at a pCO2 of 3540 mmHg.
Chest drain blood loss during the first 24 h, application of catecholamines for inotropic support, as well as the need for packed red cells and fresh frozen plasma (FFP) were documented as clinical end points.
CK and CK-MB, troponin T and myoglobin were monitored postoperatively to assess myocardial protection.
The platelet count was determined, and to show impairment of the coagulation cascade or disseminated intravascular coagulation antithrombin III (AT III) and fibrinogen, D-dimers, and prothrombin fragments F1+2 (F1+2) were assessed. The plasmin-
2-antiplasmin complex (PAP) served to quantify hyperfibrinolysis.
To assess the inflammatory whole body response, interleukin 6 and 8 (IL-6, IL-8; ELISA, R&D Systems, Minneapolis, MN) were determined. IL-6 in CPB reflects pulmonary and cardiac dysfunction. Raised IL-6 levels have been reported to correlate with post-CPB left ventricular wall-motion abnormalities and myocardial ischemic episodes, whereas IL-8 is primarily associated with cardiovascular dysfunction. IL-8 is produced in response to a wide variety of proinflammatory stimuli such as exposure to IL-1, TNF, LPS and viruses. It is a potent chemoattractant for neutrophils. In addition, IL-8 also has a wide range of other proinflammatory effects. IL-8 causes degranulation of neutrophil specific granules and azurophilic granules, enhances the adherence of neutrophils to endothelial cells and sub-endothelial matrix proteins [13]. C5b9 was employed reflecting the final step of complement activation.
Human myeloperoxidase (MPO) is a hemoprotein with a molecular weight of 140 kDa. MPO is stored in primary granules (azurophilic) of neutrophils. It is a major component of the bactericidal armamentarium of neutrophils, due to its capacity to catalyze the production of hypoclorous acid (HOCl), a powerful oxidant that is derived from chloride ion (Cl-) and hydrogen peroxide (H2O2). In a number of inflammatory situations, MPO is released in the extracellular medium where its measurement can be used as an index of neutrophil activation, a major source of reactive oxygen intermediates leading to membrane injury. It was determined with an enzyme-linked immunoassay (MPO-ELISA; R&D Systems, Minneapolis, MN).
Arterial blood samples were collected before induction of anesthesia, when CPB was installed, immediately prior to and 5 min after release of the aortic crossclamp, 1 h, 24 h, and 5 days after surgery. Concentrations were not adjusted for the hemodilution caused by the circuit prime. This resulted in the potential masking of an increase in certain variables at the onset of CPB.
S-100B protein and NSE served to quantify cerebral injury biochemically. The S-100B protein was determined using a sensitive luminometric assay (Byk-Sangtec, Lund, Sweden), which selectively measures the beta subunits (present in glial and Schwann cells), as opposed to the alpha subunits (present in striated muscle, heart and kidney). We determined NSE in serum specimens with an ELISA test (Boehringer Mannheim Immundiagnostica, Mannheim, Germany), where 95% of healthy individuals show serum concentrations below 13 µg/l (detection limit 0.5 µg/l).
2.3. Statistical analysis
Statistical analysis was carried out using the SAS software package. The MannWhitney U-test was employed to compare differences between the two groups in the absence of a normal distribution, and ANOVA for repeated measures was employed to assess increases of variables over time. Separate ANOVAs were used for each group. Bonferroni corrections were used for post hoc comparisons. Data are presented as mean±standard error of mean. Differences were considered significant, when the P value was below 0.05.
| 3. Results |
|---|
|
|
|---|
|
|
|
|
|
|
| 4. Discussion |
|---|
|
|
|---|
Since the first efforts by Gott [14] to attach heparin to CPB surfaces, several heparin coating technologies have been developed, to achieve a possible reduction of anticoagulation for CPB [15]. A second rationale behind heparin coating of CPB circuits or oxygenators alone is to reduce the interaction between blood and foreign surfaces [16], which involves modified adsorption of plasma proteins to the surface. This creates an inactive surface with reduced attachment of fibrinogen resulting in lower adsorption and activation of platelets [17]. Furthermore, the continuous formation of a complex between ATIII and thrombin is catalyzed. A third rationale behind heparin coating is a reduction of the risk of oxygenator failure, which can be achieved with both surface treatments on the QuadroxTM oxygenator employed in both groups of the present study. The surface treatment, or coating, is an additional improvement of an oxygenator with a very small pressure gradient of 39 mmHg without, and 32 mmHg with BiolineTM coating [18].
