Eur J Cardiothorac Surg 2003;23:925-934
© 2003 Elsevier Science NL
a Division of Cardiovascular Surgery, Mayo Clinic and Foundation, Rochester, MN, USA
b Anastomotic Technology Group, St. Jude Medical, Inc., Minneapolis, MN, USA
Received 3 September 2002; received in revised form 24 February 2003; accepted 3 March 2003.
* Corresponding author. Tel.: +1-507-255-8191; fax: +1-507-255-7378
e-mail: zehr.kenton{at}mayo.edu
| Abstract |
|---|
|
|
|---|
Key Words: Off-pump coronary artery bypass Minimally invasive surgical procedures Surgical anastomosis Suture techniques Surgical equipment
| 1. Introduction |
|---|
|
|
|---|
A first generation of proximal aortosaphenous vein graft connectors (Symmetry Bypass System Aortic Connector®, Anastomotic Technology Group, St. Jude Medical, Inc., Minneapolis, MN) has been used clinically in over 30 000 anastomoses performed throughout Europe and North America. Immediate results of proximal vein graft anastomoses have demonstrated rapid connector deployment (less than 10 s) with instantaneous hemostasis in 62 of 65 anastomoses [1]. The system failures were easily corrected by device removal and sutured revision. All grafts were patent at procedure's end. Long-term results are pending.
Likewise, promising results have been attained with the first generation distal SVG-to-coronary artery connector. In pre-clinical animal studies, a comparison of this device versus suture (n=12) in an on-pump canine model showed no significant inter-group differences in vessel diameters, intraoperative graft flows, graft patency, and histology following 30-day sacrifice [2]. No mechanical failures occurred, and immediate hemostasis was obtained at all connector-fashioned anastomoses. Mean time for anastomotic performance with connectors was 3 min compared to 8.4 min for sutures (P<0.0001). In a longer-term chronic study, a similar dog model found comparable results following 30-, 90-, and 180-day survival [3]. All grafts were patent on presacrifice angiography and there were no significant histologic differences between connector and sutured anastomoses at all time points. Remarkably, intimal overgrowth of connectors did not appear to substantially progress over time when comparing survival groups. One significant downside to the first generation distal connector was the requirement to load the vein onto the connector using a 10x operating microscope. This proved cumbersome and was time consuming.
Evolution in distal connector technology has produced a second generation, easy-load, 2 mm internal diameter (ID), stainless steel, anastomotic device with design improvements that facilitate rapid connector loading under minimal magnification (2x). Size reduction should allow application to most distal coronary vessels currently grafted. The purpose of this study was to evaluate this novel second generation distal connector in two distinct experimental models. First, to examine the long-term patency of the bypass graft in a standard on-pump chronic canine model, and second, to determine the acute feasibility of an off-pump SVG-to-LAD bypass performed with both proximal and distal connector devices.
| 2. Materials and methods |
|---|
|
|
|---|
|
At the appropriate target site in the bypassed vessel, an arteriotomy is made with the same preformed cutting device and dilator. The delivery system is then introduced axially through the arteriotomy until the new graft ostium intimately apposes the coronary artery wall (Fig. 2A) . With the artery intubated, the nosecone is advanced uncovering the internal fingers, and the delivery system is repositioned perpendicular to the long axis of the coronary artery. The balloon is then inflated, radially expanding the connector to a uniform cylindrical configuration (Fig. 2B). Following connector expansion, the delivery system is removed and the distal vein graft stump ligated sufficiently close to the anastomosis to avoid a large cul-de-sac without compromising the vein-anastomosis lumen (Fig. 2C).
|
2.2. Surgical procedure
Chronic distal connector studies were performed in conjunction with the Experimental Surgical Services Laboratory, University of Minnesota, Minneapolis, MN. Acute off-pump feasibility studies were performed in conjunction with the Division of Cardiovascular Surgery, Mayo Clinic, Rochester, MN.
