| Abstract |
|---|
|
|
|---|
Key Words: Roller pump Centrifugal pump Mortality Neurological outcome
| 1. Introduction |
|---|
|
|
|---|
Previous studies have assessed the effect of CPB on neurological outcomes in adult patients undergoing cardiac surgery[310]; however, little information exists on whether the type of the pump used for CPB (roller vs. centrifugal) can affect the neurological outcomes of adult patients undergoing cardiac surgery.
The aim of this study was to evaluate risk factors for perioperative mortality and adverse neurological outcomes, with special emphasis on the role of using a roller or centrifugal pump for perfusion.
| 2. Patients and methods |
|---|
|
|
|---|
2.1. Surgical procedure
On the morning of surgery, patients received their usual dose of antianginal drugs, and 25 mg morphine and 1 mg atropine as premedication. All patients received a standard moderate dose of fentanyl and benzodiazepine anesthesia, which was induced by the administration of sodium thiopental (3 mg/kg), fentanyl (0.75 µg/kg), succinylcholine (1 mg/kg), diazepam (10 mg) and pancuronium bromide (0.1 mg/kg). After endotracheal intubation, patients were ventilated to normocapnia with an oxygen and air mixture. Boluses of fentanyl (with or without droperidol), diazepam and pancuronium bromide were given when necessary. Cefuroxime (2 g) was given intravenously for infection prophylaxis. A radial artery catheter and a flow-directed pulmonary artery catheter were inserted for hemodynamic measurements. The extracorporeal circuit consisted of a roller pump (CAPS HLM; Stockert Instruments, Munich, Germany) or a centrifugal pump (Biomedicus BioPump; Medtronic, Milan, Italy), hollow fiber oxygenator with integrated heat exchanger, arterial filter, cardiotomy reservoir and polyvinyl tubing system. In all cases, an open system was used for perfusion.
The type of pump was chosen based on a consensus between the surgeon and anesthesiologist, with preference given to the centrifugal pump for CPB times which were expected to be longer (>100 min).
No heparin bonding was used in the perfusion tubing or oxygenator. The extracorporeal circuit was primed with 1500 ml electrolyte solution and 5000 IU bovine lung heparin (Liquemin; Roche, Italy).
After systemic heparinization (300 U/kg), CPB was initiated with cannulas placed in the ascending aorta and right atrium. The activated clotting time was kept at
400 s with additional heparin.
The pump flow was non-pulsatile in all operations. The flow rate was maintained at 2.4 l/min per m2 during cooling and rewarming phases, and at 2.0 l/min per m2 during stable hypothermia. The mean arterial pressure (MAP) during CPB was maintained between 60 and 90 mmHg, with CPB flows set as previously described, and vasoactive drugs were used to maintain the MAP in the desired range: if the MAP increased above the desired range, and was unresponsive to fentanyl or diazepam, sodium nitroporusside was started, but if the MAP fell below the desired range, norepinephrine, boluses or continuous infusion were added. Patients were cooled to 2830°C. CPB flows and pressures were downloaded from the monitor and recorded every 5 min during perfusion. Any significant modification of the perfusion pattern during 5 min time intervals was also recorded by the perfusionist.
For myocardial protection, patients received a first dose (1000 cc) of cool (4°C) antegrade and retrograde high-potassium cold crystalloid cardioplegia (St. Thomas Hospital Cardioplegic solution containing 110 mmol/l NaCl, 16 mmol/l KCl, 16 mmol/l MgCl2, 1.2 mmol/l CaCl2 and 10 mmol/l NaHCO3) just after aortic cross-clamping, which was repeated (250 ml retrograde) at every 20 min of the aortic cross-clamp time. Disturbances in the acidbase balance were appropriately treated, and the acidbase equilibrium was maintained by the alpha-STAT method. The hematocrit during CPB was maintained at 1825%.
After termination of CPB, heparin was antagonized with protamine sulfate at a 1:1 ratio (3 mg/kg). If necessary, inotropic support was given when patients were weaned from CPB. Autologous blood and residual volume from the extracorporeal circuit were infused into the patient when volume supplementation was necessary.
After surgery, patients were admitted to the intensive care unit (ICU) and treated according to a standardized protocol. The MAP was kept at 7090 mmHg, heart rate at 7090 beats/min, and the cardiac index was maintained at greater than 2.0 l/min per m2. Inotropic support was administered when necessary. Patients were ventilated to normocapnia, and an arterial oxygen tension of 80 mmHg with continuous positive-pressure ventilation until extubation was maintained according to the ICU regimen. Basic fluid administration consisted of 0.9% NaCl and polygelatine. Packed erythrocytes were infused when the hematocrit was <18% during CPB and <24% in the ICU. When their cardio-respiratory condition had stabilized, patients were transported to the ward for further recovery.
2.2. Statistical analysis
The data are presented as medians (25 and 75% percentiles in brackets) for continuous variables or percentages for categorical variables. A commercial statistical package (SPSS for Windows Version 8.0; SPSS, Inc., Chicago, IL) was used for data analysis.
Thirty-seven preoperative and operative variables, including the type of pump employed for CPB (roller vs. centrifugal; Tables 1 and 2) were assessed for their possible effect on the occurrence of the following outcomes, which were defined as follows:
|
|
All continuous variables were first tested individually (univariate sense) with the non-parametric MannWhitney test, while categorical variables were explored by the Chi-square (Yates continuity correction) or the Fisher's exact test when indicated.
The factors which were at least marginally significant (P
0.2) by univariate analysis were included into a multivariable forward stepwise logistic regression model. The multivariate odds ratio (OR) for each independent variable in the final regression models and 95% confidence intervals were also computed. The P value for entry of a covariate into the model was set at a significance level 0.05, while the P value for the removal of a covariate was fixed at the 0.1 significance level. Every multivariable model was tested for reliability with the HosmerLemeshow statistic [11].
| 3. Results |
|---|
|
|
|---|
The clinical variables of patients who underwent surgery using centrifugal or roller pumps are reported in Tables 3 and 4. Regarding preoperative variables, the two groups of patients were similar in age, body surface area, gender, history of hypertension, preoperative pulmonary, and renal status; on the other hand, patients who underwent surgery with the use of a centrifugal pump had a higher incidence of previous myocardial infarction, insulin-dependent diabetes mellitus, were more frequently redos, and had a higher incidence of previous transitory ischemic attack (TIAs), strokes, and overall neurological events. The analysis of the operative variables showed similar frequencies in the different types of surgery, as well as in the need to perform a circulatory arrest, while aortic cross-clamp and CPB times were, as expected, longer in cases using a centrifugal pump. Finally, there were statistically significant differences in perfusion pressures, flows and temperatures between roller and centrifugal pumps: pump flows were slightly lower in cases of centrifugal pump use, while pressures and temperatures were slightly higher; however, in no case did the differences between the two groups in CPB-related variables exceed a 10% variation, and this was reported as less than 5% in most of the comparisons.
|
|
|
|
|
|
|
|
|
| 4. Discussion |
|---|
|
|
|---|
To date, several studies assessed the effect of CPB-related variables on neurological outcomes in adult patients undergoing cardiac surgery; in particular, it could be demonstrated that longer CPB times [4,5] and severity of ascending aorta atherosclerosis [4,8] were strong predictors of perioperative neurological complications, while lack of arterial line filtration [9,12] and pH-stat acidbase management [10] could also increase the occurrence of postoperative central nervous system dysfunction. On the other hand, the possible effects of other CPB variables are still being debated, and the roles of temperature management (normothermic vs. moderately hypothermic perfusion), mean arterial pressure levels during CPB and type of perfusion (pulsatile vs. non-pulsatile) in affecting neurological complication rates have not yet been completely defined [13].
Centrifugal pumps have been widely used as the main pump in adult cardiac surgery, and are considered by some authors to be superior to the traditionally used roller pumps because of less blood trauma [14], reduced activation of the coagulation cascade [15] and improved biocompatibility [16], even if some of the studies could not document any significant benefits in terms of hemolysis [17], platelet damage [18] or immune response [19]; in addition, some recent evidence could document an increased inflammatory response to CPB in cases using centrifugal pumps [20,21].
There is less information, however, on the effect of the use of this kind of pump on clinical endpoints, and the effects on neurological function are not well established yet; even if centrifugal pumps have been shown to generate fewer microemboli than roller pumps [22], it was recently demonstrated that, for CPB times of less than 90 min, centrifugal pumps did not decrease serum S100ß release, a marker for cerebral injury, compared with roller pumps [23]. In addition, only one paper, by Klein and colleagues, has previously evaluated the effect of centrifugal and roller pumps used for CPB on many different clinical outcomes, documenting a reduced rate of neurological complications when a centrifugal pump was used for perfusion; but no additional information was given about the criteria used to define neurological complications in this paper [24].
The aim of our study was then to retrospectively review the data concerning adult patients who underwent coronary and/or valve surgery at our hospital during a 5.5-year period (1994June 1999); during that period, in fact, both types of pump (roller and centrifugal) were used at our hospital at the same time by the same team of surgeons, anesthesiologists and perfusionists. Our study could show that factors related both to the preoperative clinical status of the patients and to operative and CPB features can affect the occurrence of major neurological complications; age at intervention, previous vascular surgery, previous neurological episodes, as well as longer CPB times were risk factors for adverse neurological outcomes of adult cardiac surgery patients, as previously described [13].
In addition, our study documented that the use of a centrifugal pump can reduce the rates of the two most feared neurological complications of routine adult cardiac surgery performed with the use of CPB; univariate and multivariate analyses documented its protective effect, for the occurrence of both perioperative permanent neurological deficit and perioperative coma, reducing the risk reduction for the considered events by approximately half (multivariable ORs of 0.57 and 0.46 for perioperative permanent neurological deficit and perioperative coma, respectively).
Interestingly, the protective effect of centrifugal pump use could be documented even if there was a clear selection bias between the two pump types in our patient population, and centrifugal pumps were preferred, as previously stated, for cases with longer perfusion times. Also, the subgroup analysis comparing preoperative and operative variables by the type of pump could confirm that patients operated with centrifugal pumps, aside from the expected longer aortic cross-clamp and CPB times, also had an increased risk profile for the considered events, being more frequently redos, insulin-dependent diabetics, and having a higher rate of pre-existing neurological events; in fact, all of these three factors were previously documented as risk factors for unfavorable neurological outcome [2].
The hypothesis to explain the protective effect of the centrifugal pumps used for CPB is, that with centrifugal pumps, the embolic load to the brain is lower [22], as was previously demonstrated with ultrasonic microbubble detection in the arterial CPB line [25].
On the other hand, the potential neurological benefit of the use of a centrifugal pump for perfusion was not paralleled by a decrease in in-hospital mortality; this finding has no clear explanation to us, and further investigations will be needed to clarify this point.
In conclusion, this study, with the limits of a retrospective, non-randomized study, suggests that centrifugal pump use during routine adult cardiac surgery reduced perioperative permanent neurological deficit and coma rates. Prospective, multi-institutional, randomized studies will be needed to better define the possible protective effects of centrifugal pumps on neurological outcomes of adult cardiac surgery.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. S. Just, T. Muller, M. Hartrumpf, and J. M. Albes First experience with closed circuit/centrifugal pump extracorporeal circulation: cellular trauma, coagulatory, and inflammatory response Interactive CardioVascular and Thoracic Surgery, October 1, 2006; 5(5): 646 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Skrabal, Y. H. Choi, A. Kaminski, M. Steiner, G. Kundt, G. Steinhoff, and A. Liebold Circulating endothelial cells demonstrate an attenuation of endothelial damage by minimizing the extracorporeal circulation. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 291 - 296. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Asante-Siaw, J. Tyrrell, A. Hoschtitzky, and J. Dunning Does the use of a centrifugal pump offer any additional benefit for patients having open heart surgery? Interactive CardioVascular and Thoracic Surgery, April 1, 2006; 5(2): 128 - 134. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |