Eur J Cardiothorac Surg 2004;26:257-261
© 2004 Elsevier Science NL
Heart Lung Center Utrecht, University Medical Center Utrecht, Utrecht, The Netherlands
Received 5 March 2004; accepted 5 May 2004.
* Corresponding author. Address: Experimental Cardiology Laboratory, University Medical Center Utrecht (Room G02.523), Heidelberglaan 100, 3584 CX Utrecht, The Netherlands. Tel.: +31-30-250-7155; fax: +31-30-252-2693
e-mail: p.f.grundeman{at}hli.azu.nl
e-mail: utrecht.cardioresearch{at}hli.azu.nl
| Abstract |
|---|
|
|
|---|
Key Words: Anastomosis CABG Echocardiography Stenosis Ultrasound
| 1. Introduction |
|---|
|
|
|---|
In the 1980s, the potential value of epicardial ultrasound for the quantitative assessment of coronary artery luminal and wall dimensions has been demonstrated [4], as well as for coronary anastomosis visualization [5]. Recently, IVUS also showed to be a promising method to determine anastomosis dimensions [6]. With modern improvements in transducer size and technology [7,8], epicardial ultrasound may prove a helpful tool to assess the impact of construction errors on anastomotic geometry.
In pressure perfused porcine and human hearts, we investigated the epicardial ultrasound 2D presentation and geometry alteration of specific construction errors in coronary artery anastomoses. The constructed errors consisted of an oversutured toe, an oversutured heel, a cross-over and a purse string.
| 2. Materials and methods |
|---|
|
|
|---|
After construction, the IMA was cannulated to pressure perfuse (95±9 mmHg) the anastomosis with saline using a Langendorff setup. The LAD proximal to the anastomosis was snared.
2.2. Surgical procedure
All anastomoses were constructed end-to-side under an operating microscope (magnification x8, wild M680, Leica AG, Heerburg, Switzerland), using a running suture technique with a prolene 70 or 80 suture. The anastomoses were constructed with the intent either to be fully patent or to contain one of the following standardized construction errors; an oversutured toe, an oversutured heel, a cross-over or a purse string. In the porcine hearts, five of each anastomosis type were constructed, and in the post-mortem human hearts, two of each type were constructed.
The oversutured toe, oversutured heel and cross-over anastomoses were made by interlocking two suture bites on opposite sides of the arteriotomy in the toe area, the heel area or at the side, respectively. The purse string anastomosis was made by pulling heavily on the suture-ends before fashioning the suture.
2.3. Ultrasound equipment
As before [7], a commercially available, high frequency (up to 13 MHz in B-Mode), linear array mini-transducer (UST-5531, Aloka, Tokyo, Japan) with an image width of 10 mm (transducer dimensions: 15x6x9 mm) was used. The transducer was placed in a gel filled probe cover (Ultracover, International Medical Products, Inc., Zutphen, The Netherlands), for clear visualization of the anterior wall of the vessels. Measurements accurate to 0.1 mm were performed using the electronic calipers of the ultrasound system. The 2-point phantom resolution of the transducer is <0.25 mm (unpublished). Imaging was performed with an Aloka SSD 5000 Prosound ultrasound system (Aloka, Tokyo, Japan). Images were stored on a laptop to enable retrospective analysis.
2.4. Scanning technique and measurements
Before anastomosis construction, the external diameter of the unpressurized coronary artery, the IMA and the length of the arteriotomy were measured (accurate to 0.1 mm) using a ruler. All anastomoses were scanned by the same investigator (TCD) who was aware of which type of anastomosis was scanned.
In the porcine heart, the anastomosis was first delineated in longitudinal and transverse planes using B-mode imaging. Subsequently, an optimal longitudinal image (defined as anastomotic orifice and in- and outflow corner captured in one image) and transverse images in B-mode were obtained. In frozen images of the anastomosis site, measurements were performed with the electronic calipers of the ultrasound system. In the anastomosis (Fig. 1)
, the size of the anastomotic orifice (1), internal diameters of the coronary artery at the toe site (3), heel site (5) and midways toe/heel site (7) were measured in both planes perpendicular to each other. The internal vessel area (A1, A3, A5, and A7) at these sections was calculated using the ellipsoid area formula (
r1r2). The areas of the IMA (A2) and the coronary artery maximally 3 mm distal to the anastomosis (A4) and maximally 3 mm proximal to the anastomosis (A6) were calculated with the circle area formula (
r2) using the internal diameter measured in the longitudinal image (Fig. 1). The area ratios A1/A2, A1/A3, and A1/A4 were calculated to assess whether the specific construction error induced anastomotic orifice narrowing compared to IMA and target artery dimensions. The ratios A3/A4, A5/A6 and A7/mean A3;A5 were calculated to assess the impact on the outflow corner, inflow corner and the posterior wall of the coronary artery, respectively.
|
2.5. Statistical analysis
Statistical analysis was only done in the porcine group. Data are presented as mean±standard deviation (SD). One-way analysis of variance together with a post hoc comparison (Dunnett) was used to evaluate differences in the specific area ratios between the four construction error groups (oversutured toe, oversutured heel, cross-over and purse string) and the control group. Area A1 was normalized to the arteriotomy length, area A2 to the IMA diameter and areas A3 and A4 to the LAD diameter before statistical comparison of the area ratios A1/A2, A1/A3 and A1/A4 was made between the different error groups and the control group. A value of P<0.0125 was considered statistically significant.
2.6. Validation
A polyvinylsiloxane impression material (Kerr Co, Romulus, MI, USA) was injected into the anastomosis and coronary artery through the IMA. After the hardening process, the impression material was removed from the vessels and the obtained 3D cast of the anastomosis was used to validate ultrasound anastomotic geometry qualitatively.
| 3. Results |
|---|
|
|
|---|
|
|
|
At the anastomotic site, small septal perforators with diameters ranging from 0.2 to 0.7 mm were easily spotted. In all anastomoses, ultrasound geometry corresponded qualitatively with cast geometry findings, including the location of septal perforators and side-branches.
3.2. Post-mortem human hearts
External recipient coronary artery diameter ranged from 1.5 to 2.5 mm (median 2.0 mm).
In one control anastomosis, the complete outline of the anastomotic site could not be visualized due to the presence of abundant plaque with severe calcification. All constructed errors were properly visualized in longitudinal and transverse images. One oversutured toe anastomosis showed an adventitial flap waving up and down in the anastomotic orifice closing it off almost completely. This was confirmed by the cast. Presence and extent of coronary pathology was well visualized and quantifiable. In all anastomoses, ultrasound geometry corresponded qualitatively with cast geometry.
| 4. Discussion |
|---|
|
|
|---|
The four constructed errors did not influence the in- or outflow corner area ratios of the anastomoses (Table 1), in contrast to the anastomotic orifice (A1) area ratios which were significantly reduced for all the error groups except the oversutured heel (Table 2).
The 13 MHz resolution of the mini-transducer proved sufficient to visualize the complete outline and characteristics (septal perforators, coronary pathology) of the anastomotic site. Based on measurements made with the electronic calipers of the ultrasound system, calculations could be made to assess the impact of the specific construction error on anastomotic geometry.
Apart from the intentional construction errors, two unintended technical abnormalities (ridge due to extra stitch and adventitial flap) were revealed. Both unintended errors mimic clinical practice too. Thus, 13 MHz epicardial ultrasound may be a valuable diagnostic tool to assess anastomotic quality prior to chest closure. The small size of the present mini-transducer allows its application in between the suction pods of a cardiac stabilizer [7,10].
Besides the use for anastomotic geometry assessment [5,8,9], epicardial ultrasound may help to localize the target artery and choose the anastomotic site based on the evaluation of its wall thickness, plaque morphology (calcifications) and lumen diameter [4,5,10]. We expect high-frequency epicardial ultrasound to become a useful intra-operative diagnostic modality that is a non-invasive, fast, simple, and relatively inexpensive method in both on- and off-pump coronary artery bypass surgery. Angiography, currently considered the gold standard, is invasive, time consuming, expensive and not always immediately available in the operating room [1]. An additional disadvantage is the uncertain significance of early abnormal findings, which could lead to unnecessary revisions. Quantitative analysis of the angiogram is possible, but time consuming and displays far less anatomical detail of the anastomosis. Epicardial ultrasound in contrast, clearly displays the anatomy and dimensions of the anastomosis and therefore can differentiate for example between gross anastomotic construction errors, adventitial flaps and thrombus or spasm, which resolves after the immediate post-operative period.
Another widely used non-invasive method to assess bypass graft function is transit time flowmetry. However, flowmetry provides no anatomical information of the anastomosis. The observed flow depends on other factors as well. A crucial limitation is that flowmetry will only detect a severely stenosed anastomosis (>75%) [11].
Next to flowmetry, ultrasound can easily be employed as a complementary quality control method to provide information about bypass graft function and anastomotic geometry, respectively. For example, epicardial ultrasound may detect an adventitial flap in the anastomotic orifice in the presence of a normal flow recording. Flowmetry, on the other hand, may detect reduced volume flow in the presence of a flawless anastomosis, indicating either a proximal anastomosis problem, graft kinking, graft spasm, or reduced distal runoff.
The limitation of this study is its laboratory setting, with pressure perfusion by saline of the bypass graft. In addition, the merits and limitations of the present transducer for detection of anastomosis construction errors in the clinical setting and in an observer blinded experimental setting remain to be established.
In conclusion, ex vivo, epicardial 13 MHz ultrasound was successful in accurately visualizing and assessing the geometry of four different construction errors in the distal coronary anastomosis. All errors reduced the anastomotic orifice area.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Di Giammarco, M. Pano, S. Cirmeni, P. Pelini, G. Vitolla, and M. Di Mauro Predictive value of intraoperative transit-time flow measurement for short-term graft patency in coronary surgery. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 468 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, R. Meijer, T. C. Dessing, C. Borst, and P. F. Grundeman Detection of construction errors in ex vivo coronary artery anastomoses by 13-MHz epicardial ultrasonography J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 1078 - 1083. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, T. C. Dessing, R. Meijer, P. F.A. Bakker, C. Borst, and P. F. Grundeman Robot-assisted 13 MHz epicardial ultrasound for endoscopic quality assessment of coronary anastomoses Interactive CardioVascular and Thoracic Surgery, December 1, 2004; 3(4): 616 - 620. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||