Eur J Cardiothorac Surg 2003;23:277-282
© 2003 Elsevier Science NL
a Department of Surgery and Research Center, Montreal Heart Institute, 5000 Belanger Street East, Montreal, QC, Canada H1T 1C8
b Cardiovascular Surgery Unit, Arnaud de Villeneuve Teaching Hospital, 371 Av. Doyen G. Giraud, Montpellier, 34295, France
c Department of Pathology and Research Center, Montreal Heart Institute, 5000 Belanger Street East, Montreal, QC, Canada H1T 1C8
d Department of Cardiology and Research Center, Montreal Heart Institute, 5000 Belanger Street East, Montreal, QC, Canada H1T 1C8
Received 17 September 2002; received in revised form 27 November 2002; accepted 28 November 2002.
* Corresponding author. Tel.: +1-514-376-3330x3715; fax: +1-514-376-1355
e-mail: lpperrau{at}icm.umontreal.ca
| Abstract |
|---|
|
|
|---|
, endothelium-dependent relaxations to bradykinin were studied. Gomori trichrome and hematoxylineosin safran staining were used for histological evaluation. Results: Exposure to radiofrequency 1 mm from the coronary arteries caused a significant decrease in endothelium-independent contractions to KCl and endothelium-dependent relaxations to bradykinin compared to controls (P<0.05). No significant decrease of endothelium-dependent relaxations occurred for rings exposed to radiofrequency at a distance of 5 and 10 mm, compared to controls. Histological examination showed endothelial disruption and medial smooth muscle cells at different stages of necrosis up to 5 mm from the radiofrequency application site. Conclusions: Radiofrequency may induce coronary endothelial functional and morphological damages when applied less than 5 mm from the artery. Caution must be exerted during left atrial radiofrequency application due to the proximity of the circumflex artery.
Key Words: Coronary endothelial injury Radiofrequency Ablation atrial fibrillation
| 1. Introduction |
|---|
|
|
|---|
During RF Maze procedures, some lesions are made at close proximity to the coronary arteries, especially the circumflex [11,12]. The purpose of this study was to determine the effects of application of RF energy induced by handheld probes on coronary endothelial function and structure, and establish the relationship between coronary lesions and the distance of RF application from the coronary artery.
| 2. Materials and methods |
|---|
|
|
|---|
|
|
For each swine, three epicardial radiofrequency lesions (20 W, 20 s duration and 60 °C temperature), 2 cm in length were made with a slow mobilization of the tip of the probe on the right side of the left anterior descending artery (LAD) respectively at 1, 5, and 10 mm away from the artery (Fig. 3) . Three lesions were also created on the ventricular side of the right coronary artery (RCA). The left circumflex artery (LCX) served as control. The heart was then excised and placed in a cold modified Krebs-bicarbonate solution (composition in mmol/l: NaCl 118.3, KCl 4.7, MgSO4 1.2, KH2PO4 1.2, glucose 11.1, CaCl2 2.5, NaHCO3 25, and ethylenediaminetetraacetic acid 0.026).
|
Each arterial ring was stretched to the optimal point of its active length-tension curve as determined by measuring the contraction to potassium chloride (KCl) 30 mmol/l at different levels of stretch (stretch for optimal response was usually 3.5 g). The maximal contraction of rings was then obtained with addition of potassium chloride (KCl 60 mmol/l). After obtention of a plateau, all baths were washed twice with modified Krebs-bicarbonate solution and indomethacin (10-5 mol/l to exclude production of endogenous prostanoids), propranolol (10-7 mol/l to prevent the activation of ß-adrenergic receptors) and ketanserin (10-6 mol/l to block serotonin 5-HT2 receptors) were added in each bath.
After 45 min of stabilization, prostaglandin F2
(PGF2
, range 2x10-6 to 3x10-5 mol/l) was added to obtain a contraction averaging about 50% of the maximal contraction to KCl. Endothelium-dependent relaxations to bradykinin (BK; an agonist which binds to receptor coupled to Gq-proteins) at incremental concentrations (10-12 to 10-6 mol/l) were recorded. At the end of the experiment, endothelium-independent relaxations were studied using a bolus of sodium nitroprusside (SNP 10-5 mol/l).
2.4. Histological analysis
Histological studies were done on animals killed 5 min after the application of the RF epicardial lesions. The heart was excised rapidly and placed in a formalin solution during 24 h. Paraffin blocks were molded from each lesion and three microscopic sections were made in the plane perpendicular to the long axis of the RF ablation zone for each lesion. Gomori's trichrome and hematoxylineosin safran (HES) staining were used, to evaluate the coronary vessels and surrounding myocardium.
