Eur J Cardiothorac Surg 2005;28:701-704
© 2005 Elsevier Science NL
Department of Cardiovascular Surgery, University Hospital, 3010 Berne, Switzerland
Received 3 March 2005; received in revised form 23 August 2005; accepted 23 August 2005.
* Corresponding author. Tel.: +41 31 632 23 76; fax: +41 31 632 44 43. (Email: franzimmer{at}yahoo.de).
| Abstract |
|---|
|
|
|---|
METHODS: One hundred and thirty-six consecutive patients who underwent isolated CABG surgery at our institution were prospectively studied. Fifty-four patients (39.7%) were operated with mini-ECC. Patient characteristics were similar in both groups. The most interesting intraoperative details as well as in-hospital outcome were assessed.
RESULTS: There was no difference in mortality between the two groups. Cross-clamping time was similar in both groups (p = 0.07). Defibrillation was required in one patient in the mini-ECC group (1.9%) and in 38 patients (46.3%) in the CPB group (p < 0.001). In the mini-ECC group, the requirement of inotropic support and incidence of atrial fibrillation was significantly lower than in the CPB group. Postoperative creatine kinase isoenzyme MB (CK-MB) and cardiac Troponin I (cTnI) were significantly lower in the mini-ECC group (p < 0.05). Duration of ventilation, length of stay in the intensive care unit and total hospitalization time were significantly shorter in patients operated with mini-ECC (p < 0.05).
CONCLUSION: Mini-ECC is a safe procedure and is followed by a diminished release of CK-MB and cTnI than after CPB. Postoperative recovery is accelerated following mini-ECC and there is a significantly lower incidence of postoperative atrial fibrillation.
Key Words: CABG surgery Myocardial protection cTnI Minimal extracorporeal circulation
| 1. Introduction |
|---|
|
|
|---|
Cardiac Troponin I (cTnI) has been shown to be a sensitive and specific marker of myocardial damage during open heart surgery [68]. Additionally, cTnI is a reliable prognostic marker, and high cTnI-values have been shown to be associated with a worse mid-term outcome than lower values [8]. CABG on the beating heart (OPCAB) has gained popularity as an alternative to CABG with conventional CPB because all potential complications related to CPB may be eliminated. OPCAB significantly reduces postoperative cTnI-levels in comparison to coronary revascularization using cardioplegic arrest [9].
The minimal extracorporeal circulation (mini-ECC) system consists of a centrifugal pump, a membrane oxygenator, and an integrated optoelectrical suction system. The circuit has a reduced tubing length because the cardiotomy reservoir and the conventional suction device have been eliminated. The decrease in priming volume from 1800 to 600 ml helps to reduce the negative effects.
The aim of the present study is to compare perioperative myocardial damage observed with conventional CPB or with mini-ECC in patients undergoing isolated myocardial revascularization.
| 2. Patients and methods |
|---|
|
|
|---|
2.2 Operative technique
Surgery was performed through median sternotomy. After heparinization (mini-ECC: 200 IU/kg, CPB: 300 IU/kg) cardiopulmonary bypass was installed with an arterial cannula in the distal ascending aorta and a two-stage venous cannula inserted through the right atrium or by bi-caval cannulation in case a patent foramen ovale should be closed. In the mini-ECC group the venous cannulation is secured by a second snare around the atrial incision to ensure stabilization of the cannula and minimize air aspiration into the mini-ECC system. CPB was conducted in moderate hypothermia (32 °C). High-potassium cold blood cardioplegia was injected for 5 min in the aortic root immediately after cross clamping the aorta and repeated every 2030 min in the CPB group. A 1-min warm reperfusion with 37 °C oxygenated blood with a flow rate of 200300 ml/min was performed immediately after the last distal anastomosis while the mammary artery(ies) was unclamped. In the mini-ECC group, 100 ml of crystalloid cardioplegia was injected after cross clamping the aorta and repeated only if mechanical activity of the heart was observed, which was the case in three patients (5.5%). No warm reperfusion was done in this group. In both groups, proximal bypass grafts anastomoses to the ascending aorta were performed during rewarming. Priming volume of the CPB was 1800 ml and of the mini-ECC, 600 ml. The minimized cardiopulmonary bypass (MECC, Jostra AG, Hirrlingen, Germany) is driven by a centrifugal pump without cardiotomy reservoir. In our setting, air is detected by an ultrasound probe, placed at the highest level of the venous cannula. If air is detected within the mini-ECC, perfusion stops immediately. Up to now this system detected one air emboli in 520 patients operated with the mini-ECC system and removal of the air was successful. To prevent additional damage to the blood cells in the mini-ECC group a new suction device Cardiosmart® (Cardiosmart AG, Muri, Switzerland) was integrated into the system. Aspiration of blood is controlled by an optoelectrical sensor at the tip of the suction cannula and suction mechanism is started only when the tip of the suction cannula is in direct contact with the blood. The aspirated blood is directly retransfused into the venous line of the circuit and therefore no additional suction pump or reservoir is required. In the CPB group a classical suction unit was used, driven by a roller pump and the aspirated blood was forwarded to the cardiotomy reservoir.
2.3 Measurement of cardiac markers
Serial venous blood samples were taken preoperatively in both groups as well as after aortic unclamping at 6, 12, and 24 h. The cTnI concentration was analyzed by immunoassay using the Stratus II analyzer (Dade Behring). The upper limit of norm is 0.6 µg/l. Creatine kinase isoenzyme MB (CK-MB) was measured preoperatively and at 6, 12, and 24 h after unclamping the aorta. A 12-lead electrocardiogram (ECG) was recorded preoperatively and at 6 and 24 h after surgery. Patients with perioperative myocardial infarction (PMI) were excluded from the present study in order not to influence statistical analysis of myocardial markers between the two groups. Criteria used in the present study were new Q-waves of more than 0.04 ms and a reduction in R-waves of more than 25% in at least two leads in the ECG as well as echocardiographic criteria according to the guidelines reported by the STS. The CK-MB diagnosis criteria for PMI were CK-MB values of more than 60 IU/l at 6, 12 or 24 h postoperatively.
2.4 Statistical analysis
Data are presented as mean values ± their first standard deviation. A MannWhitney U-test and
2-test were used for comparison between groups of continuous and nominal variables, respectively. A p-value of less than 0.05 was considered significant.
| 3. Results |
|---|
|
|
|---|
Average age was 67.5 ± 8.5 years in the CPB group versus 66.0 ± 10.1 years in the mini-ECC group (p = ns). Preoperative characteristics were similar in both groups. Average EuroSCORE was 4.4 ± 1.2 in the CPB group and 4.7 ± 1.8 in the mini-ECC group (p = ns) (Table 1 ). Average extracorporeal circulation (ECC) time was 68.6 ± 24.1 min in the mini-ECC group and 84.1 ± 29.6 min in the CPB group (p = 0.06) and aortic cross-clamping (ACC) time was shorter in the mini-ECC group with 41.9 ± 17.2 min versus 53.4 ± 21.6 min in the CPB group (p = 0.07). The number of distal anastomoses (3.2 ± 1.0 CPB vs 3.2 ± 0.9 mini-ECC; p = ns) and the type of grafts, used for myocardial revascularization, were similar in both groups.
|
|
| 4. Discussion |
|---|
|
|
|---|
This is a prospective non-randomized protocol, which limits the power of this present study. However, similar results could be found in a smaller, randomized trial at our institution, including 40 patients prospectively randomized to either receive mini-ECC or CPB. In this collective similar results could be found, which may partly counterbalance this limitation.
We conclude that minimal cardiopulmonary bypass using the MECC® system with our technique of cardioplegic arrest is a safe procedure and seems to cause less damage to the myocardium than does conventional cardiopulmonary bypass. This is reflected by the lower release of cTnI in the mini-ECC group. One of the main advantages of the mini-ECC system is to maintain central and peripheral safe perfusion in contrast to the risk of hypoperfusion during OPCAB surgery. Earlier recovery and a significant lower incidence of postoperative atrial fibrillation are some additional benefits in patients being operated with the mini-ECC system.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y. Banz, R. Rieben, C. Zobrist, P. Meier, S. Shaw, J. Lanz, T. Carrel, and P. Berdat Addition of dextran sulfate to blood cardioplegia attenuates reperfusion injury in a porcine model of cardiopulmonary bypass Eur. J. Cardiothorac. Surg., September 1, 2008; 34(3): 653 - 660. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Ti, B.-L. Goh, P.-S. Wong, P. Ong, S.-G. Goh, and C.-N. Lee Comparison of Mini-Cardiopulmonary Bypass System With Air-Purge Device to Conventional Bypass System Ann. Thorac. Surg., March 1, 2008; 85(3): 994 - 1000. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. F. Immer, A. Ackermann, E. Gygax, M. Stalder, L. Englberger, F. S. Eckstein, H. T. Tevaearai, J. Schmidli, and T. P. Carrel Minimal Extracorporeal Circulation is a Promising Technique for Coronary Artery Bypass Grafting Ann. Thorac. Surg., November 1, 2007; 84(5): 1515 - 1521. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Castiglioni, A. Verzini, F. Pappalardo, N. Colangelo, L. Torracca, A. Zangrillo, and O. Alfieri Minimally Invasive Closed Circuit Versus Standard Extracorporeal Circulation for Aortic Valve Replacement Ann. Thorac. Surg., February 1, 2007; 83(2): 586 - 591. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Just, T. Muller, M. Hartrumpf, and J. M. Albes First experience with closed circuit/centrifugal pump extracorporeal circulation: cellular trauma, coagulatory, and inflammatory response Interactive CardioVascular and Thoracic Surgery, October 1, 2006; 5(5): 646 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Wahba Centrifugal blood pump use in routine cardiac surgery Interactive CardioVascular and Thoracic Surgery, June 1, 2006; 5(3): 299 - 300. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||