Eur J Cardiothorac Surg 2005;27:450-455
© 2005 Elsevier Science NL
Characteristic resistance curves of aortic valve substitutes facilitate individualized decision for a particular type
Ralf-U. Kuehnela,
Reinhard Puchnera,
Alexander Pohlb,
Max O. Wendtb,
Martin Hartrumpfa,
Manfred Pohlb,
Johannes M. Albesa,*
a Department of Cardiovascular Surgery Heart Center Brandenburg, Ladeburger Strasse 17, 16321 Bernau, Germany
b Institute of Medical Physics and Biophysics, Charite, Berlin, Germany
Received 16 September 2004;
received in revised form 18 November 2004;
accepted 1 December 2004.
* Corresponding author. Address: Herzchirurgie, Herzzentrum Brandenberg, Ladeburger Strasse 17, Bernau 16321, Germany. Tel.: +49 3338 69 4510; fax: +49 3338 69 4544. (E-mail: j.albes{at}immanuel.de).
 |
Abstract
|
|---|
Objective: Biological valves exhibit characteristic curves (CC) regarding transvalvular gradient, resistance, and effective orifice area when correlated with a physiological cardiac output range (CO). The slope of the curve of transvalvular resistance over a typical CO range characterizes the clinical performance of the valve. These information may support an individualized decision towards the most adequate valve type. Methods: In an extracorporeal mock circuit two types of stented biological aortic valves (constructed pericardial valves, Edwards Perimount: EP; porcine cusp valves, Medtronic Mosaic: MM) of 21, 23, and 25mm were investigated. Mean transvalvular gradient was measured over a range of 1.9 to 7.2l/min CO at a simulated heart rate of 70beats/min. Transvalvular resistance was calculated and presented as characteristic curves in a loglog-plot against cardiac output. Results: EP valves of all sizes demonstrated low slopes (resistance range; slope: 21mm: 5379dynesscm5; 0.29; 23mm: 4456dynesscm5; 0.12; 25mm: 3845dynesscm5; 0.12) while MM valves exhibited steep slopes (resistance range; slope: 21mm: 46169dynesscm5; 0.97; 23mm: 36 146dynesscm5; 0.95; 25mm: 2764dynesscm5; 0.68) Conclusions: While constructed pericardial valves demonstrate sufficient hemodynamic performance especially in the higher CO range porcine cusp valves exhibited minor resistance in the lower CO range. Patients who exercise regularly may therefore profit from a pericardial valve while patients with a small body surface area and little exercise who therefore remain in the lower CO range may be adequately treated with a porcine cusp valve.
Key Words: Aortic valve Biological valve replacement Hemodynamics Cardiovascular pathology Cardiovascular model
 |
1. Introduction
|
|---|
Demographic changes observed in western industrialized nations have inevitably resulted in an increase of age and morbidity of patients in need for aortic valve replacement [1]. As elder patients do exhibit a considerable susceptibility regarding cerebral or intestinal hemorrhage as well as bone injuries lifelong avoidance of anticoagulation has gained renewed interest [2]. As much evidence has been accumulated determining the average patency rates of biological valves to appear in the proximity of 1012 years surgeons are increasingly inclined to offer biological valves to their elderly patients assuming that the life expectance may very well match the patency of the prosthesis while offering a reasonable quality of life wihout the inherent risks of coumarin derivatives [3]. In the routine setting stented aortic valves appear to be preferably used because they facilitate and expedite implantation thereby reducing operative risks [4,5]. Interestingly, two entirely different types of stented biological valve substitutes have been engineered and refined in the last three decades. Valves constructed from bovine pericardium aligned on an artifical frame as well as porcine valves mounted on an artificial scaffold [6]. In numerous studies it has been shown that both valve types demonstrate a comparable clinical short and long term performance particularly in the aortic position [3,69]. Although pericardial valve substitutes tend to exhibit lower transvalvular gradients the assumed differential hydraulic behaviour has not yet been sufficiently and comprehensively revealed. It can be hypothesized that the particular hemodynamic properties which can be theoretically derived from the different engineering principles may very well account for a different in vivo behaviour [10]. It can even be speculated as to whether these differences may be utilized to match the particular hemodynamic advantages with the individual requirements of the patient while diminishing the disadvantages of the particular design [11]. This study therefore intended to analyse flow-dependent hemodynamic parameters in a systematic fashion in order to derive characteristic resistance curves serving as decision tool towards an individualized indication for aortic valve replacement [10,1315].
 |
2. Materials and methods
|
|---|
2.1. Heart valve testing device and set-up
An established extracorporeal mock circuit designed for highly precise in-vitro measurements of the hemodynamic performance of heart valves and heart valve substitutes was utilized. The apparatus developed by Schichl and Affeld has been characterized in detail previously and is shown schematically in Fig. 1 [18]. Briefly, a piston-pump generated a constant stroke volume which was repeatedly ejected through the test valve. This process was controlled by a computer-driven programmable disk armature motor (F12M4, MATTKE AG, Freiburg, Germany) simulating a physiological aortic flow profile. After completion of the forward flow phase, the controller switched from flow control to pressure con-trol so that a physiological diastolic pressure gradient was maintained by a computer-controlled retrograde motion of the piston. Upon termination of diastole, a new cardiac cycle was initiated in order to generate several consecutive beats by the device.

View larger version (31K):
[in this window]
[in a new window]
|
Fig. 1. Heart valve testing device Schematic drawing of the mock-circuit according to Schichl and Affeld.
|
|
2.2. Parameter assessment and measuring protocol
Measurements during the first beat were discarded to attain a steady pulsatile state. In every experimental setting, measurements were conducted in 10 consecutive beats. The pressure difference across the valve were measured with two separate pressure transducers (PR 10, KELLER Ges. für Druckmesstechnik mbH, Jestetten, Germany). The amplification of one transducer was set according to the signal strength to achieve a higher resolution of the pressure signal during the systolic phase. The measurement of the piston displacement was accomplished with a digital angular transducer (GiO 40, TWK Messelektronik GmbH, Düsseldorf, Germany).
The resultsflow and pressure differencewere given as curves (Fig. 2), and as tables with a resolution of 1ms. The output diagram included the integrated data (see appendix): mean flow rate (MF; ml/s), root mean square flow (RMS; ml/s), stroke volume (SV; ml), simulated heart rate (HR; bpm), cardiac output (CO; l/min), mean pressure gradient (MTG; mmHg), closure time (CT; ms), closing volume (CV; ml), and leakage volume (LV; ml).

View larger version (16K):
[in this window]
[in a new window]
|
Fig. 2. Curves of flow and pressure difference of the test set-up Schematic drawing of the time curves of instantaneous systolic pressure drop and transvalvular flow. Characteristic time points in the cardiac cycle.
|
|
Data were recorded in an electronic database for further processing. A dataset consisted of at least 12 different cardiac output measurements in order to assess the flow-dependency of EOA and TR of the respective study valve.
The precision of the parameters measured was as follows: volume ±1.0%, time ±1.5%, and pressure ±1.5%. The testing device enabled investigations over a broad spectrum of test parameters. Heart rate could be adjusted within a range of 30150beats/min with variable systolic ejection period (SEP; s/min). Stroke volume ranged from 30100ml while the pressure differences across the valve resembled the physiological systolic/diastolic values of 120/80mmHg.
In this study and in accordance to the ISO and FDA standards for cardiac valve testing, measurements with aortic valves were carried out at a frequency of 70beats/min and a SEP of 23.78±0.11s/min. Isotonic saline was used as the test fluid to achieve full transparency because it has been shown by us that fluid viscosity has only little influence on hemodynamic parameters obtained in ex-vivo mock circuits [21,22].
2.3. Data processing
Off-line analysis of the recorded data was performed using a self-developed software package. Full details of calculations and formulae were already provided in detail recently [2123] (see Appendix A). Single cardiac cycles in every data set were mathematically combined to form an average beat which then served as a basis for all subsequent calculations.
2.4. Heart valve substitutes
Commercially available biological heart valve substitutes were used. Edwards Lifesciences Perimount Aortic Valves of 25, 23, 21mm size (Edwards Lifesciences, Irvine, CA) and Medtronic Mosaic Aortic Valves of 25, 23, 21mm size (Medtronic Inc., Minneapolis, MN). 19mm valves were not investigated according to our clinical policy not to use 19mm biological valve substitutes in adult patients.
2.5. Hemodynamic assessment and data acquisition
Transvalvular resistance (TR; dynesscm5) was calculated by 1333xMTG/MF. Effective orifice area (EOA; cm2) was calculated by RMS/(51.6xMTG1/2) (see Appendix A).
The characteristic curves of a valve (CC) TR versus CO were plotted as loglog-plot in a CO-range of 1.57l/min and linearized by a power function: TR=cCOß. The weight factor c and the exponent ß (the slope of the curve) were calculated by non-linear regression analysis ({c, ß+SD}=rgp(log(TR); log(CO); true; true); Excel for Windows, statistic package, Microsoft Inc, Seattle, WS). Area utilization index (AI) related to the obturator area (AI-O) was assessed as: AI-O=100EOA/OA (%); AI related to sewing ring area (AI-SR) (Performance Index acc. to Yoganathan [24]) was assessed as: AI-SR=100EOA/sewing ring area (%).
2.6. Assessment of physical orifice dimension
The physical orifice diameter (POD; mm) of the respective valve type and size was measured with a Hegar obturator (diameter sizes: 1631mm; in 1mm increments) while the outer diameter including the sewing ring was measured with a ruler. The obturator area (OA; cm2) was computed by:
(diameter/2)2 (cm2).
2.7. Data acquisition and statistical analysis
All data were compiled by means of specifically designed software with a standard personal computer. Data were organized in a calculation sheet (Excel for Windows, Microsoft Inc, Seattle, WS). Descriptive and comparative statistical analysis were performed with standard software (SPSS for Windows). Flow independent parameters were analyzed using two-sided Student's t-test in case of normal distribution or using MannWhitney test if non-normally distributed. Data are presented as mean values ±standard deviation (SD). Significant results were indicated at 5, 1, and 0.1% error level. *P<0.05; **P<0.01; ***P<0.001.
 |
3. Results
|
|---|
3.1. Hydraulic valve parameters in the physiological range dependent on valve type and size
Mean transvalvular gradients (MTG) of the EP valves were typically lower than those of MM valves. The differences, however, diminished with increasing valve size. Regurgitation time (RT) of MM valves was shorter than that of EP valves. The differences increased with valve size. Regurgitation volume (RV) and leakage volume (LV) of MM valves were smaller than those of EP valves. Again, the differences increased with valve size (Table 1).
3.2. Flow dependency of effective orifice area and resistance with respect to valve type and size
In general, a 2mm larger obturator could be passed through the valve's orifice in EP valves compared to MM valves of identical sewing ring size. Minimum computed EOA of MM valves was larger than minimum computed EOA of EP valves while maximum EOA of EP valves was larger than maximum computed EOA of MM valves. The flow-dependent increase of EOA of the MM valve was 0 (21 and 23mm) or low (25mm) while the flow-dependent increase of EOA of EP valves of all sizes was significantly higher than 0. The flow-dependent increase of TR of the MM valve was 1(21 and 23mm) or 0.7 (25mm) respectively, while the flow dependent increase of TR of EP valves of all sizes was significantly lower than 1. Small MM valves exhibited a flow-independent EOA whereas small EP valves demonstrated nearly flow-independent TR (Table 2).
View this table:
[in this window]
[in a new window]
|
Table 2. Flow-dependency of EOA and TR. Flow-range of 1.57l/min compared to the physical orifice area. Values and slope of linearized CC in loglog-plot
|
|
Transvalvular resistance of all studied valves were plotted against cardiac output. Distinct slope differences of the two investigated valve types were observed which appeared to be rather independent of the valve size. MM valves showing an almost linear curve with steep slope while EP valves exhibited a non-linear curve with almost absent slope in the low CO range while exhibiting a steady increase of slope towards higher CO values (Fig. 3).

View larger version (28K):
[in this window]
[in a new window]
|
Fig. 3. Flow-dependence of transvalvular resistance: Transvalvular resistance of all study valves is plotted against cardiac output. Note the distinct slope differences of the two investigated valve types independent of the valve size. EP valves showing an almost linear curve with steep slope while MM valves exhibit a non-linear curve with absent slope in the low CO range while exhibiting a steady increase of slope towards higher CO values.
|
|
3.3. Area utilization index
In general, area utilization index related to the obturator area (AI-O) and area utilization index related to the sewing ring (AI-SR) increased with increasing flow. The increase of AI-O and AI-SR was higher in Edwards Perimount valves than in Medtronic Mosaic valves. Particularly the AI-O diminished in the same valve type with increasing size. Both, AI-O and AI-SR appeared to be advantageous in the low-flow range of the Medtronic Mosaic valve. The differences were most pronounced in small valves (21mm). In contrast, AI-O und AI-SR were advantageous in Edwards Perimount valves of both smaller sizes (25mm) in the high-flow-range (Table 3).
View this table:
[in this window]
[in a new window]
|
Table 3. Area utilization index (AI) of the valves under low-flow LF (1,5l/min) and high-flow conditions HF (7l/min) in percent of the obturator-area/sewing ring area
|
|
 |
4. Discussion
|
|---|
During the last 15 years stented biological aortic valve substitutes have been investigated in an astounding number of publications [13,68]. It was shown that both types of biological valves demonstrate reasonable hemodynamic performance in the clinical setting [13,6]. It was also shown that the observed differences in transvalvular gradients were not followed by significant differences in left ventricular remodelling as well as clinical outcome [7]. Instead, evidence was accumulated that left ventricular muscle mass regression as one of the key indicators for adequate clinical recovery of the heart did not differ significantly [20,1617]. Investigations, however, further revealed that degeneration of biological valves appeared after a typical time frame of 815 years notwithstanding the particular valve type [3]. Fortunately, it was also demonstrated that elderly patients fared reasonably well with biological valves in terms of longevity and quality of life [6,7]. A paucity of evidence, however, has been provided, regarding a potentially different clinical behaviour of both valve types when looking at the individual situation of a patient [1,2]. The transvalvular gradient appeared almost exclusively in a variety of publications as a tool towards a differential decision. It was assumed that patients with a small annulus require a valve with the largest effective orifice area available for a given size [6,8]. As pericardial valves showed smaller transvalvular gradients when compared to porcine valves of the same size pericardial tissue valves were often proposed as the substitute of choice in an annulus of limited dimensions [5]. This, however, may be a superficial attitude. Already in 1978 Rahimtoola had inaugurated the concept of patient-prosthesis mismatch as one of the key factors for appropriate prosthesis selection [24]. Dumesnil could later show that a small decrease in the effective orifice indexed for body surface area was accompanied by large increments of transvalvular pressure gradient when looking at the increase of stroke volume during exercise. From his ex-vivo investigations he recommended the indexed effective orifice area to be not less than 0.91.0cm2/m2 for aortic prostheses [25]. The concept of effective orifice area index being the clinical equivalent for the behaviour of a valve of too small a size over the entire range of exertion the patient generates, however, has been heavily debated over the following years. While proponents repeatedly advocated to abide by this threshold the majority of clinical investigations could demonstrate that small prosthesis can indeed be implanted and that patient-prosthesis mismatch does not play as much a clinical role as assumed [6,7,9,12,19]. Small valves are not necessarily bad as demonstrated by Medailon and others [4,5,8]. The current perception of patient-prosthesis mismatch may thus be reconsidered at least for selected patient cohorts as quoted by Freed and coworkers [16].
In our study we therefore looked at the hydraulic properties of biological aortic valves over a typical and physiological range of cardiac output. We were able to prove the data in the literature and could demonstrate that pericardial valves indeed exhibit lower transvalvular gradients because of the larger effective orifice area [11]. We could furthermore support the data in the literature regarding the opening and closure characteristics of Medtronic Mosaic and Edwards Perimount valves [3]. However, in our study it became evident that the behaviour of the effective orifice area, which resembles the opening and closure characteristics of the different types of biological valves was responsible for a flow dependent hemodynamic performance [1315]. By means of a resistance/CO loglog-plot distinct slope differences were visualized showing differential behaviour of both valve types independent of the size. Our study indicated that the effective orifice area of pericardial valves is flow-dependent. This indicates that the valve does not open entirely under low-flow conditions while it can adapt to a wider cardiac output range by an effective orifice area increase. In contrast, porcine cusp valves exhibit the entire effective opening of the valve already under low-flow conditions while not showing a cardiac output dependent effective opening area reserve. The different characteristics of the valves allow therefore for an individualized choice of valve type: Patients, in whom CO-ranges in the proximity of 36l/min can be assumed, may be candiates for porcine cusp valves while patients exhibiting a larger CO range between 410l/min may consequently be candiates for pericardial valves. It was demonstrated that Medtronic Mosaic valves appear to be particularly advantageous in the smaller range of cardiac output while our data indicated that Edwards Perimount valves appear to be advantageous in the higher CO range.
4.1. Limitations of the study
Artificial models offer a wide spectrum of experimental settings while providing accurate data. Furthermore, they can reduce the number of animal experiments. However, the transfer to the in-vivo situation is limited. Aspects of the central and peripheral circulation, such as the Windkessel effect play a major role in valve behaviour but are only simulated in our model. The outflow tract of our model is straight and not naturally curved. As a consequence, the in-vivo hydraulic behaviour of a valve may differ from our idealized assumptions. Furthermore, an altered configuration of the subvalvular left-ventricular outflow tract due to the implantation procedure itself was not simulated. Isotonic saline was used in our model. In contrast, blood is a Non-Newtonian fluid exhibiting velocity dependent viscosity. Instead of blood one can utilize 30% glycerol solution which resembles the viscosity of blood. However, Werner has already demonstrated that the fluid viscosity has little influence on systolic valve parameters. By using 0.9% saline, however, pressure gradients may be slightly underestimated and leakage volumes overestimated [21,22]. Most certainly, our in vitro study can provide but first insights into the assumed potential of biological valves regarding a differential indication. In vivo studies are required to verify our initial findings and theoretical assumptions. Other biological valves fabricated with pericardial tisssue or assembled from porcine aortic valves may exhibit different properties. Although not yet proven, we however, believe that our findings may perhaps have exemplary character regarding both main constructive types.
4.2. Conclusion
In the clinical scenario rather small and fragile elderly patients who may exhibit a cardiac output of hardly more than 3l while sleeping and who do live with only moderate exertion may be best served with a porcine cusp aortic valve while elderly individuals of larger size and weight who perform demanding exercise on a regular basis such as biking or hiking may profit from a pericardial aortic valve. Characteristic curves have the potential to supplement the indexed effective orifice area (EOAI) for verification of patient-prosthesis mismatch or -in other words- to predict proper patient-prosthesis match.
 |
Appendix A. List of abbreviations, symbols, and the formula used (Fig. 2)
|
|---|
- HKP+
- heart valve tester
- OA/POD
- opening surface of the valves (Hegar)/diameter of Hegar's obturator (cm2/mm)
- Q(t)/MF/RMS
- instantaneous volumetric flow/mean flow rate/root mean square-flow rate (ml/s)
- SV
- stroke volume (Ml)
- CO
- cardiac output (l/min)
P(t)/MTG- instantaneous systolic pressure drop/mean transvalvular gradient (mmHg)
- EOA
- effective orifice area (cm2)
- ET/CT/DFP
- ejection time/time to close of the valve/diastolic filling period (ms)
- HR
- heart rate (min1)
- SEP
- systolic ejection period (s/min)
- CV
- closing volume (Ml)
- LV
- leakage volume (Ml)

- density of the fluid (g/ml)
mean flow rate MF, root mean square-flow RMS, stroke volume SV, and cardiac output CO:
valve closure volume CV, leakage volume LV:
mean transvalvular gradient MTG, transvalvular resistance TR, and effective orifice area EOA:
 |
Appendix B. Conference discussion
|
|---|
Dr H. Vanermen (Aalst, Belgium): I've got a question about your conclusions. Old ladies, most of the time, have very small annuluses. So most of the time we have the tendency to pick up a valve that's got better hemodynamic performances even at high cardiac outputs because most elderly people now are getting more and more active, too.
Dr Kuehnel: The small annulus diameter with old ladies is a point of interest, but I think then the gradients shows us that the Medtronic valves in the lower flows are better. In other studies, also, the small sizes you need in this situation take good results.
Dr Vanermen: One may anticipate more calcification to develop in the porcine valves versus the pericardial valves as well.
Dr Kuehnel: Yes. This is an in vitro study following hemodynamic parameters. It is not only one point. We have a discussion picture and you can see here, this is from nonpublished data from another study, you can see here the pericardial valve and the Medtronic Mosaic valve. And you can see here under low-flow condition the pericardial valve opens only a little bit and here it opens greater. And this could be a reason for calcification, yes.