2.5. Statistics
Continuous variables are expressed as mean±standard error of the mean (SEM). Preliminary two-way repeated measures analysis of variance (ANOVA) showed interaction between group and bradykinin concentrations. Due to this interaction, one-way repeated measures ANOVA was used to analyze the data. Contrasts were built to make predetermined comparisons of the means (Fig. 4)
. One-way repeated measures ANOVA and contrasts was also used to analyze the data of Table 1. Differences were considered statistically significant if the P-value was less than 0.05.
|
|
| 3. Results |
|---|
|
|
|---|
(range 2x10-6 to 3x10-5mol/l) was recorded for all groups (group 1, 5, 10 mm, and controls). There was a statistically significant lower amplitude of contractions to KCl and PGF2
in the rings exposed to RF at 1 mm, compared to the other three groups. There was no statistically significant difference in contractions to KCl and PGF2
between the rings exposed to RF at 5 and 10 mm, when compared to the control group (Table 1).
3.1.2. Relaxations
3.1.2.1. Endothelium-dependent relaxations
There was a statistically significant decrease (P<0.05) in endothelium-dependent relaxations to BK in rings exposed to RF at 1 mm, compared with the control group and the others groups. There was no statistically significant decrease of relaxations to BK between rings exposed to RF at 5 and 10 mm, compared with the control group (Fig. 4).
3.1.2.2. Endothelium-independent relaxations
No statistically significant differences in relaxation to SNP was observed between groups with all rings achieving 100% relaxation (data not shown).
3.2. Macroscopy and histology
Application of radiofrequency energy caused a pale yellow discoloration of the epicardium in an area extending 5 mm around the tip of the ablation probe, inducing a 10-mm wide rigid lesion 20 mm in length (Fig. 3).
After formalin fixation, the lesions were readily visible on cut sections. The lesions were stiff and pale compared to the surrounding myocardial tissue. Microscopic examination showed a well delimited RF lesion on the myocardium. The borders of these lesions were precisely identifiable with the Gomori Trichrome stain. The depth of lesions was 5±1 mm (Figs. 5 and 6) . At the center of the RF lesion, in a radius of 4 mm around the application point, myocardial cells showed complete coagulation necrosis surrounded by edema and hemorrhage. At the periphery of the lesion, contraction band necrosis was visible at a depth of 1 mm in depth encircling the previously described area of coagulation necrosis. The section of LAD artery located away from the zone of RF application, was morphologically normal (Fig. 7A) . The section located inside the RF effect area (Fig. 7B), showed arterial wall damage with the variable extent of disruption of the endothelial layer, ranging from partial detachment to total wrench. Medial smooth muscle cells showed different stages of necrosis with lysis of the cytoplasm and nuclear pycnosis.
|
|
|
| 4. Discussion |
|---|
|
|
|---|
The size of lesions induced by RF energy depends of various factors, anatomical as well as type of devices and procedure (power, duration, temperature) used. In the present study, in which RF was applied on the ventricular wall, which is thicker than the atrium, RF lesions penetrated deeply (5 mm) into the myocardium. The parameter used (20 W power and 20 s duration) were less aggressive than the settings used in previous reports [1214] in which the size of the lesion increases with the duration of the ablation [15] even when adjusted to the thickness of the myocardial wall. Wittkampf et al. have demonstrated, that the size of the RF lesion produced with a 2-mm large catheter tip electrode also depends on the power applied by the generator and that maximum lesion areas are achieved in less than 20 s [16]. However, in the present study, constant duration and power were selected for each lesion created to compare the effects of RF relative to its proximity to the coronary artery. The site of application was the same for all procedures on the LAD and RCA and only the distance of application from the artery varied. Furthermore, there is no difference between the endothelium-dependent reactivity of the LAD and RCA in swine as previously demonstrated in our laboratory (data not shown).
The better understanding of the role of the endothelium in regulation of the vascular tone and coagulation control has led to the knowledge that structural and functional integrity of the endothelial layer are fundamental to avoid spasms or thrombosis of the coronary artery during or after surgical procedures [9,10]. The proximity of the circumflex artery to lesions made during the Maze procedure, especially when a RF is applied between the pulmonary veins to the mitral annulus [11] may jeopardize its integrity. As observed in the present experiments, when the coronary artery is involved in the necrotic process, both muscular and endothelial layers bear morphological lesions with edema, intimal disruption, exposure of the subendothelial layer, and muscular necrosis. These findings have functional implications including the decrease of muscular dependent contraction to KCl and decrease or suppression of endothelium-dependent relaxation in arteries located 1 mm away from the site of RF application and not at 5 mm, demonstrating that the section of the artery located outside the zone of RF effect maintains its functionality. The decreased contractions at 1 mm reflect morphological alterations of vascular smooth muscle cells described above but the level of contraction obtained was sufficient after exposure to PGF2
to study endothelium-dependent relaxations, which was also significantly altered.
4.1. Limitations
The first limitation of this study concerns creation of RF lesions in normal porcine ventricular myocardium with normal coronary arteries. However, patients treated with RF for chronic atrial fibrillation may also be free of coronary disease. Secondly, in clinical practice, RF is usually applied on the atrial endocardium rather than on the ventricular epicardium. The ventricle was chosen as the site of radiofrequency ablation to allow perfect visualization of the coronary arteries and a reproducible accuracy of the measurement of the distance between RF application in relation to the artery. The choice of ventricular wall also allowed accurate assessment of the maximum depth of extension of the lesions. Since application of RF ablation in these experiments involved the ventricular myocardium where the thermal spread is very limited, compared with the atrial wall, the usual site of radiofrequency ablation in the surgical practice, which has a great propensity for thermal spread, consequently, the present study is probably an underestimation of the actual risk of coronary lesions induced by radiofrequency during clinical procedures. The lesions induced in this beating heart model may not reflect exactly those made in the atrium during cardiac surgery in humans in which lesions are performed on an arrested heart after cardioplegia administration and at lower myocardial temperatures in case of hypothermic cardioplegia, which could potentially reduce the size of the lesions. Obviously this would not be the case when normothermic or tepid cardioplegia are used. In preliminary experiments of endocardial left atrial RF application performed under cardiopulmonary bypass and cardioplegic arrest there was no evidence of endothelial dysfunction of the circumflex artery, when compared to LAD and RCA used as controls (data not shown). Direct anatomical measurements of the distance between the endocardium and the circumflex artery varied from 6±1 mm in porcine explanted hearts, which could explain the lack of endothelial dysfunction observed in this set of experiments. The circumflex artery was used as control in organ chamber experiments with epicardial application of RF because its limited length in swine precluding its use for the study of the distance-lesion relationship.
4.2. Conclusion
Radiofrequency ablation with a 20 W power, 60 °C temperature and 20 s duration for 2 cm ablation induces coronary endothelial dysfunction associated with morphological damages when applied less than 5 mm from the coronary artery, in a beating heart porcine model. As the extent and depth of RF-induced lesions may vary depending on anatomical characteristics, the presence of blood, the probe temperature, the power, and duration of application, this study shows that a RF ablation can induce endothelial dysfunction and structural lesions of the coronary endothelium. Caution must be exerted during RF application in the left atrium due to the proximity of the circumflex artery and during right atrial RF application due to the proximity of the RCA. Techniques avoiding creation of lesions at these critical sites, especially near the mitral annulus, should be favored to prevent arterial damage and clinical complications related to coronary artery injury.
| Acknowledgments |
|---|
| Footnotes |
|---|
| Appendix A. Conference discussion |
|---|
|
|
|---|
Dr Demaria: I think all energy would be very dangerous for the artery, and probably cryotherapy should be a good option for that when we are obliged to do ablation near the circumflex artery. We do that in clinical practice.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. C. Roberts-Thomson, D. Steven, J. Seiler, K. Inada, B. A. Koplan, U. B. Tedrow, L. M. Epstein, and W. G. Stevenson Coronary Artery Injury Due to Catheter Ablation in Adults: Presentations and Outcomes Circulation, October 13, 2009; 120(15): 1465 - 1473. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Europace, June 1, 2009; 11(6): 771 - 817. [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
S. C. Lall, S. J. Melby, R. K. Voeller, A. Zierer, M. S. Bailey, T. J. Guthrie, M. R. Moon, N. Moazami, J. S. Lawton, and R. J. Damiano Jr The effect of ablation technology on surgical outcomes after the Cox-maze procedure: A propensity analysis J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 389 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Kirsh, G. J. Gross, S. O'Connor, R. M. Hamilton, and Cryocath International Patient Registry Transcatheter cryoablation of tachyarrhythmias in children: Initial experience from an international registry J. Am. Coll. Cardiol., January 4, 2005; 45(1): 133 - 136. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Khargi, A. Laczkovics, K. Muller, and T. Deneke A possible surgical technique to avoid esophageal and circumflex artery injuries using radiofrequency ablation to treat atrial fibrillation Interactive CardioVascular and Thoracic Surgery, June 1, 2004; 3(2): 352 - 355. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |