Eur J Cardiothorac Surg 2000;17:580-586
© 2000 Elsevier Science NL
a Department of Cardiothoracic Surgery, University of Gothenburg, Sahlgrenska University Hospital, SE 413 45 Gothenburg, Sweden
b Scandinavian Heart Center, Carlanderska Hospital, Gothenburg, Sweden
Corresponding author. Tel.: +46-31-342-1556; fax: +46-31-417-991
e-mail: ali.belboul{at}sahlgrenska.se
| Abstract |
|---|
|
|
|---|
Key Words: Biocompatibility Lipid peroxidation Cardiac enzyme Heart surgery Heparin coated circuits
| 1. Introduction |
|---|
|
|
|---|
| 2. Patients and methods |
|---|
|
|
|---|
Patients were randomized into two groups as described below.
2.1.1. CPB with heparin-coated membrane oxygenator (H)
Twenty patients (all men) were operated with the use of heparin-coated CPB circuits (Duraflo II, Bentley/Baxter Inc., Irvine, CA). The mean age was 65.8±6.6 years. Five patients were in NYHA-class II, 15 in NYHA-class III. The mean ejection fraction in the H group was 59.2±14.7%.
2.1.2. CPB with non-coated membrane oxygenator, control group (C)
Nineteen patients (16 males and three female) were operated with non-coated CPB circuits (Bentley/Baxter Inc., Irvine, CA). The mean age was 66.7±9.2 years. Three patients were in NYHA-class II and 16 patients were in NYHA-class III. The mean ejection fraction in the C group was 63.9±13.4%.
2.2. Anaesthesia
Premedication consisted of morphine 0.1 mg/kg i.m. and scopolamine 0.4 mg. Anesthesia was induced with thiopental 35 mg/kg body wt., followed by pancuronium 0.1 mg/kg. Fentanyl was given in incremental doses during induction of anesthesia and after intubation up to a total amount of 15 µg/kg. All operations were performed through a midline sternotomy using ECC with cannulation of the ascending aorta and right atrium.
2.3. Cardiopulmonary bypass (CPB)
A Gambro heart-lung machine (Lund, Sweden) with a roller pump was used. The CPB circuit consisted of a membrane oxygenator (BOS-CM50; Bentley/Baxter Inc., Irvine, CA) a venous reservoir (BMR-1900; Bentley/Baxter Inc.) and a cardiotomy reservoir (BCR-3500; Bentley/Baxter Inc.). The CPB circuit was primed with 2000 ml Ringerdex (Pharmacia, Uppsala, Sweden). A modified St. Thomas solution (Pharmacia, Uppsala, Sweden) was used to achieve cardioplegia and general hypothermia (3034°C) was instituted. Patients were given the standard dose of systemic heparin (300 IU/kg). The CPB was started when the activated coagulation time (ACT) was above 450 s. Additional heparin was given if the ACT was below 400 s. At the end of CPB, the effect of heparin was neutralized by protamine. Alpha-stat was used for blood gas management. The hematocrit was aimed for 1825%.
2.4. Measurement of complement activation
The classical complement activation (c-function) was determined by a hemolytic method [9]. Values were expressed as percentage of the activity of pooled normal serum (75125%). The C3, C4 and C3d fractions were determined by the immuno-electrophoresis technique [9] and values were expressed in g/l or mg/l. Measurements were made at the start of ECC, every 30 min during ECC and at the end of bypass.
2.5. Measurement of routine hematological parameters
Blood samples for total hemoglobin (HGB) white cell count (WCC), platelet count, plasma hemoglobin (P-Hb) and cell differentiation were taken preoperatively, 30 and 60 min of bypass and at the end of CPB, and 1, 3 and 5 days postoperatively.
2.6. Measurement of malondialdehyde (MDA)
Blood samples for assay of plasma MDA as a marker of the free radicals inducing lipid peroxidation (LPO) were taken preoperatively, 30 and 60 min of bypass and at the end of CPB. The blood samples were centrifuged and analysed using the thiobarbituric acid method [12]. LPO in plasma is expressed by MDA (nmol/ml).
