Eur J Cardiothorac Surg 2002;22:228-232
© 2002 Elsevier Science NL
a Department of Cardiac Surgery, Oxford Heart Centre, The John Radcliffe Hospital, Headleyway, Headington, Oxford OX3 9DU, UK
b Texas Heart Institute, Houston, TX, USA
c Department of Cardiovascular Surgery, University of Freiburg, Freiburg, Germany
d Jarvik Heart Inc., New York, NY, USA
Received 18 September 2001; received in revised form 8 February 2002; accepted 9 April 2002.
* Corresponding author. Tel.: +44-1865-220269; fax: +44-1865-220268
e-mail: swestaby{at}ahf.org.uk
| Abstract |
|---|
|
|
|---|
Key Words: Heart failure Left ventricular assist device
| 1. Introduction |
|---|
|
|
|---|
Axial flow pumps have already entered the clinical arena [1,2]. The Micromed-Debakey and Thermo Cardio Systems II left ventricular assist devices (LVADs) are smaller than the Novacor and Thermo Cardio Systems pusher plate pumps but have the same anatomical connections. Both are implanted via median sternotomy with an inflow cannula in the left ventricular apex, an external pocket and a long outflow graft to the ascending aorta. The Jarvik 2000 Heart adopts a completely different approach. Measuring only 2.4 cm in diameter by 5.5 cm in length and weighing 85 g, this is the first blood pump to fit within the native heart without an inflow cannula. The pump offloads to the descending thoracic aorta and only one body cavity is opened during implantation [3].
The Jarvik 2000 Heart is now undergoing clinical trials for both bridge to transplantation (United States) and permanent mechanical circulatory support (Europe). This report describes our early clinical experience with specific reference to the benefits of the intraventricular position.
| 2. Patients and methods |
|---|
|
|
|---|
|
|
2.2. The implant procedure
The pump was inserted via left thoracotomy into the apex of the dilated left ventricle with careful alignment parallel to the interventricular septum (Fig. 2)
. The outflow graft makes a gentle turn around the costophrenic angle to the descending thoracic aortic. The power line was brought through either the upper right abdominal quadrant or scalp skin via the titanium button. The distal graft anastomosis was made first. A retaining cuff was then sewn onto the apex of the left ventricle and the circumscribed muscle punched out with a cork bore instrument avoiding the papillary muscles. The system was then de-aired and the device switched on before weaning from cardiopulmonary bypass.
|
| 3. Results |
|---|
|
|
|---|
3.1.1. Bridge to transplantation
Seven patients have undergone heart transplantation and were discharged from the hospital. Two have died during support, and one continues with the device while awaiting heart transplantation. The seven transplant recipients have average follow-up period of 5.6 months (range 0.711.2 months), and all remain in NYHA Class I. For the series as a whole, the average duration of support has been 84 days (range 13214), and total duration of support has been 826 days (Table 2).
|
3.1.2. Permanent implants
One patient died during support, and six are ongoing. The death occurred in a 120-kg male who developed a slowly accumulating subdural haematoma at the skull pedestal site. This required surgical evacuation on two occasions. With residual neurological disability, he remained ventilated and eventually died from tricuspid regurgitation and right heart failure 94 days after the implant.
3.2. Clinical outcome
3.2.1. Bridge to transplantation
Of the bridge-to-transplant patients the duration of support has ranged between 13 and 214 days (mean 84 days). Two patients had superficial abdominal power cable infection but this did not alter the clinical course. One patient had a major haemorrhage from a gastric ulcer and a separate arteriovenous malformation in the small intestine. This patient required multiple blood transfusions followed by laparotomy, cauterization of the ulcer and resection of the arteriovenous malformation. Anti-coagulation was discontinued completely in this patient whose international normalized ratio ranged from 1.0 to 1.2.
There has been no thromboembolism or infection in the device itself. No thrombus was found in explanted the devices or within the left ventricle of the seven patients successfully transplanted. These patients have an average follow-up of 5.6 months after transplantation (range 0.711.2 months) and all are NYHA Class I.
3.2.2. Permanent implants
The three surviving Oxford patients left hospital within 4 weeks and are NYHA I or II living in the community. Duration of support ranges from 94 to 455 days (mean 208 days). All three Freiburg patients are NYHA I or II between 14 and 93 days postoperatively (mean 41 days). There have been no device-related complications in these patients. The skull-mounted percutaneous power delivery system has healed satisfactorily without infection in all six patients and has proven user-friendly. None of the patients can hear or feel the device within the chest. All of these dilated cardiomyopathy patients have shown marked improvement in native left and right ventricular function. With the device temporarily switched off, retrograde functional aortic regurgitation from the descending thoracic aorta was measured estimated at between 600 and 800 ml/min. This was well tolerated by the improving left ventricle. Accidental interruption of power supply led to the device being off for 45 min in one patient. There were no problems on restarting the pump.
3.3. Haemodynamic assessment
Compared to preoperative values, measured cardiac output improved considerably with LVAD support. Continuous cardiac output monitoring in the early postoperative period documented pump outputs varying between 3 and 8 l/min at a fixed pump speed of 10 000 rpm depending on afterload. Mean arterial pressure varied correspondingly from 50 to 80 mmHg. Though the pump provides continuous non-pulsatile flow reduced, pulsatility was discernible in the systemic circulation at speeds of 8000 to 11 000 rpm. At 12 000 rpm, the arterial pressure wave was virtually flattened with less than 15 mmHg pulse pressure. The progressive decrease in pulse pressure within increasing pump speed reflected the increased capture of blood flow from the left ventricle through the pump during diastole. At low pump speeds (8000 and 9000 rpm), the aortic valve opened with ejection through the left ventricular outflow tract and a diacrotic notch on the arterial pressure trace.
Even though pulsatility remained in the systemic circulation at 10 000 rpm, the diacrotic notch disappeared indicating that the aortic valve remained closed. At 10 000 and 11 000 rpm, flow through the left ventricular outflow tract was minimal. At 12 000 rpm the aortic valve remained closed, demonstrating that left ventricular volume was too low for ejection. Pulsatility was produced by ventricular preload to the pump during systole. For most patients, a pump speed of 9000 or 10 000 rpm was chosen to allow ejection through the left ventricular outflow tract. The device had no significant negative impact on left ventricular dynamics. Echocardiography confirmed left ventricular unloading at these speeds and Doppler imaging showed continuous flow through the pump with systolic augmentation. At all pump speeds, the left atrial pressure was decreased by 50% below preoperative levels.
The average plasma free haemoglobin level was 14.1 mg/dl in the bridge-to-transplant patients and 8.4 mg/dl in the permanent implants. No patient had significant haemolysis and the variety of plasma free haemoglobin level probably occurred through different laboratory baselines. The highest recorded plasma lactate dehydrogenase was 605 IU.
| 4. Discussion |
|---|
|
|
|---|
The intraventricular position conveys major advantages. The alignment of the pump inflow remains constant and is not altered by a change in left ventricular shape during unloading. The consistent flow pattern mitigates against turbulence and haemolysis [5]. The apex of the native heart is not tethered by an extracardiac device in a fixed pocket. Left ventricular mechanics are virtually unaltered, so much so that discontinuation of power and functional aortic regurgitation can be tolerated for long periods as ventricular function improves. The pump itself is encapsulated by blood and myocardium, lessening the risk of device and pocket infection. Only the left pleural cavity is entered at implantation so that the median sternotomy for cardiac transplantation is easier and safer.
The skull-mounted post auricular power line mimics a hearing aid and is virtually invisible when the hair grows. There has been no infection with this approach [6]. However, the miniature pump with a fine power line and no external pocket also seems less prone to conventional drive line infection. The external battery and controller are easily portable on a belt or shoulder bag. There is only one speed knob for the patient to regulate and this is usually kept constant at 900010 000 rpm. The user-friendly nature of the system allows discharge from hospital in less than 4 weeks followed by an unrestricted life in the community. This includes international travel by air and reinstatement of a driving licence.
Given the functional status and quality of life of our early patients, the importance of systemic pulse pressure is questionable. Most of the patients had prolonged periods with pulse pressure less than 15 mmHg with the aortic valve persistently closed during the cardiac cycle. Our preferred approach is to allow antegrade ejection through the aortic valve and promote a synergistic relationship between the pump and native left ventricle. By partially offloading, the Jarvik 2000 Heart improves the position of the left ventricle on the FrankStarling relationship. This is a different strategy from the pusher plate LVADs which empty and completely replace the native left ventricle [7]. Echocardiography was used to ensure that lower pump speeds (900010 000 rpm) provided adequate offloading. In this setting lack of pulsatility suggests inadequate pre-load, inappropriate afterload or persistently poor left ventricular function. On exercise, intrinsic mechanisms boost cardiac output through increased heart rate and enhanced contractility of the offloaded left ventricle. Flow increases both through the device and via the left ventricular outflow tract. This means that sophisticated activity-responsive electronic mechanisms to increase pump speed are unnecessary. We also find that an increase in pump speed by the patient during exercise is unusual.
In summary, this early experience with the Jarvik 2000 Heart confirms the user-friendly nature of the device and low event rate. The novel intraventricular position without an inflow cannula or graft may be the key to overcoming major complications such as haemolysis, thromboembolism, and device infection.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Haj-Yahia, E. J. Birks, P. Rogers, C. Bowles, M. Hipkins, R. George, M. Amrani, M. Petrou, J. Pepper, G. Dreyfus, et al. Midterm experience with the Jarvik 2000 axial flow left ventricular assist device J. Thorac. Cardiovasc. Surg., July 1, 2007; 134(1): 199 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Clegg, D. A. Scott, E. Loveman, J. L. Colquitt, P. Royle, and J. Bryant Clinical and cost-effectiveness of left ventricular assist devices as a bridge to heart transplantation for people with end-stage heart failure: a systematic review and economic evaluation Eur. Heart J., December 2, 2006; 27(24): 2929 - 2938. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. El-Magharbel Ventricular Assist Devices and Anesthesia Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2005; 9(3): 241 - 249. [Abstract] [PDF] |
||||
![]() |
B. A. J. M. de Mol Mechanical Support in Acute Perioperative Heart Failure: Are Assist Devices Smart Enough to Heal the Heart? Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2003; 7(1): 105 - 109. [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||