Eur J Cardiothorac Surg 2001;19:302-306
© 2001 Elsevier Science NL
Service de Chirurgie Thoracique et Cardiovasculaire, Groupe Hospitalier Pitié-Salpêtrière, 47-83 Bd de l'Hôpital, 75651 Paris Cedex 13, France
Received 30 November 2000; received in revised form 15 January 2001; accepted 16 January 2001.
Corresponding author. Tel.: +33-1-4217-7005; fax: +33-1-4217-7030
e-mail: pascal.leprince{at}psl.ap-hop-paris.fr
| Abstract |
|---|
|
|
|---|
Key Words: Left ventricular assist device Total artificial heart Expanded polytetrafluoroethylene Pericardial substitute Adhesions
| 1. Introduction |
|---|
|
|
|---|
In patients in whom a circulatory support device is used for only a short time, problematic pericardial adhesions generally do not develop. However, with the increase in the time to transplantation to months after implantation of a support device, device-related adhesions may represent a considerable challenge on resternotomy, increasing the risk of tissue injury and damage to the mechanical device tubes. In addition, adhesions extend the time required for dissection of the native heart, thereby prolonging the risk of prolonged ischemia of the donor organ, and they also increase the risk of bleeding after transplantation [2]. Moreover, in patients in whom a total artificial heart (TAH; Cardiowest Technologies, Inc, Tucson, AZ) has been implanted for 3 months or longer, we have observed considerable pericardial thickening, which can lead to cardiac allograft restriction requiring pericardectomy.
In 1997, in an attempt to limit adhesions in patients undergoing bridge to cardiac transplantation, we began to use expanded polytetrafluoroethylene (ePTFE) pericardial substitutes (Preclude Pericardial Membrane, W.L. Gore & Associates, Inc, Flagstaff, AZ) to wrap surfaces of circulatory support devices at implantation. We here report our 3-year experience with this technique in patients given a TAH or left ventricular assist device (LVAD; Novacor, Baxter Healthcare Corp, Oakland, CA).
| 2. Patients and methods |
|---|
|
|
|---|
When a potentially suitable donor heart became available, the donors hemodynamic status and the quality of the graft were assessed by the harvesting team. If the findings were satisfactory, the heart recipient was placed under anesthesia as soon as possible and sternotomy for transplantation was performed. In some cases, the ePTFE membranes removed during this procedure were preserved for histologic study.
2.2. ePTFE wrapping technique in patients receiving a TAH
After homeostasis is achieved and placement of the ventricles tested during temporary chest closure, the sternum is retracted again to wrap the TAH with an ePTFE membrane (Fig. 1) Three interrupted monofilament 4/0 sutures are passed into the pericardium into the left side of the chest-one close to the left inferior pulmonary vein, one close to the diaphragm, and the third between the other two. The sutures are placed as deep as possible, posteriorly to the phrenic nerve. Sutures are then passed through the edge of an ePTFE membrane (a sheet of 15x20 cm that is 0.1 mm thick). The membrane is slid down the sutures, between the artificial ventricles and pericardium, and secured. The same technique is used to place a second ePTFE membrane between the pericardium and right atrium on the right side of the chest. The free edges of the two membranes are then overlapped anteriorly along the middle line. The inferior edge of the membrane on the left side is secured with two interrupted sutures on the anterior edge of the diaphragm, over the drive line for the TAH. The superior edge of the membrane on the right side is passed over the aorta and pulmonary artery and secured on the left side. In some cases, a few other sutures are added to avoid creases and folds in the ePTFE membranes, but no attempt is made to produce a watertight seal around the heart.
|
At chest closure in all patients in whom ePTFE membranes are used to wrap a TAH, a pericardial drainage tube is inserted into the membrane-limited cavity. Another drainage tube is placed anteriorly to the membranes, posteriorly to the sternum.
2.3. ePTFE wrapping technique in patients receiving an LVAD
Because the LVAD implantation procedure involves leaving the native heart in place and the strongest adhesions develop between the mechanical device and the pericardium, only the inflow and outflow tubes of the LVAD are wrapped with ePTFE membranes. A full sheet of ePTFE is wrapped around the entire length of the outflow tube, and a smaller piece is wrapped around the inflow tube. The ePTFE membranes are either secured with interrupted monofilament sutures or left free.
| 3. Results |
|---|
|
|
|---|
|
Another of the 23 patients in whom ePTFE membranes were used underwent reoperation after 51 days of circulatory support with a TAH because of increasing pericardial effusion. Computerized tomography showed clots between the TAH and an ePTFE membrane and compression of the right atrium. Reoperation was performed through a subxiphoid incision. After the plane of dissection between the diaphragm and the ePTFE membrane was established, the membrane was opened easily and the clots were removed by suction. No other adverse events occurred, and the patient underwent transplantation 15 days later.
Six explanted ePTFE membranes were studied histologically. The surfaces of the membranes were covered with loose accumulations of fibrous connective tissue that did not infiltrate the interstices of the membranes. Loosely adherent fragments of collagen were occasionally observed in focal regions along one surface (Fig. 3). The interstices of the membranes were multifocally infiltrated by amorphous proteinaceous material and proteinaceous fluid. There was no evidence of bacterial infection or calcification.
|
| 4. Discussion |
|---|
|
|
|---|
Membrane technology allows control of adhesions between tissues and artificial surfaces by modulation of the size of a membranes interstices. In our patients, we used ePTFE membranes with interstices of less than 1 µm. This pore size minimizes adhesions because it is too small to allow tissue penetration. The explanted ePTFE membranes that were examined histologically showed loosely adherent cells at the tissue-membrane interface and no evidence of collagen or cellular infiltration.
The most commonly used pericardial substitutes are made of either preserved animal pericardium (chiefly bovine) or a synthetic membrane (e.g. polyurethane, ePTFE, rubber, and polyester). In a study in ewes that compared bovine pericardium, ePTFE, and two types of polyglycolic acid mesh (Dexon), Bunton et al. [3] showed that after 6 months, marked adhesions to the bovine pericardium prosthesis and one type of polyglycolic acid mesh developed. Results with the other type of polyglycolic acid mesh were inconsistent. In addition, one of the bovine pericardium substitutes was found to have calcified by 6 months after implantation. In a multicenter survey, Heydorn et al. [4] reported comments from 89 surgeons on 236 reoperations in patients given a variety of pericardial substitutes, including devices made of silicone rubber or polyester, three different bovine pericardium products, and ePTFE. There was a high level of dissatisfaction with the substitutes, especially the silicone rubber devices and the xenografts, because of dense adhesions, intense pericardial reactions, pericarditis-related perioperative fever, graft rejection, and infection. The level of satisfaction with ePTFE (86%) was significantly higher (P=0.0004 by chi-square analysis) than that with the other substitutes (38 to 50%).
The usefulness of ePTFE pericardial substitutes in providing a clear plane of dissection and minimizing adhesions has also been documented in both experimental investigations [5] and clinical studies in patients in whom additional cardiac surgery may be necessary, such as those with congential heart defects [610] and those undergoing coronary artery bypass grafting [7,11,12]. Moreover, Vitali et al. [13], like us, had no complications in 20 patients in whom ePTFE membranes were used to wrap the outflow cannula of an LVAD. Vitali et al. also observed, as we did, no problematic adhesions between the membranes, chest wall, extracardiac conduit, and underlying epicardium at resternotomy. Their study did not include patients in whom a TAH was used or a histologic assessment of explanted ePTFE membranes.
Holman et al. [14] used ePTFE membranes to reconstruct the pericardium, but not to wrap the ventricular support device, in seven patients. They found that the membranes protected grafts to the great vessels and the right ventricle from injury on resternotomy and that the plane of dissection between the heart and the pericardium was easy to discern. Infection occurred in two patients, but one developed only after a second operation and the other, like the one in our series, was managed successfully with antibiotic therapy. Indeed, implantation of a circulatory support device is known to increase the risk of infection [15,16], regardless of whether ePTFE membranes are used. Fortunately, such infections generally do not preclude heart transplantation [16]. We do not believe that the use of ePTFE membranes increased the risk of infection in our patients.
One patient in our series had to undergo reoperation because of a pericardial effusion. It is known that patients in whom a circulatory support device is implanted have an increased risk of pericardial infusion. Even in patients given a TAH, this complication can lead to atrial compression and cardiac tamponade. We did not observe an increase in the pericardial effusion rate among our patients in whom ePTFE membranes were used to wrap a support device. Furthermore, since the new cavity created with such membranes is not watertight, membrane use does not increase the risk of compression.
In summary, we found that use of ePTFE membranes in patients undergoing bridge to transplantation with either an LVAD or a TAH prevented adhesions between tissues and device surfaces without increasing the risk of infection. As a result, reoperation was easier and safer. We initially wrapped only the surfaces of TAHs with ePTFE membranes, but because of the good results achieved with that procedure, we now wrap the native heart in all patients given an LVAD.
| References |
|---|
|
|
|---|
ur artificiel total Cardiowest: expèrience de 29 cas. Arch Mal Coeur Vaiss 1998;91:397-404.[Medline]
This article has been cited by other articles:
![]() |
M. E.W. Kirsch, Z. Kostantinos, F. Ali, E. Vermes, G. Bajan, and D. Y. Loisance Kinetic assisted venous drainage for orthotopic heart transplantation in patients under mechanical circulatory support: a double-edged sword Eur. J. Cardiothorac. Surg., March 1, 2008; 33(3): 418 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Leprince, A. Combes, N. Bonnet, A. Ouattara, C. E. Luyt, P. Theodore, P. Leger, and A. Pavie Circulatory support for fulminant myocarditis: consideration for implantation, weaning and explantation Eur. J. Cardiothorac. Surg., September 1, 2003; 24(3): 399 - 403. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||