Thirty-three adult male canine subjects (weight 2634 kg) were used, eighteen for chronic studies and fifteen for acute. All animals were obtained from licensed suppliers and treated in accordance to the principles stated in the Guide for the Care and Use of Laboratory Animals published by the National Institute of Health (NIH publication 85-23, revised 1985). The Animal Care and Use Committee of Mayo Clinic and Mayo Foundation and the University of Minnesota approved all procedures and protocols. Mayo Clinic and the University of Minnesota comply with the Animal Welfare Act of 166 (PL 89544) and the Animal Welfare Act amendment of 1985 (PL 99198).
Animals underwent induction of anesthesia with sodium thiopental (15 mg/kg to effect). After orotracheal intubation, the animals were ventilated mechanically and anesthesia maintained with halothane (11.5%). The electrocardiogram and blood pressure through a femoral arterial line were monitored continuously throughout the procedure. Chronic animals received Cefazolin (1 gm IV) during induction of anesthesia. Initial muscle relaxation was obtained with a solitary injection of pancuronium bromide (0.050.1 mg/kg).
Procedures were performed under sterile conditions in chronic animals. Animals were placed in the supine position, and both saphenous veins were harvested. The proximal and distal outer diameter (OD) of the veins were measured. To facilitate loading of the distal connector device, veins with a minimum OD of 4 mm were chosen as grafts. Veins were loaded with both proximal and distal connectors in acute studies and with distal connectors in chronic studies. After animals were repositioned, the heart was exposed through a left fourth intercostal space thoracotomy, and systemic anticoagulation was obtained with heparin (300 U/kg).
In chronic animals, an arterial perfusion cannula was placed in the right femoral artery and a venous return cannula was inserted through the right atrial appendage for extracorporeal circulation. During normothermic cardiopulmonary bypass, fibrillatory arrest was induced to obtain a still field. The middle third of the LAD was dissected, and the OD of the vessel measured. Device modifications of the second generation distal connector allow anastomoses to coronary vessels with an ID of 2 mm or greater; therefore, a region of the LAD with an OD of 2.5 mm (estimated ID of 2 mm) was chosen as the anastomotic site.
Proximal and distal 4-0 Prolene (Ethicon, Somerville, NJ) snares were placed to control bleeding through the coronary artery during construction of anastomoses. Hearts were cardioverted following completion of distal anastomoses. Aortosaphenous vein graft anastomoses were performed with running 6-0 Prolene suture.
Chronic animals were weaned off bypass, and once normal hemodynamics were obtained, flow through the grafts were measured using a Doppler flow probe (Small Animal Blood Flow Meter T-206, Transonic Systems, Inc., Ithaca, NY) before and after ligating the native coronary arteries proximal to the graft anastomosis. Protamine sulfate (1 mg for each 100 U of heparin) was then given, and meticulous hemostasis was achieved. The pericardium was approximated, a small drainage tube was placed in the left pleural space, and the wound was closed in layers. After the animal satisfactorily recovered form anesthesia, the chest tube was removed.
Acute animals maintained their own native circulation off-pump throughout the procedure. The middle third of the LAD was dissected and proximal and distal 4-0 Prolene snares were placed as in chronic animals to control bleeding through the coronary artery during anastomosis construction. The upper descending aorta was then partially occluded with a side-biting clamp and the aortosaphenous vein graft anastomoses performed with the proximal connector system. SVG-to-LAD anastomoses were then performed on the beating heart with the distal connector and flows obtained as in the chronic animals.
2.3. Postoperative care
Chronic animals received Cephazolin (1 gm IV) every 8 h for the first 24 h. Buprenorphine Hydrochloride (0.005 mg/kg intramuscularly) was used for pain control. Aspirin (325 mg orally) and Dypiridamol (20 mg orally) were given once a day until the day of sacrifice.
2.4. Conclusion of survival period
Chronic animals were subdivided into three survival groups and maintained for a minimum 30, 90, or 180 days prior to sacrifice. At the end of the survival period, chronic animals underwent general anesthesia as during the surgical procedure and a coronary angiogram was obtained with standard technique. With the animal heparinized (10 000 units IV) and still under deep general anesthesia, an overdose injection of pentobarbital followed by 40 mEq of potassium chloride was given.
Following euthanasia, the heart, bypass grafts, and thoracic aorta were excised en bloc and pressure-perfused at physiologic pressures with 10% neutral-buffered formalin for a minimum of 2 h. The specimens were then sent for gross and microscopic evaluation. Sub-gross photographs of the distal connector were prepared. Sections from the proximal and distal anastomoses, vein grafts, and coronary arteries immediately distal to the anastomoses were examined.
The connectors were processed and embedded in methylmethacrylate, cut using a diamond band saw (ExaktTM System), and ground and surface stained with hematoxylin and eosin (H&E). The remainder of tissues were processed through graded alcohols, embedded in paraffin, and stained with H&E, Masson's trichrome, and Verhoeff's elastic stains. The entire myocardium was examined grossly for the presence of infarction.
2.5. Statistical analysis
Continuous data are presented as mean plus or minus the standard deviation and range where indicated. Distal connector device loading and deployment times and graft blood flow following native LAD ligation were compared between off- and on-pump studies using Student's t-test (JMP Application 4.0.4, SAS Institute, Inc., Cary, NC). The incidence of intraoperative device-related complications for distal connectors, including device mechanical failure, leak, or rupture were compared between off- and on-pump procedures using Fisher's exact test. Two-tailed P values less than or equal to 0.05 were considered statistically significant. Chronic graft angiographic patency and distal anastomosis patency as determined from pathologic examination were descriptively reported.
| 3. Results |
|---|
|
|
|---|
|
|
|
All chronic animals survived to the minimum anticipated date of sacrifice (n=8, n=5, and n=5 for 30, 90, and 180 days, respectively) (Table 3). Angiography performed prior to euthanasia demonstrated patent grafts in all cases (Fig. 4) . All distal anastomoses were photographed after opening the opposite coronary artery wall longitudinally; representative photographs from each survival group are displayed in Fig. 5 . Regardless of survival period, distal anastomoses were widely patent on gross examination in all animals except one animal within the 180-day survival group. In this case (animal number 17), the anastomosis was mild-to-moderately stenotic (5 mm ID diameter approximately), and the connector appeared compressed.
|
|
|
| 4. Discussion |
|---|
|
|
|---|
Unlike previous preclinical investigations evaluating these devices, the acute phase of this study demonstrated the feasibility of a single SVG-to-coronary artery bypass on the beating heart, employing off-pump techniques, using the most current connector prototypes to fashion both graft anastomoses. Proximal anastomoses were performed on average in 56 s with no mechanical failures or anastomotic leaks. Average time for distal anastomotic completion was substantially longer, 194 s, but this reflects differences in connector design and delivery technique, not an increase in technical difficulty with anastomosis to a moving target. Indeed, off-pump distal anastomoses were performed just as rapidly as those created on-pump with no significant difference in deployment time noted between the two groups [P=0.86]. More importantly, graft function was not sacrificed in this model as mean Doppler blood flow through these conduits following native LAD ligation proved to be equivalent to their on-pump counterparts [P=0.65]. All the animals were hemodynamically stable during flow measurements and the blood pressure was not supported by inotropes. These flows are very similar to our previously published data comparing a hand sewn anastomosis to a larger 3 mm ID first generation distal device created anastomosis [2].