The biochemical benefit of heparin coated CPB circuits has been shown by several researchers in animal models [19] and in clinical studies. An important difference between various clinical studies is the lower [5], or equal ACT level [12] for heparin coated versus uncoated CPB circuits. When using different ACT levels in study groups, one cannot distinguish between the effects of heparin coating alone, and those of reduced anticoagulation. Other authors could not confirm some of the postulated benefits of heparin coating, particularly that heparin attached to the blood contact surface may limit complement activation [20] or change beta thromboglobulin or fibrinopeptide A levels [21].
The treatment of artificial surfaces with synthetic proteins may be an alternative to heparin coating, because it is capable of avoiding the adsorption of blood cells on the oxygenator surface. The original idea behind using the synthetic protein treatment, was to prevent the high pressure drop phenomenon, a rise in pressure difference between the inlet and outlet of membrane oxygenators, probably due to a pathologic fibrin layer increasing the pressure gradient across the membrane.
The high standard of current CPB systems has made it increasingly difficult to test technical improvement in clinical studies involving relatively small patient groups. In most cases, the statistical power of such studies will not suffice to show a significant clinical benefit associated with changes in the CPB circuit. The complications that become clinically evident represent merely the tip of the iceberg, whereas the large ice mass under the water represents those laboratory findings that have not become clinically relevant. For the future, there is the choice to rely on laboratory outcome variables, which is the more unsatisfactory option, or to study larger patient groups to achieve the statistical power to reveal a clinical benefit. Another option pursued in the current protocol is a strict patient selection regarding concomitant disease and risk factors for CPB related complications.
The biochemical effects of heparin coating versus untreated surfaces have been shown previously [22] by comparing untreated oxygenators to otherwise identical heparin coated circuits and oxygenators. The coated circuits were associated with reduced leukocyte and complement activation, and less cerebral and/or blood brain barrier changes. In the relatively small number of routine CABG patients studied, a clinical benefit for heparin coating was not evident. Other authors have demonstrated that oxygenators and circuits heparin coated with various technologies, lead to biochemical, but not clinical improvement [23].
The study by Steinberg et al. [24] assessed interleukins 1, 2, 4 and 6, and found an increase of IL6 due to lung dysfunction. Increased production of interleukin 6 has been associated with post-bypass lung dysfunction and impaired hemostasis in additional studies, and IL8 levels correlated with post-bypass cardiovascular dysfunction [25]. We therefore chose those two cytokines for the current study, and observed a significant increase of both variables after CPB. IL8 was markedly lower with heparin coating, but did not correlate with post-bypass cardiovascular dysfunction. The different reaction pattern of these cytokines is open for speculation, because of the similar mechanism of their activation.
Leukocyte activation, as determined by myeloperoxidase release, showed a similar time course in both groups (Fig. 1), demonstrating the need for further control of activated leukocytes by prevention of activation and/or filtration of these cells. The amount of complement activation, as measured by membrane attack complex, differed significantly between groups after release of the aortic crossclamp (Fig. 3), indicating a possible effect of surface bound heparin on complement activation. The interaction between complement and leukocyte activation would suggest a reduced myeloperoxidase release in heparin coating patients, which was not observed. Therefore, other mechanisms of complement activation may be involved in the reduced release of C5b9 in group B.
Postoperative platelet count was reduced in all patients examined, showing a significant difference between groups 1 h postoperatively (P=0.029; Table 4), reflecting a platelet sparing effect due to decreased platelet adhesion to heparin coated surfaces. The degree of disseminated intravascular coagulation as measured by AT III, and fibrinogen was identical in both groups, which was also true for D-dimers, F1+2, and PAP, representing coagulation activation and hyperfibrinolysis associated with CPB. These domains are often considered among the variables affected by IV heparin. The lack of changes between groups suggests that surface bound heparin does not significantly alter the interference of IV heparin with coagulation and fibrinolysis. Based on the present data, the quality of extracorporeal surfaces in contact with blood is not the dominant factor determining the activation of the coagulation cascade.
S-100B protein has become a frequently used marker of cerebral injury and/or blood brain barrier dysfunction. Serum concentrations react rapidly not only to neuronal injury, but even transient and clinically irrelevant changes in blood brain barrier permeability lead to increased S-100B release [26]. Healthy individuals show serum concentrations below the detection limit of 0.2 µg/l. Drawbacks of this variable, such as the sensitivity to cardiotomy blood processing (unpublished data), need to be considered. The current double-blind protocol allowed us to eliminate this problem, so that at least the differences between groups supply reliable information. As opposed to previous studies comparing S-100B between heparin coated and untreated CPB circuits/oxygenators, the difference between protein treatment and heparin coating was not significant before, during and after surgery.
NSE is found predominantly in neurons and its release is a marker of outright neuronal damage, such as in TIA, stroke, or intracerebral hemorrhage. Concentrations correlate with the volume of the lesion [27]. It is a sensitive indicator of even small brain infarcts [28].
Our results imply that heparin, as a pharmacological agent, may not be the only source of improvement associated with heparin coating. We assume that the synthetic protein treatment contributes significantly to improved biochemical results, when compared to previous studies, where completely untreated surfaces served as controls [24].
We conclude, that both polypeptide coating and heparin coating are safe and effective surface treatments for CPB. Both comprise an improvement as compared to CPB circuits and oxygenators examined previously. However, the polypeptide coating approach is the current standard treatment of the QuadroxTM oxygenator, and is not associated with additional cost. SafelineTM is therefore appropriate for use in routine patients. In patients undergoing long-term extracorporeal support or ECMO, the advantages of heparin coating may become more evident.
| Footnotes |
|---|
| Appendix A |
|---|
|
|
|---|
Dr Wimmer-Greinecker: I don't have any data on the neutrophil numbers, no.
Dr L. von Segesser (Lausanne, Switzerland): We had heard in the postgraduate course on Sunday from the Regensburg group, that there was a significant reduction in the transoxygenator gradient for coated oxygenators. Can you confirm this?
Dr Wimmer-Greinecker: We haven't looked at this either. This was not the aim of this study.
Professor Taylor: Were you surprised to find no change in neuropsychological test results from preop to 5 days postoperatively?
Dr Wimmer-Greinecker: Actually we were not surprised, no. We have done other studies not only on heparin-coated systems, we also did studies on surface modification, and we also have done some studies on different oxygenators. In all those cases, we didn't find any significances between groups evaluating S100 beta protein or our testing, so we actually were not surprised about those findings.
Professor Taylor: So your experience with cognitive testing is that you haven't been able to pick up any change in any patient you studied undergoing open heart surgery?
Dr Wimmer-Greinecker: Among this study, there was one patient who had a transient neurologic damage. It was a paralysis of his left arm for 6 days. In this patient, we did have a correlation concerning the tests. This patient also showed an increase as well in S100 beta and in neuron-specific analysis.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F De Somer, A Van Landschoot, G Van Nooten, and J Delanghe Interaction of plasma proteins with commercial protein repellent polyvinyl chloride (PVC): a word of caution Perfusion, July 1, 2008; 23(4): 215 - 221. [Abstract] [PDF] |
||||
![]() |
J. W. Hammon Extracorporeal Circulation: Perfusion System Card. Surg. Adult, January 1, 2008; 3(2008): 350 - 370. [Full Text] |
||||
![]() |
O. Mangoush, S. Purkayastha, S. Haj-Yahia, J. Kinross, M. Hayward, F. Bartolozzi, A. Darzi, and T. Athanasiou Heparin-bonded circuits versus nonheparin-bonded circuits: an evaluation of their effect on clinical outcomes Eur. J. Cardiothorac. Surg., June 1, 2007; 31(6): 1058 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. Hogue Jr, C. A. Palin, and J. E. Arrowsmith Cardiopulmonary bypass management and neurologic outcomes: an evidence-based appraisal of current practices. Anesth. Analg., July 1, 2006; 103(1): 21 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G Raja and G. D Dreyfus Modulation of Systemic Inflammatory Response after Cardiac Surgery Asian Cardiovasc Thorac Ann, December 1, 2005; 13(4): 382 - 395. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. N Hoel, V. Videm, S. T Baksaas, T. E Mollnes, F. Brosstad, and J. L Svennevig Comparison of a Duraflo II-coated cardiopulmonary bypass circuit and a trillium-coated oxygenator during open-heart surgery Perfusion, May 1, 2004; 19(3): 177 - 184. [Abstract] [PDF] |
||||
![]() |
E. A. Hessel II and L. H. Edmunds Jr. Extracorporeal Circulation: Perfusion Systems Card. Surg. Adult, January 1, 2003; 2(2003): 317 - 338. [Full Text] |
||||
![]() |
G Matheis, M Scholz, A Simon, O. Dzemali, and A Moritz Leukocyte filtration in cardiac surgery: a review Perfusion, September 1, 2001; 16(5): 361 - 370. [Abstract] [PDF] |
||||
![]() |
H. Suhara, Y. Sawa, M. Nishimura, H. Oshiyama, K. Yokoyama, N. Saito, and H. Matsuda Efficacy of a new coating material, PMEA, for cardiopulmonary bypass circuits in a porcine model Ann. Thorac. Surg., May 1, 2001; 71(5): 1603 - 1608. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Matheis, M Scholz, A Simon, D Henrich, G Wimmer-Greinecker, and A Moritz Timing of leukocyte filtration during cardiopulmonary bypass Perfusion, January 1, 2001; 16(1_suppl): 31 - 37. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||