Eur J Cardiothorac Surg 2003;24:481-486
© 2003 Elsevier Science NL
a Heart Institute, Instituto do Coração, Av. Prof. Reynaldo dos Santos 27, 2795-563 Carnaxide, Portugal
b Santa Cruz Hospital, Carnaxide, Portugal
Received 30 September 2002; received in revised form 13 May 2003; accepted 26 May 2003.
* Corresponding author. Tel.: +351-214-165-900; fax: +351-214-165-918
e-mail: teresa.santiago{at}incor.pt
| Abstract |
|---|
|
|
|---|
23 mm) in ablations at 80 and 85 °C and on thicker atria (
5 mm) in ablations at 90 °C. Lesions measured 0.851.98 mm, all showed epicardial and myocardial damage but none were transmural. Mitral patients: Lesions measured 0.383.25 mm and 13/25 induced at 70 °C, 2/8 at 80 °C, 1/4 at 85 °C and 0/1 at 90 °C were confined to the epicardium leaving the myocardium undamaged. The remaining had damage of the epicardium and of variable portions of the myocardium, and three were transmural. Conclusions: The application temperature and the intra-tissue temperature are not the sole factors that determine lesion depth. The thickness and the composition of the epicardium and of the myocardium are major determinants in the formation of the lesion.
Key Words: Radiofrequency ablation Tissue temperature Lesion histology
| 1. Introduction |
|---|
|
|
|---|
There are several publications on electrode/tissue temperature using in vitro and in vivo experimental [24] and theoretical models [5,6] simulating percutaneous approaches, as well as in vivo/in vitro experiments of epicardial radiofrequency (RF) ablation in animals [7,8]. However there is, to our knowledge, no data relating intra-tissue temperatures with histological assessment of the lesions induced by surgical epicardial RF applications in human atrial myocardium.
We studied the temperatures measured sub-endocardially and sub-epicardially under different settings of radiofrequency epicardial applications and their relation with the thickness of the atrial wall. We used fragments of human atrial myocardium in our experimental model. Similar applications were performed in selected mitral patients with concomitant atrial fibrillation submitted to valve surgery and bilateral isolation of the pulmonary veins with endocardial RF ablation. The various lesions were histologically studied to evaluate acute tissue alterations induced by the RF application, to measure the lesion depth and assess transmurality.
| 2. Materials and methods |
|---|
|
|
|---|
The atrial tissue was mounted with its endocardial surface against the lateral opening of a custom built bath filled with circulating saline solution kept at 37 °C. In order to simulate conditions found in surgical epicardial RF ablation on a beating heart a roller pump was used to keep the saline solution circulating with a debit of 3 l/min. Radiofrequency currents were delivered to the tissue between the catheter placed over the three points marked on the epicardial surface and the dispersive electrode placed underneath the bath. We used a malleable Thermaline probe connected to a radiofrequency generator [9] (Boston Scientific, San Jose, CA) to perform epicardial ablations at set temperatures of 80 (n=10), 85 (n=12) and 90 °C (n=6) for 2 min.
Twelve samples from RF applications at 80, 85 and 90 °C were histologically assessed.
2.2. RF ablations in mitral patients
RF epicardial applications at set temperatures of 70 (n=25), 80 (n=8), 85 (n=4) and 90 °C (n=1) were performed for 2 min on the edge of the LA incision in front of the right pulmonary veins of 38 patients (aged 3373 years (58.3±10.9) of which 27 were females) with concomitant atrial fibrillation submitted to mitral valve surgery. The epicardial ablations were always performed under normothermia on a beating heart during partial cardiopulmonary bypass (CPB). The heart was arrested after the epicardial ablation and fragments of tissue were removed from the zone of ablation of the 38 patients and were observed histologically.
Radiofrequency endocardial applications were then performed at a set temperature of 70 °C for 2 min in all patients to achieve bilateral isolation of the pulmonary veins as previously described by the authors [10,11].
2.3. Histopathological assessment
The samples were fixed in 10% buffered formalin, and fragments were serially taken from the whole line of the sample (containing a fragment of the RF induced lesion) in sections that were perpendicular to the line and included the whole thickness of the atrial wall. After paraffin embedding 2 µm cuts were stained with histochemical dyes haematoxylin/eosin, Gomori's trichrome and elastic van Gieson. The sections were analyzed under light microscopy by two observers. A metric eyepiece (with a precision of 0.02 mm) was used to measure the thickness of the left atrial wall, of its layers and of the lesions caused by radiofrequency.
2.4. Data analysis
Throughout the text values pertaining to patients age and temperatures are presented as average±standard deviation. The latter were computed from the instant a temperature steady state had been reached (approximately 1015 s after the beginning of the RF ablation) until the end of the ablation.
Lesion depth is presented as average±standard deviation or average±range/2 whenever sample size was small (n
5).
Atrial wall thickness is presented as average±measurement error. The measurement error was taken as the error of the digitizer or the standard deviation whichever was the largest.
| 3. Results |
|---|
|
|
|---|
23 mm) at 80 and 85 °C (Fig. 1)
and on thicker atria (
5 mm) in ablations at 90 °C.
|
|
None of the 12 lesions histologically assessed were transmural despite the fact that in five of them the sub-endocardial temperatures reached values of and above 50 °C (55.9±6.5 °C).
3.2. RF ablations in mitral patients
Histologically the lesions showed similar features to the in vitro ones, but in mitral patients (Fig. 2
left) the myocardial interstitium at the damaged area contained hemorrhagic foci and thrombosis of the small vessels (Fig. 2 right).
|
|
|
| 4. Discussion |
|---|
|
|
|---|
4.1. RF ablations in vitro
Our results show that for the same ablation temperature the sub-endocardial temperatures are usually (but not consistently) higher in ablations performed on thinner walls than on thicker walls. Moreover, on atrial walls of similar thickness the sub-endocardial temperatures tend to increase with the set ablation temperature but not consistently. In other words, the sub-endocardial temperatures do not depend uniquely on the atrial wall thickness and on the ablation temperature. This is in agreement with our findings concerning endocardial ablations in human tissues [12].
Histology showed that there is a large variability associated to the depth of the epicardially induced lesions. Although the deepest lesions tend to occur at higher application temperatures, it cannot be stated that the average depth of the lesions increases with the application temperature. It seems that the ablation temperature and the intra-tissue temperatures are not the only factors that influence the lesion depth. In fact, sub-endocardial temperatures above 50 °C were measured in five epicardial applications without achieving transmurality. We encountered a similar situation in previous in vitro studies of endocardial ablations on human tissues [12] and again this raises the issue of whether 50 °C as the temperature threshold between viable and non-viable tissue [2,13,14] is true for all biological tissues. It is worth noting that all the referred studies were performed in healthy canine ventricular myocardium and rabbit skeletal muscle and not in human myocardial tissue. Further studies are needed to clarify this issue.
It might be argued that the temperatures opposite the application surface were incorrectly measured during the epicardial applications that led to the non-transmural lesions. We have no reason to believe that this was the case. The very thin thermocouples were always visible under the endocardium and were inserted at three different points approximately 5 mm apart (to measure the tissue temperature under the area ablated by the same electrode) in order to compensate for the inevitable imprecision in their insertion.
4.2. RF ablations in mitral patients
The thickness of the normal human left atrial wall, and the proportion of its three layers, vary from person to person, depending on age and on related diseases but its average value is approximately 3 mm. In the majority of patients with mitral valve pathology the endocardium is thicker than in normal subjects due to a higher content in elastic fibers, in collagen, in fat content and to eventual smooth muscle cells hyperplasia. The thickness of the myocardium is also highly variable, due to myocite hypertrophy and fibrosis or lipomatosis of the interstitium. The thickness of the epicardium depends on its fat content.
Histology showed that RF epicardial applications in mitral patients at 70, 80 and 85 °C led, in each case, to myocardial lesions of variable depth:
Half of the ablations at 70 °C produced lesions that were confined to the epicardium (Group A, Table 2), whereas the remaining cases produced lesions that damaged the epicardium and variable portions of the myocardium (Group B, Table 2). It is interesting to note that the average thickness of the epicardium pertaining to samples from the two groups was similar. Furthermore, the two deepest myocardial lesions did not occur in walls with the thinnest epicardium as might be expected. In one of them the epicardium measured 1.00 mm and in the other one it was twice as thick. The depth of the myocardial lesion does not depend solely on the thickness of the epicardium.
In applications at 80 °C only one quarter of the samples showed lesions confined to the epicardium leaving the myocardium undamaged, and one of the remaining lesions was transmural.
At 85 °C a similar proportion of the lesions was confined to the epicardium and one of the remaining lesions was also transmural. Again there is no correlation between the depth of the myocardial lesion and the thickness of the epicardium.
Similarly to the in vitro applications, it cannot be stated that increasing the application temperature will increase the average depth of the myocardial lesion. Nevertheless, in mitral patients the number of lesions confined to the epicardium appears to decrease when the temperature of RF application is raised.
When lesions obtained in vitro are compared with the ones similarly obtained in mitral patients limitations of the in vitro model, i.e. the lack of microcirculation and the differences in blood and saline flow must be taken into account. Ablations in patients were performed under partial CPB meaning that the saline flow in vitro and its cooling effect were probably higher than the blood flow through the atrium. This may be the reason why transmurality was never achieved in vitro.
Another important difference between the in vitro model and the in vivo results is that the former used atrial tissue from organ donors without previous pathology, other than age-related alterations, whereas the in vivo results came from mitral patients that had disease atrial walls. This may explain why all the in vitro lesions damaged the whole epicardium plus variable portions of the myocardium, whereas a significant number of lesions in mitral patients were confined to the epicardium.
Our findings suggest that the thickness and the composition of the epicardium and of the myocardium play an important role in the formation of the myocardial lesion induced by epicardial RF applications and that this may account for the large variability in the clinical results reported by different groups [1517]. Since the histological content varies from patient to patient it is logical to assume that this feature is one of the reason for such variability. From a biophysical point of view, it is likely that atrial tissues with different compositions will have different electrical properties, which will affect the conduction of RF electrical current in the tissue. For the aforementioned reasons areas of epicardial fat should be removed or avoided during RF epicardial ablation.
When we compare the depth of the epicardially induced lesions in mitral patients at 80, 85 and 90 °C (Table 2) with that of the endocardially induced lesion at 70 °C (1.06±0.46 mm) also in mitral patients [12] it becomes apparent that the depth of the lesions is similar. It may therefore be argued that, from the point of view of the lesion depth, it does not matter whether the ablation is performed endocardially or epicardially. However, because the endocardium of mitral patients is often very thickened it is sometimes more likely to obtain a deep lesion epicardially than endocardially. An example of such a situation in which transmurality was achieved with an epicardial ablation is illustrated in Fig. 3. Had we attempted to induce such a lesion endocardially it is unlikely that transmurality would have been achieved due to the very thickened endocardium.
Conversely, in re-operated patients the epicardium may be so thickened that an epicardial ablation will have little chance to be transmural. Although the future lies on epicardial ablations on beating heart these considerations will have to be taken into account.
The development of a definitive treatment of atrial fibrillation with radiofrequency requires that the ideal ablation lines be defined. Until transmurality is consistently achieved a scientific comparison between surgical techniques will remain a clinical challenge.
| Acknowledgments |
|---|
| Footnotes |
|---|
| Appendix A. Conference discussion |
|---|
|
|
|---|
So if we now have the experience that at least with various energy sources we don't reach maybe not a true linear transmural defect in the myocardium, then we should look out for the easiest or simplest ablation line, because then we at least are sure that we, for example, exclude the pulmonary veins in this box technique as it is now cited. Do you think that is reasonable?
Dr Melo: I don't have the same opinion, because, there is not a relation with length unless you are using too long energy sources, which is not the case for any tool that is on the market. The depth is one issue, the length is another because the energy delivered is really proportional to the length of the tool you are using.
Dr Moritz: But the less perfect your transmurality is the more chance you have that it doesn't work.
Dr Melo: Because I believe the minimal lesion set is bilateral isolation of the pulmonary veins, I should agree with you, but we have to go further because there are many patients, at least 30% of patients, in whom that is not enough.
Dr F. Mohr (Leipzig, Germany): I have a question and a remark. You maybe are the only one who can talk the same language in between all of surgeons and electrophysiologists, and I would like to make a comment, and it is a comment of your friend, Robert Dion, and I absolutely agree. Talking about your in vitro tests right now, others have done similar testing, comparing two radiofrequency probes, comparing the same probes you were using, and it was clear that contact to the tissue makes a major difference, and as soon as you don't have close contact, it varies very much. So if you are transmural, how soon the epicardial or transmural temperature is reached, number one. Can you comment on that?
And number two, I just want to have your comment. Should the surgeons really think about a different concept as compared to the electrophysiologists, who have agreed upon a linear lesion, more or less, which is connected to the mitral valve? Should we go ahead and do something else as the electrophysiologists have defined in the catheter laboratory? Please comment on that.
Dr Melo: Answering to your first question, I fully agree that the contact pressure in between the probe and tissue is a major determinant of tissue temperatures. And because of the problems you have reported, we will be soon reporting experiments with varying pressures. If you make a lot of pressure, your temperatures will be rising very high, and mainly if you are near any other structures, the temperatures will spread. So pressure is really dangerous, especially if you are near a vital organ.
As for the second question, I must say that in the field we have two problems again: persistent atrial fibrillation and permanent atrial fibrillation. Until last year, cardiologists or the most advanced electrophysiologists would say that permanent atrial fibrillation was not amenable for the percutaneous approach.
The first group in the world to report outstanding results in this patients group was Carlos Pappone in San Raffaele, doing what is called the electroanatomical approach(more or less what we described 2 years ago by doing isolation of the pulmonary veins only). And he has reported in the setting of permanent atrial fibrillation without concomitant organic disease a success rate in the range of 70%. Last month at the European Society of Cardiology in Berlin, Michel Haissaguere, showing exactly the same lines but together with a connecting line to the mitral annulus, has shown exactly the same results. So it is very clear in my mind that the pulmonary veins are a major component of permanent atrial fibrillation. In paroxysmal atrial fibrillation, we have either foci or single wavelength entries.
I think that we cannot address a patient in a different way as the cardiologists do. I know, that, the concept of your lines are slightly different, but they are affecting the pulmonary veins.
I don't agree with the concept of performing the Maze operation in every patient. We have to develop a stepwise approach, There are completely different groups of patients. I think atrial fibrillation is a symptom.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Benussi, S. Nascimbene, A. Galanti, A. Fumero, E. Dorigo, V. Zerbi, M. Cioni, and O. Alfieri Complete left atrial ablation with bipolar radiofrequency Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 590 - 595. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Kiser, L. W. Nifong, J. Raman, V. Kasirajan, N. Campbell, and W. R. Chitwood Jr Evaluation of a Novel Epicardial Atrial Fibrillation Treatment System Ann. Thorac. Surg., January 1, 2008; 85(1): 300 - 303. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
P. Suwalski, G. Suwalski, N. Doll, F. Majstrak, A. Kurowski, and K. B. Suwalski Epicardial Beating Heart "Off-Pump" Ablation of Atrial Fibrillation in Non-Mitral Valve Patients Using New Irrigated Bipolar Radiofrequency Technology Ann. Thorac. Surg., November 1, 2006; 82(5): 1876 - 1879. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Melby, A. Zierer, S. P. Kaiser, R. B. Schuessler, and R. J. Damiano Jr Epicardial microwave ablation on the beating heart for atrial fibrillation: The dependency of lesion depth on cardiac output. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 355 - 360. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Jahangiri, G. Weir, K. Mandal, I. Savelieva, and J. Camm Current strategies in the management of atrial fibrillation. Ann. Thorac. Surg., July 1, 2006; 82(1): 357 - 364. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi and O. Alfieri Off-pump connection of the pulmonary veins with bipolar radiofrequency: Toward a complete epicardial ablation J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 177 - 178. [Full Text] [PDF] |
||||
![]() |
H. Aupperle and N. Doll Reply to the Editor J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 213 - 214. [Full Text] [PDF] |
||||
![]() |
S. L. Gaynor, G. D. Byrd, M. D. Diodato, Y. Ishii, A. M. Lee, S. M. Prasad, J. Gopal, R. B. Schuessler, and R. J. Damiano Jr Microwave Ablation for Atrial Fibrillation: Dose-Response Curves in the Cardioplegia-Arrested and Beating Heart Ann. Thorac. Surg., January 1, 2006; 81(1): 72 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Aupperle, N. Doll, T. Walther, P. Kornherr, C. Ullmann, H.-A. Schoon, and F. W. Mohr Ablation of atrial fibrillation and esophageal injury: Effects of energy source and ablation technique J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1549 - 1554. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi and O. Alfieri Concomitant ablation of atrial fibrillation during mitral surgery MMCTS, November 29, 2005; 2005(1129): 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Aupperle, N. Doll, T. Walther, C. Ullmann, H.-A. Schoon, and F. Wilhelm Mohr Histological findings induced by different energy sources in experimental atrial ablation in sheep Interactive CardioVascular and Thoracic Surgery, October 1, 2005; 4(5): 450 - 455. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Accord, R.-J. van Suylen, T. J. van Brakel, and J. G. Maessen Post-Mortem Histologic Evaluation of Microwave Lesions After Epicardial Pulmonary Vein Isolation for Atrial Fibrillation Ann. Thorac. Surg., September 1, 2005; 80(3): 881 - 887. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi, S. Nascimbene, G. Calori, P. Denti, Z. Ziskind, S. Kassem, G. La Canna, C. Pappone, and O. Alfieri Surgical ablation of atrial fibrillation with a novel bipolar radiofrequency device J. Thorac. Cardiovasc. Surg., August 1, 2005; 130(2): 491 - 497. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bugge, I. A. Nicholson, and S. P. Thomas Comparison of bipolar and unipolar radiofrequency ablation in an in vivo experimental model Eur. J. Cardiothorac. Surg., July 1, 2005; 28(1): 76 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi Editorial comment Eur. J. Cardiothorac. Surg., July 1, 2005; 28(1): 80 - 82. [Full Text] [PDF] |
||||
![]() |
T. J. van Brakel, G. Bolotin, L. W. Nifong, A. L.A.J. Dekker, M. A. Allessie, W. R. Chitwood Jr, and J. G. Maessen Robot-assisted epicardial ablation of the pulmonary veins: is a completed isolation necessary? Eur. Heart J., July 1, 2005; 26(13): 1321 - 1326. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. van Brakel, G. Bolotin, K. J. Salleng, L. W. Nifong, M. A. Allessie, W. R. Chitwood Jr, and J. G. Maessen Evaluation of Epicardial Microwave Ablation Lesions: Histology Versus Electrophysiology Ann. Thorac. Surg., October 1, 2004; 78(4): 1397 - 1402. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||