2.7. Measurement of cardiac enzyme changes
Aspartate aminotransferase (ASAT) and creatine kinase myocardial-brain isoenzyme (CK-MB) were drawn from the patients preoperatively, and 6, 12, 24 and 72 h postoperatively. The reference values of ASAT and CK-MB in our hospital are <0.8 µkat/l and <5 µg/l, respectively. Myocardial infarction was empirically considered when CK-MB level rose over 100 µg/l and ASAT over 2.5 µkat/l with or without ECG changes. Smaller rises in the enzymes were considered as markers of occult damage due to ischemia and reperfusion.
2.8. Postoperative morbidity and complications
The frequencies of postoperative morbidity (postoperative bleeding, respirator usage, blood transfusion, rhythm disturbance, cerebrovascular accidents, and infection) were registered and compared between H and C groups.
2.9. Statistical analyses
All means are expressed with one standard deviation (SD). Comparisons between two groups were performed using the Student's unpaired t-test or the MannWhitney test. Serial changes within each group were statistically evaluated by the analysis of variance for repeated measurements. A P-value of less then 0.05 was considered to be significant. Correlations were performed to bring out the association between variables tested.
| 3. Results |
|---|
|
|
|---|
|
|
3.2. White blood cells
There was an increase in circulating white blood cells during the operation with a maximum at the end of bypass in both H and C groups (P<0.0001). When comparing the two groups, a significantly higher leucocytosis was seen in the C group (P<0.0001) (Table 3). The increase in white blood cells was caused by an increase in circulating neutrophils, which showed a significant difference between the groups being higher in the C group, (P<0.0001) (Table 3).
|
|
3.4. Changes in cardiac enzymes
The mean CK-MB value was significantly raised in both groups from a preoperative value of less then 5 µg/l to 25±14 in the H group and 36±16 in C group 6 h postoperatively (P<0.001). When comparing the two groups, there were significant difference starting at 6 h postoperatively (Table 5).
|
3.5. Plasma hemoglobin (P-Hb)
The mean P-Hb increased significantly with the duration of CPB in both groups (P<0.0001). The P-Hb in the H group was significantly less elevated when compared to the C group, (P<0.02) (Table 6).
|
|
|
|
| 4. Discussion |
|---|
|
|
|---|
Both groups had similar preoperative and intraoperative characteristics and this could explain their similar clinical postoperative outcome with respect to the duration of ventilation, blood loss, transfusion requirements, myocardial infarction and arrhythmic events.
In a previous publication [8], we had shown that a patient group operated with the heparinized equipment spent a shorter time postoperatively on the ventilator and that this could be related to a reduction in lung damage, resulting in a better condition and oxygenation of the patients. Furthermore, it was shown that the heparin-coated circuits had a significantly lower increase in neutrophil levels, as seen in the present study, but the finding was not emphasized in the pulmonary report.
In this study we also demonstrated increasing levels of leucocytes and MDA during CPB which could support the hypothesis of oxygen free radical damage on the jeopardized myocardium. The main origin of oxygen free radicals during CPB has been credited to activated neutrophils and this is borne out in this study by the strong correlation of MDA and neutrophil generation. Furthermore, the relationship of MDA to CK-MB suggests very strongly that CPB induced white cell damage and free radical generation can damage the myocardium following reperfusion after aortic declamping.
Several mechanisms could contribute to ischaemia and reperfusion injury: The physical occlusion of microvessels by activated leukocytes which occur under conditions of reduced perfusion pressure where the capillary may behave like a sieve. This trapping phenomenon during ischaemia and reperfusion may become even more pronounced as the release of adhesions molecules lead to increased leucocyte-endothelial interactions [13]. The release of active mediators by activated leucocytes, and granular enzymes, reactive oxygen metabolites and products of membrane phospholipases [14] could contribute to the damage of the myocardium. ECC stimulated neutrophils would also transiently aggregate and adhere to vascular endothelial cells and to myocardial cells after infiltration across the endothelial barrier into the reperfused myocardium [15]. In the presence of the activated complement fragments, activated neutrophils would more easily release their granule contents and generate oxygen free radicals, enhancing the reperfusion injury [14]. These reactive oxygen mediates have been shown experimentally to induce arrhythmias [16] which could be inhibited by free radical scavengers such as allopurinol [17].
The lipid peroxidation pattern suggests a clue to the time when it is mostly generated. When comparing the groups, the first 30 min gives an average rate of a 60% higher production with uncoated circuits, and following this, the generation appears to be stable in both groups with a 5% higher generation in the controls. This finally rises to a 30% higher production. in the controls on reperfusion and rewarming following aortic declamping. This suggests two peak periods of relatively intense free radical generation. The main leucocyte generation appears in the first 60 min and is higher in the controls and then falls at the end of CPB. This suggests that leucocytosis tends to stabilize faster in the heparin-coated group. The reason for this is not known but it can be speculated that the signal mediators of the leucocytosis response are probably made less active in the heparin-coated circuits. Since complement generation was similar in both groups, it is probably the lower production of free radicals and probably other signal substances that modify the intensity of leucocytosis. The cytokines and leukotrines were not monitored in this study. The question arises if heparinized circuits can adsorb such mediators and this needs to be addressed in future studies.
Most studies show that cardiac enzymes do not tend to leak during cardioplegic protection but rather after declamping the aorta. Furthermore, enzyme leakage during CABG without ECC is strikingly absent in the majority of cases suggesting that the long ischemic periods due to global ischemia is a major determinant for reperfusion injury during ECC and aortic clamping. However, the intensity of enzyme leakage was well tolerated in this study in both groups suggesting that cardioplegic protection was adequate during the period of cross clamping not leading to significant complications. It is still evident that cardioplegia could not prevent occult myocardial damage with some enzyme loss.
Breda et al. found that reperfusion of isolated ischemic hearts with leukocyte-depleted blood prevented ultrastructural damage and resulted in better myocardial function than when normal blood was used [18]. The sequestration of leukocyte-platelet aggregates in the ischemic heart after declamping [19] has been demonstrated during CPB. The leukocytes are thought to be attracted by free radical generation in the transient ischemic heart after cardiopulmonary bypass giving the occult ischemic damage which may play a role in the pathogenesis of postoperative morbidity after open heart surgery. Lazar et al. showed that the use of heparin-bonded circuits in animals, significantly decreased myocardial ischaemic damage during acute surgical revascularization [20].
In addition, heparin-coated ECC circuits, by reducing neutrophil stimulation, and subsequently free radical release (expressed as increased MDA in this study), was found to reduce occult myocardial damage as reflected by the reduction of myocardial enzyme as seen in this study. This study verifies that the use of the CPB technique is still a potential hazard which can, however, be modified to the patients advantage by using heparin-coated circuits.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
O. Mangoush, S. Purkayastha, S. Haj-Yahia, J. Kinross, M. Hayward, F. Bartolozzi, A. Darzi, and T. Athanasiou Heparin-bonded circuits versus nonheparin-bonded circuits: an evaluation of their effect on clinical outcomes Eur. J. Cardiothorac. Surg., June 1, 2007; 31(6): 1058 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Jessen Pro: Heparin-Coated Circuits Should be Used for Cardiopulmonary Bypass Anesth. Analg., December 1, 2006; 103(6): 1365 - 1369. [Full Text] [PDF] |
||||
![]() |
R. Taneja and D. C. Cheng Con: Heparin-Bonded Cardiopulmonary Bypass Circuits Should Be Routine for All Cardiac Surgical Procedures Anesth. Analg., December 1, 2006; 103(6): 1370 - 1372. [Full Text] [PDF] |
||||
![]() |
S. G Raja and G. D Dreyfus Modulation of Systemic Inflammatory Response after Cardiac Surgery Asian Cardiovasc Thorac Ann, December 1, 2005; 13(4): 382 - 395. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Ngaage Off-pump coronary artery bypass grafting: the myth, the logic and the science Eur. J. Cardiothorac. Surg., October 1, 2003; 24(4): 557 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Johnell, G. Elgue, R. Larsson, A. Larsson, S. Thelin, and A. Siegbahn Coagulation, fibrinolysis, and cell activation in patients and shed mediastinal blood during coronary artery bypass grafting with a new heparin-coated surface J. Thorac. Cardiovasc. Surg., August 1, 2002; 124(2): 321 - 332. [Abstract] [Full Text] [PDF] |
||||
![]() |
L.-C. Hsu Heparin-coated cardiopulmonary bypass circuits: current status Perfusion, September 1, 2001; 16(5): 417 - 428. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||