However, we did observe a trend towards an increased incidence of distal connector device-related complications during off-pump procedures [P=0.08], which we defined as mechanical failure or anastomotic leak or rupture. Two leaks developed immediately following device deployment; both were adequately controlled employing a single suture without disturbing the integrity of the anastomosis. In one instance, difficulty was encountered withdrawing the delivery system, resulting in irreparable damage to the connector. Simply removing the first device and replacing it with a second easily salvaged this anastomosis. This graft was successfully and rapidly loaded with a new connector and delivery system within the chest without disrupting the proximal anastomosis. We recognize that creating an anastomosis in a mobile field presents a technical challenge that may have been pivotal in these events, however, it is anticipated these results will improve once more experience is gained with these devices in off-pump procedures. In all three cases, intraoperative graft flow was not substantially affected, falling within one standard deviation of the off-pump group average.
Design modifications of this second generation, stainless steel distal connector permitted anastomoses to smaller caliber vessels and improved the ease of device loading compared to its predecessor, while maintaining the advantages of side-to-side technique, including distal graft ostium uniformity and avoidance of trauma to the endothelium within the graft body. Target sites on the LAD were chosen specifically for an ID of 2 mm in all cases, a downsizing of 0.5 mm from the earlier prototype [2]. We anticipate future generations of distal connectors will accommodate coronary artery diameters of 1.5 mm.
Easy load capabilities resulted in a sizable improvement in graft loading time, from 8.5 min (range 416 min) reported for the first generation device [2] to 1.5 min (range 0.64.1 min) for procedures in both groups, and loading time was similar whether off- or on-pump techniques were employed [P=0.69]. Loading of the first generation distal device proved cumbersome. It required loading on the back table under microscopic magnification by an experienced assistant during exposure of the heart. However, the improved design made it possible for loading to be completed by the operating surgeon within the chest while minimally prolonging graft completion off-pump or aortic cross-clamp time in the on-pump procedures. In fact, the ability to re-load the graft rapidly within the chest without disrupting a previously fashioned proximal anastomosis was demonstrated in the one case in which a damaged distal connector was replaced.
Long-term graft patency appeared excellent at all time intervals evaluated during the chronic phase of this study. Postoperative angiography demonstrated widely patent distal anastomoses. Effectively, the side-to-side anastomoses became smoothly tapered end-to-side grafts at 30, 90, and 180 days. There was no evidence of encroachment by thrombus from the blind end of the ligated vein graft. Pathologic examination confirmed these results, revealing a thin-to-moderate layer of neointima covering the endolumenal device surfaces. No substantial inflammatory or foreign body reaction to the stainless steel connectors was identified. Only one anastomosis was discovered to be stenotic on gross inspection, but the connector appeared ovoid and elongated as if externally compressed. This finding likely represented an artifact produced by damage inflicted to the connector during graft harvesting as the corresponding postoperative angiogram showed a widely patent anastomosis. Overall, graft patency was essentially unchanged from that seen with the first generation distal connector [2], indicating that design modifications have not negatively altered device function.
All chronic animals received a standard prophylactic regimen of antiplatelet medications including Aspirin and Dipyridamol, prohibiting this study from drawing any meaningful conclusions regarding thrombogenicity of device anastomosis or methods for thrombus prevention. Certainly, this may become an important issue with longer periods of survival. Advances in antithrombotic management of coronary stents likely may be applicable to these connectors.
The omission of an off-pump group employing only manual suturing techniques did not allow for direct comparisons to be made between traditional and connector-fashioned anastomoses on the beating heart. However, based on our results and those published previously with similar devices, in an off-pump setting we would anticipate anastomoses created with connectors to provide a substantial time-savings for bypass grafting, while proving to be more easily reproducible, as durable, and equally functional to anastomoses fashioned with manual suturing techniques. Further study will be needed to evaluate these uncertainties.
Sutureless SVG-to-coronary artery bypass is feasible, rapid, and reproducible with on- and off-pump surgical techniques using a 2 mm ID, stainless steel, distal connector. In this model, early graft patency was 100% with either technique, while grafts performed on-pump were all patent on angiographic and pathologic examination at 30, 90, and 180 days. Few device-related complications occurred, each easily managed without compromising the graft integrity, and the incidence of these events was similar whether on- or off-pump techniques were employed. This or similar technologies may become an important addition to the armamentarium for the management of coronary artery disease, particularly in off-pump or minimally invasive approaches.
| Acknowledgments |
|---|
| Footnotes |
|---|
1 Dr Bonilla and Messrs Berg, Cornelius, and Hindrichs disclose that they have a financial relationship with St. Jude Medical, Inc. ![]()
| Appendix A. conference discussion |
|---|
|
|
|---|
Dr Zehr: Well, I don't think that cost has been released for the distal connectors. The cost is about $450 for the proximal connectors which attach the saphenous vein to the aorta. The distal connectors, are still in an investigatory mode, and I don't think any patients have been billed for them thus far.
Dr J. Bachet (Paris, France): Isn't it paradoxical to make big efforts to develop connectors for the saphenous vein when at the same time an enormous amount of papers try to prove that only arterial conduits work?
Dr Zehr: I agree with you a hundred percent if our efforts were to end with saphenous vein to coronary artery anastomoses. However, our goal is to develop connectors for arterial grafts. This is a work in progress. We have already begun work in the experimental laboratory with arterial conduits. We have been able to connect canine internal thoracic arteries to coronary arteries. In a similar way to this presentation, we come in the end of the artery with the device and create a side-to-side and then make it a functional end-to-side anastomosis by clipping the manipulated end of the artery. I believe that we can do this safely in radial arteries as well as internal thoracic arteries and avoid injuring the intima. This can possibly play a significant role for less invasive procedures, potentially thoracoscopic placement of saphenous vein and radial conduits and in conjunction with the use of the robot for a LITA to LAD.
Mr A. Murday (Glasgow, United Kingdom): It is a slightly cynical question, but what I want to ask is what assessments can you make to show that it's equal or better than a hand sewn anastomosis? It is a genuine question. How are you going to make that assessment?
Dr Zehr: Well, we have not made that assessment clinically. We have made this assessment experimentally in a previously published study evaluating the 3-0 distal connector. We looked at the difference between a hand sewn anastomosis and the connected anastomosis. In that study the flows were identical both before and after ligating the native coronary artery using either technique. In this experiment we did not have the same hand-sewn anastomosis as an internal control as all of the experiments were a single aorta to left anterior descending anastomosis. However, we examined the anastomoses of the chronic animals in this study by angiography, gross pathological examination, and histology. Intra-operatively we evaluated flows by transonic Doppler ultrasound in all experiments. The flows were very similar to our historical hand-sewn control anastomoses. Clinically the gold standard is angiography and that is going to tell the story.
Mr Murday: Ultrasound of the animals?
Dr Zehr: Did we ultrasound these?
Mr Murday: Yes.
Dr Zehr: We did angiography on them. We did ultrasound at the time of the initial operation, and that is where the flows values came from, a transonic ultrasonic probe.
Dr G. Bloch (Paris, France): In your opinion is there a risk of stenosis in the long-term follow-up like restenosis in the stent in PTCA?
Dr Zehr: I think that whenever you put metal in any anastomosis there is concern about restenosis caused by pseudointimal hyperplasia. Clearly when we look at the first generation proximal, there have been incidences of occluded grafts and stenotic grafts, however, when we hand sew our anastomoses there is also an incidence of stenosis and occluded grafts. The longest term follow-up of a similar proximal connector device is by Dr Carlos Antona's group in Italy. They did angiography on 11 first generation proximals and found 10 of 11 to be widely patent at 1 year. The first generation distal connectors have also been shown to have excellent patency at 3 month evaluation. So I believe that the connectors can indeed have widely patent anastomoses at long-term follow-up, but clearly we need more long-term angiographic data to prove this to be an important technical advance.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. D. Protopapas Anastomotic devices for coronary bypass: lethal complications have been previously reported! Eur. J. Cardiothorac. Surg., January 1, 2004; 25(1): 145 - 145. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||