Eur J Cardiothorac Surg 2001;19:678-683
© 2001 Elsevier Science NL
a Oxford Heart Centre, John Radcliffe Hospital, Headley Way, Headington, Oxford OX3 9DU, UK
b Terumo R&D Centre, Kanagawa, Japan
Received 10 October 2000; received in revised form 2 February 2001; accepted 23 February 2001.
Corresponding author. Tel.: +44-1865-220269; fax: +44-1865-220268
e-mail: swestaby{at}ahf.org.uk
| Abstract |
|---|
|
|
|---|
Key Words: Left ventricular assist device Centrifugal pump Sheep
| 1. Introduction |
|---|
|
|
|---|
Since 1995 the research group of the Terumo Corporation, Japan have sought to overcome the drive shaft and bearing problem by developing a magnetically suspended centrifugal pump (MSCP), the impeller of which is suspended by a magnetic field within the housing [3]. Because the MSCP provides contact free rotation of the impeller without material wear, it is expected to be one of the most durable blood pumps. Early prototypes of MSCP have already operated for more than 2 years in an ex vivo animal model [4,5]. Early experience with intrathoracic implantation of a second generation device (Model 2 MSCP) was terminated because of blood leakage through an intrahousing connector. However, the compact device fitted easily into the thorax of a 45-kg sheep and operated without thrombus formation or thromboembolism. After introduction of a dual connector system and sensorless motor the modified model 2 (Model III) was able to operate in vivo for more than 14 months without any sign of mechanical failure or thromboembolism [4,5]. The MSCP now has significant potential for long-term use as an implantable circulatory assist system.
| 2. Materials and methods |
|---|
|
|
|---|
|
Blood samples were collected serially to determine haematological and biochemical indices, prothrombin times, and markers of haemolysis. The surgical procedures and post-operative care were undertaken humanely by licensed personnel in compliance with United Kingdom Home Office guidelines.
2.3. Operation
The sheep were anaesthetized with thiopentone then intubated and ventilated with halothane in oxygen. A thermodilution pulmonary arterial catheter and arterial cannula were introduced into the left internal jugular vein and the left common carotid artery, respectively.
The animals were positioned for left thoracotomy. The sixth intercostal space was entered to provide access to both the apex of the left ventricle and the descending thoracic aorta.
The Dacron velour-covered power cable was first tunnelled under the scapula towards the midline to exit the skin of the left flank. The power line was then connected externally to the control system and power supply. Attention was then turned to the descending thoracic aorta which was partially mobilized for the outflow graft anastomosis. An aortic side clamp was applied and a 1.5-cm linear incision made in the vessel wall. To this the pre-clotted Dacron outflow graft of the device was anastomosed with 3-0 polypropylene.
The pericardium was then opened to gain access to the apex of the left ventricle. The apex was elevated and circumscribed with interrupted Teflon-pledgetted mattress sutures of 2-0 Tycron. These were used to secure the inflow cuff of the device. A cruciate incision was made into the apical muscle and the inflow cannula of the TILVAS (Fig. 2) easily inserted without blood loss. This was secured with a tape around the apical cuff to align the inflow cannula axially within the left ventricular cavity. The system was carefully de-aired before releasing the clamp on the outflow graft and switching the device on at a speed between 1000 and 2500 rev./min. With the pump in situ we implanted an ultrasonic flow probe around the outflow graft to determine pump flow at different rotational speeds. With normalized circulating blood volume and central venous pressure we determined the pump speed (18001900 rev./min) at which the aortic valve remained closed through capture of all transmitral flow. At greater speeds pulse pressure disappeared from the systemic circulation. An intercostal drain was inserted and the wound then closed in layers. Anaesthesia was discontinued and the sheep transferred from the operating room to the pen.
|
Neurological status was determined after recovery from anaesthetic by observing behaviour, mobility and balance. Renal and hepatic function were assessed serially by measurement of blood urea, creatinine, bilirubin and liver enzymes. After removal of the indwelling lines, blood samples were obtained by direct puncture of the internal jugular vein. Pump speed and power requirements were recorded twice daily. Auscultation was used to check the tone of the device. Neither heparin nor warfarin were used for anticoagulation after the surgical procedure.
2.5. Studies of safety, efficacy and long-term reliability
Data from device including motor suspension current, energy requirement and calculated pump flow were monitored constantly by on-line computer. Mechanical reliability was determined by recording parameters of pump function, by auscultation of the chest, and by inspection of power lines and driveline exit site. The sheep were examined for signs of heart failure, thromboembolism or abnormal behavior.
2.6. Device retrieval and autopsy studies
Four sheep were electively euthanized at 90 days after implantation. One was electively euthanized at day 210 and another at 360 days.
Prior to sacrifice at 3 months the sheep were given heparin (5000 IU/kg) intravenously to avoid clot formation on the blood-contacting surfaces of the pump and cannulae. After elective sacrifice the pump and vascular graft were removed by careful dissection. The interior of the native heart, vascular graft and aortic anastomosis were examined for thrombus formation. The pump was then perfusion-washed with saline by pumping at 2000 rev./min followed by fixation with 2% glutaraldehyde solution. The pump and computer were then returned to the Terumo Corporation for disassembly and inspection for mechanical wear, thrombus formation and component reliability.
Detailed post mortem examination was performed with macroscopic and histologic examination of the sheep aorta, carotid and renal arteries, together with kidney, liver, lungs and brain.
2.7. Statistical analysis
All results for continuous variables are expressed as mean±standard deviation. The Student's paired or unpaired t-test, or MannWhitney test if appropriate, were used to compare continuous variables between two subgroups. P-values of less than 0.05 were considered to indicate statistical significance.
| 3. Results |
|---|
|
|
|---|
3.2. Heamodynamic studies
Haemodynamic studies before chest closure showed the relationship between pump speed (rev./min) and cardiac output. Systemic blood flow was observed to be non-pulsatile at pump rates exceeding 1500 rev./min (Fig. 3).
|
|
3.4. Indices of haemolysis
Daily observation showed the urine to be clear. Intermittent random testing showed no haemoglobinaemia. Weekly blood tests showed the haemogloblin level to remain constant with no significant rise in lactate dehydrogenase after the peri-operative period. Plasma free haemoglobin levels are shown in Table 1. These remained <5 mg/dl apart from the early postoperative period when factors such as non-cross-matched blood transfusion caused some haemolysis.
|
|
|
| 4. Discussion |
|---|
|
|
|---|
Until recently it was inconceivable that a rotary blood pump could function for years in the circulation without anticoagulation, thrombus formation or thromboembolism. The new blood pumps (Jarvik 2000 and TILVAS) are extremely promising in this respect, and we expect them to become the first realistic devices for permanent mechanical circulatory support [6]. Currently we suggest that these left ventricular assist devices are positioned in the left pleural cavity bridging between the apex of the dilated left ventricle and the descending thoracic aorta.
This position allows for short vascular out flow graft, and leaves the mediastinum clear in the event of subsequent cardiac transplantation. It also avoids a potentially thrombogenic cul-de-sac in the aortic root as may occur when an outflow graft is joined to the ascending aorta when the valve remains permanently closed. Similarly, intrapleural placement on the diaphragm avoids the need to open the abdominal cavity either during implantation or device removal. This reduces the amount of surgical trauma and expedites recovery by avoiding gastrointestinal ileus and delayed feeding.
In practice it is now evident that the patient with a non-pulsatile pump can develop pulsatility in the systemic circulation within hours or days of left ventricular offloading. Our experience with our first two Jarvik 2000 heart patients has confirmed an early return to pulsatility even in absolute end-stage dilated cardiomyopathy [7]. This, together with positive outcomes after prolonged non-pulsatile flow in the sheep model, gives us confidence that both axial flow and centrifugal blood pumps will provide safe and effective long-term circulatory support in the patient with heart failure.
In conclusion, The TILVAS is a novel and effective non-pulsatile blood pump which performs reliably without anticoagulation in the sheep model.
| Footnotes |
|---|
| Appendix A. Conference discussion |
|---|
|
|
|---|
Dr Saito: For the first question, the only heparin we gave in these animal studies was 3000 units during surgery. That is all. And so far as our examination of the inside the device is concerned, the heparin-coating system was stabilized. This is not a large number, but as far as we can tell, the detailed study of the surface of the device using the electron microscope showed the stability of the heparin-coated system.
Dr O. Frazier (Houston, TX, USA): What was the appearance of the aortic valve and the myocytes at the time of sacrifice?
Dr Saito: Well, as you know, the sheep left ventricle is quite small, but this device is so powerful, it sucked all the blood into this device. The aortic valve always remained closed.
Dr Frazier: what did it look like at sacrifice?
Dr Saito: At sacrifice, grossly there was no fusion of the aortic valve leaflets but microscopically the aortic wall was thinner than normal.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Morshuis, A. El-Banayosy, L. Arusoglu, R. Koerfer, R. Hetzer, G. Wieselthaler, A. Pavie, and C. Nojiri European experience of DuraHeartTM magnetically levitated centrifugal left ventricular assist system Eur. J. Cardiothorac. Surg., June 1, 2009; 35(6): 1020 - 1028. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Patel, E. S. Weiss, J. Schaffer, S. L. Ullrich, D. C. Rivard, A. S. Shah, S. D. Russell, and J. V. Conte Right Heart Dysfunction After Left Ventricular Assist Device Implantation: A Comparison of the Pulsatile HeartMate I and Axial-Flow HeartMate II Devices Ann. Thorac. Surg., September 1, 2008; 86(3): 832 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Ootaki, K. Kamohara, M. Akiyama, F. Zahr, M. W. Kopcak Jr., R. Dessoffy, and K. Fukamachi Phasic coronary blood flow pattern during a continuous flow left ventricular assist support Eur. J. Cardiothorac. Surg., November 1, 2005; 28(5): 711 - 716. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. H. Huber, P. Tozzi, M. Hurni, and L. K. von Segesser No drive line, no seal, no bearing and no wear: magnetics for impeller suspension and flow assessment in a new VAD Interactive CardioVascular and Thoracic Surgery, June 1, 2004; 3(2): 336 - 340. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Goldstein Worldwide Experience With the MicroMed DeBakey Ventricular Assist Device(R) as a Bridge to Transplantation Circulation, September 9, 2003; 108(90101): II-272 - 277. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Saito, S. Westaby, D. Piggot, S. Dudnikov, D. Robson, P. A. Catarino, C. Clelland, and C. Nojiri End-organ function during chronic nonpulsatile circulation Ann. Thorac. Surg., October 1, 2002; 74(4): 1080 - 1085. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. McCarthy and W. A. Smith Mechanical Circulatory Support--a Long and Winding Road Science, February 8, 2002; 295(5557): 998 - 999. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||