Dr Vanermen: It's a very interesting finding.
Dr A. Wechsler (Philadelphia, PA, USA): Did you measure the orifice area of each valve?
Dr Kuehnel: Yes, we measured with a Hegar probe; and we visualized this and can planimetrate this area also.
Dr Wechsler: And did you find that a 21 is exactly 21?
Dr Kuehnel: Yes. The Medtronic Mosaic are 1mm diameter lower than the pericardial valve each size.
 |
Footnotes
|
|---|
☆ Presented at the joint 18th Annual Meeting of the European Association for Cardio-thoracic Surgery and the 12th Annual Meeting of the European Society of Thoracic Surgeons, Leipzig, Germany, September 1215, 2004.
 |
References
|
|---|
- Seitelberger R, Bialy J, Gottardi R, Seebacher G, Moidl R, Mittelböck M, Simon P, Wolner E. Relation between size of prosthesis and valve gradient: comparison of two aortic bioprosthesis. Eur J Cardiothorac Surg 2004;25:358-363.[Abstract/Free Full Text]
- Chaudhry MA, Raco L, Muriithi EW, Bernacca GM, Tolland MM, Weathley DJ. Porcine versus pericardial bioprostheses: eleven-year follow up of a prospective randomized trial. J Heart Valve Dis 2000;9:429-438.[Medline]
- Reichenspurner H, Weinhold C, Nollert G, Kaulbach HG, Vetter HO, Boehm DH, Reichart B. Comparison of porcine biological valves with pericardial valvesa 12 year clinical experience with 1123 bio-prostheses. Thorac Cardiovasc Surg 1995;43:19-26.[Medline]
- Medalion B, Lytle BW, McCarthy PM, Stewart RW, Arheart KL, Arnold JH, Loop FD, Cosgrove 3rd DM. Aortic valve replacement for octogenarians: are small valves bad?. Ann Thorac Surg 1998;66:699-705[discussion 7056].[Abstract/Free Full Text]
- Medailon B, Blackstone EH, Lytle BW, White J, Arnold JH, Cosgrove DM. Aortic valve replacement: is valve size important?. J Thorac Cardiovasc Surg 2000;119:963-974.[Abstract/Free Full Text]
- McDonald ML, Daly RC, Schaff HV, Mullany CJ, Miller FA, Morris JJ, Orszulak TA. Hemodynamic performance of small aortic valve bioprostheses: is there a difference?. Ann Thorac Surg 1997;63:362-366.[Abstract/Free Full Text]
- Vermeulen F, Bennink G, Ernst S, Jaarsma W, Chevalier PA, Lutz DV. The intact porcine bioprosthesis: early world-wide clinical experience and analysis of a single institution's experience. Eur J Cardiothorac Surg 1992;6(Suppl 1):S124-S127.
- Gelsomino S, Morocutti G, Masullo G, da Col P, Frassani R, Spedicato L, Livi U. Patient-Prosthesis mismatch after small-size stentless aortic valve replacement. J Card Surg 2004;19:91-97.[CrossRef][Medline]
- Frapier JM, Rouviere P, Razcka F, Aymard T, Albat B, Chaptal PA. Influence of patient-prosthesis mismatch on long-term results after aortic valve replacement with a stented bioprosthesis. J Heart Valve Dis 2002;11:543-551.[Medline]
- Eriksson MJ, Rosfors S, Radegran K, Brodin LA. Effects of Exercise on Doppler-derived pressure difference, valve resistance, and effective orifice area in different aortic valve prostheses of similar size. Am J Cardiol 1999;83:619-622.[CrossRef][Medline]
- Marquez S, Hon RT, Yoganathan AP. Comparative hydrodynamic evaluation of bioprosthetic heart valves. J Heart Valve Dis 2001;10:802-811.[Medline]
- Pibarot P, Dumesnil JG, Cartier PC, Metras J, Lemieux MD. Patient prosthesis mismatch can be predicted at the time of operation. Ann Thorac Surg 2001;71:S265-S268.[Abstract/Free Full Text]
- Pohl M, Meyer R, Kuehnel R, Talukder NK, Wendt MO. Different types of aortic stenosis and simulation of their morphological-hydrodynamic interdependencein vitro study with allografts and stenotic valve models. Int J Artif Organs 2001;24:870-877.[Medline]
- Wendt MO, Pohl M, Werner S, Kühnel R, Toppel D. Differentiated assessment of heart valve stenoses by the expanded Bernoulli equationan in vitro study of stenosis models. Z Kardiologie 1995;84:686-693.
- Burwash IG, Hay KM, Chan KL. Hemodynamic stability of valve area, valve resistance, and stroke work loss in aortic stenosis: a comparative analysis. J Am Soc Echocardiogr 2002;15:814-822.[CrossRef][Medline]
- Freed DH, Tam JW, Moon MC, Harding GE, Ahmad E, Pascoe EA. Nineteen-millimeter prosthesic aortic valves allow normalization of left ventricular mass in elderly patients. Ann Thorac Surg 2002;74:2022-2025.[Abstract/Free Full Text]
- Albes JM, Hartrumpf M, Rudolph V, Krempl T, Hüttemann E, Vollandt R, Wahlers Th. Are mechanical valves with enhanced inner diameter advantageous in the small sized aortic annulus?. Ann Thorac Surg 2003;76:1564-1570.[Abstract/Free Full Text]
- Schichl K, Affeld K. A computer controlled versatile pulse duplicator for precision testing of artificial heart valves. Int J Artif Organs 1993;16:722-728.[Medline]
- Zeitani J, Bertoldo F, Nardi P, Polisca P, de Paulis R, de Peppo AP, Chiariello L. Influence of patient-prosthesis mismatch on myocardial muscle mass regression and clinical outcome in physically active patients after aortic valve replacement. J Heart Valve Dis 2004;13(Suppl 1):S63-S70.
- Hanayama N, Christakis GT, Mallidi HR, Joyner CD, Fremes SE, Morgan CD, Mitoff PR, Goldman Bs. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thorac Surg 2002;73:1822-1829[discussion 1829].[Abstract/Free Full Text]
- Werner S, Wendt O, Schichl K, Pohl M, Koch B. Testing the hydrodynamic properties of heart valve prostheses with a new test apparatus. Biomed Tech (Berl) 1994;39:204-210.
- Pohl W, Wendt MO, Werner S, Koch B, Lerche D. In vitro testing of artificial heart valves. Comparsion between Newtonian and non-Newtonian fluids. Artif Organs 1996;20:37-46.[Medline]
- Hartrumpf M, Albes JM, Krempl T, Rudolph V, Wahlers T. The hemodynamic performance of standard bileaflet valves is impaired by a tilted implantation position. Eur J Cardiothorac Surg 2003;23:283-291.[Abstract/Free Full Text]
- Dumesnil JG, Yoganathan AP. Valve prosthesis hemodynamics and the problem of high transprosthetic pressure gradients. Eur J Cardiothorac Surg 1992;6(Suppl 1):S34-S37.
- Rahimtoola SH. The problem of valve prosthesis-patient mismatch. Circulation 1978;58(1):20-24.[Abstract/Free Full Text]
This article has been cited by other articles:

|
 |

|
 |
 
T. Bottio, V. Tarzia, G. Rizzoli, and G. Gerosa
The changing spectrum of bioprostheses hydrodynamic performance: considerations on in-vitro tests
Interactive CardioVascular and Thoracic Surgery,
October 1, 2008;
7(5):
750 - 754.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Bakhtiary, O. Dzemali, U. Steinseiffer, C. Schmitz, B. Glasmacher, A. Moritz, and P. Kleine
Opening and closing kinematics of fresh and calcified aortic valve prostheses: An in vitro study
J. Thorac. Cardiovasc. Surg.,
September 1, 2007;
134(3):
657 - 662.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio
Reply to the Editor
J. Thorac. Cardiovasc. Surg.,
March 1, 2007;
133(3):
847 - 849.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio
Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves
J. Thorac. Cardiovasc. Surg.,
May 1, 2006;
131(5):
1058 - 1064.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. H. Rahimtoola
The Year in Valvular Heart Disease
J. Am. Coll. Cardiol.,
January 17, 2006;
47(2):
427 - 439.
[Full Text]
[PDF]
|
 |
|