|
|
||||||||
Eur J Cardiothorac Surg 2001;20:956-960
© 2001 Elsevier Science NL
a Department of Cardiothoracic Surgery, University of Cape Town, Cape Town 7925, South Africa
b Herzzentrum, University of Leipzig, Leipzig, Germany
Received 11 June 2001; received in revised form 6 August 2001; accepted 7 August 2001.
Corresponding author. Current address: Cardiac Directorate, University Hospital of Wales, Heath Park, Cardiff, CF14 4XW, UK. Tel.: +27-21-406-6181; fax: +27-21-448-1145
e-mail: uvonopp{at}thoracic.cts.uct.ac.za
| Abstract |
|---|
|
|
|---|
Key Words: Heart Atrium Atrial fibrillation Treatment Radiofrequency Hyperthermia
| 1. Introduction |
|---|
|
|
|---|
The principle of the left atrial radiofrequency procedure is to produce a confluent transmural line of cellular death by raising tissue temperature to greater than 5255°C thereby [2], similar to a surgical incision, preventing electrical conduction across this line. Resistive heating is the primary mechanism of radiofrequency energy delivery and occurs in the subendocardial tissue where there is high current density, within a millimetre of the electrodetissue interface [2]. Deeper tissue heating occurs as a result of passive heat conduction from this subendocardial region along a tissue temperature gradient.
The left atrial radiofrequency procedure is still in a developmental phase in terms of both the ablation line configuration and mechanism of ensuring a confluent transmural line of cellular death. The ablation line configuration originally described by Melo et al. was to merely isolate the pulmonary veins [3]; however, a number of different ablation line configurations are being used clinically and the ideal configuration has not yet been established. Furthermore, individual radiofrequency generators, probes, and the duration of application for each system used may not be similar, and currently there are no published comparative guidelines when using them. Comparing the efficacy of different techniques is thus complicated by both the effectiveness of obtaining transmural cellular death, a confluent line of ablation lesions that may be operator dependent, as well as the configuration of the ablation lines. Clinically, the transmurality of the radiofrequency ablation lesion is not easily or routinely evaluated and therefore the specific mechanism of failure when it occurs is frequently unknown.
Most radiofrequency temperature studies have used ventricular myocardium and have been performed with catheters designed for closed intravascular use. In this study we evaluate the predictability of both the Boston Scientific Thermaline (Boston Scientific, Boston, USA) and Osypka (Grenzach-Wyhlen, Germany) systems to in vitro uniformly raise transmural atrial temperature above 55°C by open endocardial application, the objective being to establish comparative guidelines for the use of these tools, in terms of more consistently obtaining a transmural ablation lesion.
| 2. Methods |
|---|
|
|
|---|
The Osypka radiofrequency unit has a single rigid unipolar electrode. This resterilizable stainless steel 2x10-mm probe has a thermocouple incorporated into the central distal tip. In contrast, the Boston Scientific Thermaline probe is a non-reusable flexible probe (approximately 2 mm in diameter) consisting of seven unipolar (a single distal 8-mm, and six 12.5-mm) electrodes separated by 2-mm intervals in which two thermocouples are incorporated at the edge of each electrode. Any single or multiple combination of electrodes in the Thermaline probe can be selected for activation and are individually monitored.
The respective hand-held electrode was placed on the opened endocardial surface of either the right or left atrium, not in contact with saline, and a thermocouple (Voltcraft; -40 to 1200°C) placed on the corresponding epicardial surface, either central or at the edge in relation to the position of the radiofrequency electrode. Contact pressure was sufficient to cause an indentation of the endocardial surface by the electrode and maintained by hand, similar to the clinical situation and not specifically controlled. A fresh section of atrial tissue of between 4 and 6 mm in thickness was used for each measurement. When evaluating the Thermaline probe, only a single electrode (electrode no. 4) was activated for central epicardial measurements. However, two adjacent electrodes (electrode nos. 4 and 5) were activated for edge temperature measurements in between these two electrodes, in order to simulate the clinical situation when a linear ablation line would be made with more than one electrode.
Six measurements were made with each probe, three epicardial central measurements and three edge measurements.
The measurements with the Osypka system were repeated using a reference set point of both 65 and 60°C.
2.1. Statistics
Statistical analysis was done with Statistica 5.1 for Windows, 1998 (StatSoft Inc., Tulsa, OK, USA). Means±standard deviations are provided and comparisons between groups was by analysis of variance. Statistical significance was assumed at P<0.05.
| 3. Results |
|---|
|
|
|---|
|
|
|
|
| 4. Discussion |
|---|
|
|
|---|
Melo and co-workers originally reported using the Osypka HAT 200S for 30 s [4]. and then the Boston Scientific multi-electrode unipolar probe at 70°C for 60 s [3], and currently report the maintenance of sinus rhythm in approximately 75% of patients with atrial diameters of less than 5.0 cm, but only in 25% of patients with atrial dimensions of more than 5.5 cm (J. Melo et al., personal communication). Hemmer and co-workers report a 3-month success rate of 75% in patients with chronic atrial fibrillation and mitral valve disease using the Boston Scientific system with a 70°C set point and 120 s duration per application [5]. These are better results than our experience in a limited series of patients at the University of Cape Town using the Boston Scientific probe for 60 s per application, albeit with similar ablation line configurations. In this study we have also shown that the epicardial temperature measured in-between two active electrodes (epicardial edge temperature with the Boston Scientific unit) closely approximates the central measured temperature, suggesting that the multi-electrode configuration of the Boston Scientific probe will produce a confluent linear ablation lesion provided the duration of application is sufficient to produce transmural cellular death. Tissue temperatures of greater than 55°C result in denaturation of cellular proteins and cellular death [2]; however, the required duration of hyperthermia is not exactly known. In the in vivo clinical situation, the macro- and microcirculation would act as a heat sink and conduct heat away from the area, and therefore when extrapolating these in vitro measurements to the in vivo situation one could expect a potential reduction in transmural heat transmission. Hence, increasing the duration of application required to obtain the desired epicardial temperature by 23-fold would be appropriate in the clinical in vivo situation, provided this did not result in excessively high temperatures predisposing to perforation. Extrapolation of our results to the in vivo situation suggest that the Boston Scientific system should be used with a set point of at least 70°C and for a duration of at least 23 times that required to reach 55°C in vitro. Excessive epicardial temperatures were not obtained in our study at 120 s, and measured epicardial temperature at 120 s had plateaued and were no different to that at 80 s. Hence, clinical experience and extrapolation of our experimental in vitro observations support a 120 s duration of application with this probe as a useful safe guideline.
In contrast, the Osypka system (set point 70°C) appears to transfer energy and therefore heat more rapidly into the tissue as evidenced by epicardial temperatures being greater than 55°C within 10 s of application, and approximating probe temperature. This also suggests that the depth of the more rapid resistive tissue heating is greater with the Osypka system compared to the Boston Scientific system where slower time-dependent conductive heating is important. The earlier study by Kottkamp and co-workers (4 mm tip catheter electrode, 500 kHz Osypka HAT 200S set for 80°C) required approximately 20 s for subendocardial ventricular tissue at a depth of 2.53.0 mm to reach 50°C, and approximately 50 s for 5.56.0-mm deep tissue to reach these temperatures [6]. Other studies have required between 60 and 120 s to reach steady-state temperatures at myocardial depths of 23 mm [7,8]. Our observations of an even more rapid temperature rise with the Osypka unit is probably because we used a larger 10-mm electrode, which utilizes greater radiofrequency power and creates larger lesions [2]. However, thinner walled atrial tissue may also have a different distribution of current density and therefore temperature response compared to ventricular tissue.
The tendency for some individual epicardial temperatures to exceed probe temperature with the Osypka system, which was not observed with the Boston Scientific system, also suggests more effective energy transfer into the tissue, possibly because of the higher frequency used by the Osypka system, 560 as opposed to 460 Hz. The electrode size of both the Osypka and Boston Scientific single electrode were similar, albeit of different design. Overshoot of both the probe and epicardial temperature with the Osypka system, in terms of the selected set point, is a concern in terms of safety and rapidity of feedback control. Excessive tissue temperature could result in necrotic perforation.
Kottkamp and co-workers originally reported their clinical use of the Osypka system at a set point of 6075°C for 2030 s per application and reported an 86% success rate with this procedure, in patients undergoing combined mitral valve surgery who where previously in chronic atrial fibrillation for at least 1 year prior to surgery, with apparently no major influence of left atrial size [9]. Our results support the use of a slightly lower set point of 65°C with the 10-mm Osypka electrode because of the observed tendency for hyperthermic overshoot. An epicardial temperature of 55°C was reached within 10 s, 25% of the time required by the Boston Scientific system, and extrapolation of our results to the in vivo situation, as previously discussed, would suggest that a recommended clinical duration of application with this probe is 30 s per application.
The limitations of this current study is that 46 mm cadaver porcine atrial wall as opposed to living human atrial tissue was used and that a single point, as opposed to an array, epicardial thermocouple was used. Nevertheless, we have shown that the two radiofrequency generators evaluated have different efficiencies of thermal energy transfer into tissue.
In conclusion, the Boston Scientific Thermaline system (set point of 70°C) requires an in vitro application of at least 40 s to elevate epicardial temperature of 46 mm tissue to 55°C. In contrast, the Osypka system is more effective in terms of energy transfer, requiring less than 10 s to achieve a similar in vitro epicardial temperature. Extrapolation of our results to the in vivo situation suggests that a 120-s duration of application at a set point of at least 70°C would appear to be a reasonable safe clinical recommendation for the Boston Scientific system. Higher epicardial temperatures, equal to the selected set point, and a tendency to overshoot both epicardial and probe temperatures suggest that both a lower set point of 65°C and a 75% shorter duration of application of 30 s is more appropriate and safer in vivo with the Osypka system. Application times and temperature set points used to achieve transmural radiofrequency ablation lesions differ with different radiofrequency generators and probes.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y. Inoue, I. Kiso, R. Takahashi, A. Mori, and K. Motogami Beating-heart epicardial radiofrequency ablation: optimal temperature setting Ann. Thorac. Surg., July 1, 2004; 78(1): 308 - 311. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Khargi, A. Laczkovics, K. Muller, and T. Deneke A possible surgical technique to avoid esophageal and circumflex artery injuries using radiofrequency ablation to treat atrial fibrillation Interactive CardioVascular and Thoracic Surgery, June 1, 2004; 3(2): 352 - 355. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Laczkovics, K. Khargi, and T. Deneke Esophageal perforation during left atrial radiofrequency ablation J. Thorac. Cardiovasc. Surg., December 1, 2003; 126(6): 2119 - 2120. [Full Text] [PDF] |
||||
![]() |
N. Doll, M. A. Borger, A. Fabricius, S. Stephan, J. Gummert, F. W. Mohr, J. Hauss, H. Kottkamp, and G. Hindricks Esophageal perforation during left atrial radiofrequency ablation: Is the risk too high? J. Thorac. Cardiovasc. Surg., April 1, 2003; 125(4): 836 - 842. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |