Eur J Cardiothorac Surg 2002;21:478-482
© 2002 Elsevier Science NL
a Department of Cardiology, Interventional Cardiology Lab and Cardiomyoplasty Project Unit, Legnago General Hospital, Verona, Italy
b Department of Biomedical Sciences, University of Padua, Padua, Italy
c Department of Cardiology, Cittadella General Hospital, Padua, Italy
d Division of Rehabilitation, Maugeri Foundation, Montescano Medical Center, Pavia, Italy
Received 7 July 2001; received in revised form 27 September 2001; accepted 10 December 2001.
* Corresponding author. Division of Cardiology and Advanced Heart Failure Center, Legnago General Hospital, Via Gianella, 37040 Legnago, Verona, Italy. Tel.: +39-442632329; fax: +39-442632311
e-mail: jackyheart{at}katamail.com
| Abstract |
|---|
|
|
|---|
Key Words: Heart failure Ultrasound
| 1. Introduction |
|---|
|
|
|---|
In order to maintain long-term partial transformation of the LD wrap avoiding excessive muscular degeneration and an extremely long contractionrelaxation cycle, fewer impulses per day than in standard clinical stimulation should be delivered. This can be obtained by providing the LD wrap with daily periods of rest (demand stimulation protocol), based on a heart rate cut-off [6].
Despite the encouraging preliminary results in sheep and humans [6,7], there are no definitive data on the real systolic assistance that demand dynamic cardiomyoplasty (DDCMP) offers, limiting its acceptability by cardiologists. There are two main difficulties in the assessment of DDCMP assistance in vivo using non-invasive methods such as echocardiography. Firstly, the movement of the ascending aorta during assisted beats does not allow correct alignment of the Doppler signal during transthoracic echocardiography. Secondly, beat-by-beat measurements during normal stimulation are relatively inaccurate because they are greatly dependent on variations in stroke volume. In order to overcome these difficulties, we measured the aortic flow velocity with an intravascular Doppler flow wire. This is a 0.018 inch wire with a Doppler transducer near its tip that permits the measurement of blood flow velocities inside the coronary and peripheral vessels [8]. We describe our experience on the use of the Doppler flow wire in the evaluation of the systolic assistance provided to the heart by cardiomyoplasty in a series of seven patients; in these patients, demand protocol was introduced, in some between 6 and 12 months after operation and in others after years of continuous stimulation.
| 2. Patients and method |
|---|
|
|
|---|
In three patients who came from the Department of Cardiovascular Disease of Legnago General Hospital, Verona, Italy and were operated at the Institute of Cardiovascular Surgery of the University of Padua, the LD was stimulated with a single impulse at a 1:3 synchronisation ratio after a healing period of 1014 days. An impulse was added every week at a 23-ms pulse interval (43 Hz) for a final burst of four impulses with a cardiac amplitude >5 V and pulse width of 1.5 ms. After 612 months of this light daily stimulation, the patients were submitted to the demand regime allowing the LD wrap a daily period of rest.
The remaining four patients were referred to us from the Cardiology Rehabilitation Center, IRSS Fondazione Maugeri, Montescano, Pavia, Italy, due to their clinical deterioration. They had been operated several years before in another institute. Demand stimulation protocol was introduced in these patients after many years of continuous stimulation (6 pulses at 31 ms of interval, 5100 ms of synchronisation delay, LD/heart ratio 1:21:4, cardiac amplitude >5 V, cardiac pulse width 1.5 ms). Individual data are showed in Table 1.
|
After having obtained informed consent, a mechanogram registration was performed to establish the LD wrap properties. Contractile characteristics of LD wrap were monitored using a standard polygraph (MegaCart or Mingophon, Siemens Elema, Solna, Sweden): the electrocardiogram and pressure changes due to LD contraction were simultaneously recorded as described previously [9]. The dynamic characteristics of the LD wrap were determined from the LD response to stimuli delivered at increasing frequency up to tetanic fusion frequency (TFF). In this way, a smooth contraction curve was plotted: the faster the fibres, the higher the TFF.
A 0.018 inch peripheral FlexTM Doppler flow wire (Cardiometrics Inc, Mountain View, CA, USA) was advanced under fluoroscopy into the descending thoracic aorta through a 4F catheter introduced via a 5F femoral sheath. The catheter served to maintain the wire at a steady position, coaxial to the vessel throughout the entire cardiac cycle. A peripheral type flow wire was chosen for its stiffness and easy parameter setting for aortic measurements. It was connected to its digital display, and registration of aortic flow velocity was performed at standard settings. Three 1-min periods with the stimulator off and three 1-min periods with the stimulator on (ratio of assisted:unassisted beats=1:3) were recorded, avoiding haemodynamic changes due to external factors. Quantitative data were expressed as percentage variation and calculated as mean±SD. Continuous values were compared using the Student's t-test and Pearson's correlation coefficient. A value of P<0.05 was considered significant. Individual data are shown in Table 1.
| 3. Results |
|---|
|
|
|---|
|
|
|
Mean NYHA class at follow-up was significantly lower when compared with pre-operative and post-DCMP: 1.4±0.5 versus 3±0 (P<0.01) and 2.0±0.8 (P=0.03) whereas mean EF at follow-up was significantly higher: 37.7±5.8% versus 26.8±2.4% (P<0.01) and 27.1±2.1% (P=0.004), respectively (Fig. 2). Moreover, the value of TFF was significantly higher at follow-up (23±8.9 versus 35±7.2 Hz, P<0.05), (Table 3; Fig. 3 ).
|
|
Mean fluoroscopy time was <1 min and mean procedural time was <15 min. No complications were observed during and after the procedure.
| 4. Discussion |
|---|
|
|
|---|
There are two principal hypotheses on the mechanism by which DCMP relieves symptoms in patients who benefit from it: limited heart dilatation and enhanced systolic contraction [1215]. The enhancement of systolic contraction has been suggested but not proven by measurements of ejection fraction [16,17], stroke volume [17,18] and study of the P/V loops [19] in DCMP patients. Unfortunately, despite their results these studies were unable to explain the great variability in the response to classic cardiomyoplasty, probably related to graft degeneration.
Demand stimulation protocol was introduced [6,17] in order to avoid complete LD wrap transformation caused by the continuous stimulation protocol of the FDA phase II trial in use by the American Cardiomyoplasty Group [5] and to homogenise the patient response. Up to now, no proof of real systolic assistance using this protocol has been produced.
It is well established that a muscle fully transformed by continuous stimulation displays significant loss in power, generally attributed to fibre-type change or loss of type 2 myofibrils (fast contracting myofibrils). The insertion of rest periods during chronic electrical conditioning preserves myofibril cross-sectional area and produces fatigue-resistant fibre distributions; thus, a more powerful fatigue-resistant muscle is created. The improved performance of such a LD is thought to be due to the maintenance of an intermediate level of transformation of the LD wrap, thanks to the demand regimen. Moreover, the differences in TFF between patients who started the demand regimen 612 months after operation and those who switched to it after years of continuous stimulation support our theory: activityrest stimulation protocol is superior to daily continuous stimulation in minimising incremental activity-induced muscle damage and thus maintaining an intermediate level of LD transformation.
Due to the difficulty in measuring a parameter directly correlated to stroke volume and cardiac output by standard echocardiography, we decided to assess the efficacy of the flow wire in measuring aortic blood flow velocity in a feasibility study [20]. In fact, aortic flow velocity is a term in the simplified continuity equation: Q (cardiac output)=V (aortic flow velocity) x A (cross-sectional area of the vessel). The flow wire was chosen as a technique that is independent of aortic movements secondary to LD contraction and capable of beat-to-beat measurements. In this feasibility study, the flow wire technique was tested in non-heart failure patients. Every two normal cardiac contractions, an ectopic ventricular beat was introduced using a temporary cardiostimulation catheter: the flow wire was sensitive enough to measure the changes in cardiac output between normal and post-ectopic beats, with an excellent correlation to transthoracic Doppler echocardiographic findings.
Our data suggest that flow wire measurements in DDCMP patients are able to reveal the real improvement in cardiac assistance provided by the powerful, partially transformed LD. The activityrest demand stimulation protocol based on heart rate seems to enhance muscle capabilities over time, reversing fast-to-slow muscle transformation.
Combined mechanographic evaluation proved that cardiac assistance is related to LD properties: the equation resistant muscle=good muscle could be changed to faster muscle=improved systolic assistance. In fact, comparison between pre-operative and follow-up data demonstrated a consensual increase in EF, TFF and improved NYHA class.
One limit of our study may be the late initiation of demand stimulation in the last four patients that came from centres other than our own. This could explain certain dispersion in our data. Moreover, in our study application of the protocol was not done in patients immediately after surgery: on one hand, geographical dispersion of the centres involved in the research on DCMP made immediate post-operative inclusion in the protocol impossible; on the other hand it was decided to introduce the demand protocol after a period of light stimulation in order to partially condition the LD and provide sufficient benefit in the early post-operative period.
Another limit is the lack of a direct comparison of flow wire data between the FDA and demand protocol. On the one hand, at the time of the application of the FDA protocol, the Doppler flow wire technology available in non-academic hospitals such as our own did not permit such measurements. On the other hand, in the previous years there has been a lack of interest in DCMP, which still remains an experimental technique.
| 5. Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Rigatelli, U. Carraro, R. Riccardi, and G. Rigatelli Demand dynamic biogirdling: ten-year results. J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): e58 - e59. [Full Text] [PDF] |
||||
![]() |
L. Barandon, T. Couffinhal, P. Dufourcq, P. Alzieu, D. Daret, C. Deville, and C. Duplaa Repair of Myocardial Infarction by Epicardial Deposition of Bone Marrow Cell-Coated Muscle Patch in a Murine Model Ann. Thorac. Surg., October 1, 2004; 78(4): 1409 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Carraro Video-assisted thoracoscopic transplantation of myoblasts into the heart Ann. Thorac. Surg., July 1, 2004; 78(1): 14 - 16. [Full Text] [PDF] |
||||
![]() |
G. Rigatelli, G. Rigatelli, M. Barbiero, A. Cotogni, A. Bandello, R. Riccardi, and U. Carraro "Demand" stimulation of latissimus dorsi heart wrap: experience in humans and comparison with adynamic girdling Ann. Thorac. Surg., November 1, 2003; 76(5): 1587 - 1592. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Benicio, L. F. P. Moreira, F. Bacal, N. A. G. Stolf, and S. A. Oliveira Reevaluation of long-term outcomes of dynamic cardiomyoplasty Ann. Thorac. Surg., September 1, 2003; 76(3): 821 - 827. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F. P. Moreira, A. Benicio, F. Bacal, E. A. Bocchi, N. A.G. Stolf, and S. A. Oliveira Determinants of long-term mortality of current palliative surgical treatment for dilated cardiomyopathy Eur. J. Cardiothorac. Surg., May 1, 2003; 23(5): 756 - 764. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Rigatelli, U. Carraro, M. Barbiero, R. Riccardi, F. Cobelli, M. Gemelli, and G. Rigatelli A Review of the Concept of Circulatory Bioassist Focused on the "New" Demand Dynamic Cardiomyoplasty: The Renewal of Dynamic Cardiomyoplasty? Angiology, May 1, 2003; 54(3): 301 - 306. [Abstract] [PDF] |
||||
![]() |
G. Rigatelli, M. Barbiero, G. Rigatelli, R. Riccardi, F. Cobelli, A. Cotogni, A. Bandello, and U. Carraro Maintained benefits and improved survival of dynamic cardiomyoplasty by activity-rest stimulation: 5-year results of the Italian trial on 'demand' dynamic cardiomyoplasty Eur. J. Cardiothorac. Surg., January 1, 2003; 23(1): 81 - 85. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||