Eur J Cardiothorac Surg 2005;28:790-796
© 2005 Elsevier Science NL
Cardiovascular Research Institute Maastricht, Department of Cardiothoracic Surgery, Academic Hospital Maastricht, P. Debyelaan 25, 6229HX Maastricht, The Netherlands
Received 9 August 2005; received in revised form 14 September 2005; accepted 16 September 2005.
* Corresponding author. Tel.: +31 43 3875070; fax: +31 43 3875993. (Email: erik.vanderveen{at}ctc.unimaas.nl).
| Abstract |
|---|
|
|
|---|
Key Words: Extracorporeal life support Hemodynamics Pressurevolume relation Counterpulsation Myocardial energetics
| 1. Introduction |
|---|
|
|
|---|
In today's clinical practice, mostly, non-pulsatile ECLS circuits are employed. As set forth by Wright [6] and others [7], the choice for this technology is bound to a historical context. In the early years, pulsatile circuits were bulky and expensive systems, which required elaborate setup and tuning. The advent of roller pumps and later centrifugal pumps paved the way for smaller, cheaper, and easier to setup non-pulsatile systems. However, little scientific evidence was available to back up the trend towards continuous flow perfusion because research efforts lagged behind the technological progress. Today, the debate remains controversial, as solid evidence is still void [6,7].
The application of (continuous flow) centrifugal pumps in ECLS systems has gained widespread use clinically. Whereas continuous flow technology has valuable features like good reliability and lower cost; ECLS therapy is frequently combined with intra-aortic balloon pumping (IABP), which suggests that the specific cardiac/coronary demands cannot always be met by the extracorporeal circuit alone [1,5,8].
We sought to investigate the mechanics of pulsatile extracorporeal life support (pECLS) and specifically the hemodynamic interaction with the left ventricle and coronary circulation. Our hypothesis is that left ventricular function is impaired when extracorporeal pulse flow occurs during cardiac systole and that this will adversely affect myocardial working conditions.
| 2. Methods |
|---|
|
|
|---|
2.2 Circulatory support
The T-PLS system is a recently developed ECLS system (NewheartBio Corporation, Seoul, Korea) that features two tubular shaped pumping chambers in pushpull configuration [9,10]. The pumps fill passively and are emptied by an undulating pendulum, which pushes the sacs in alternating mode. The maximum volume of the chambers is 70 ml. The pump rate is specified as selectable from 20 to 120 bpm in steps of 1 bpm but cannot be synchronized by external inputs. The system features a suction alarm and a feedback circuit that reduces pump rate to avoid severe suction.
The ECLS system was configured as a femoro-femoral veno-arterial ECMO circuit, which was primed through the use of a venous reservoir. During support, the reservoir was bypassed. Femoral arterial access was via a 21 Fr cannula (Jostra, Hirrlingen, Germany). A femoral vein, a jugular vein, or both were cannulated for venous drainage (2427 Fr or 32/40 Fr double stage venous cannulae). An external pressure sensor (Baxter International Inc., Deerfield, IL) and an ultrasonic flow probe (Transonic Systems, Ithaca, NY) were placed on the arterial cannula to measure pump output pressure and flow. Pump output was adjusted to 4060% of total output. Fig. 1 sketches the basic experimental arrangement and presents the idea of interacting pulses from the heart and the extracorporeal circuit.
|
2.4 Conductance calibration
Parallel conductance was determined by injecting 7.5 ml of 6.5% hypertonic saline into the pulmonary artery [12]. A 5 ml blood sample was collected in a sampling cuvette for blood resistivity measurement (CD Leycom). The gain calibration factor was determined by comparing conductance derived left ventricular output (LVO) with the aortic flow at baseline. The conductance data were analyzed with the Circlab 2000 software package (Paul Steendijk, Leiden University Medical Center, Leiden, the Netherlands).
2.5 Measurements and analysis
2.5.1 Asynchronous protocol
Animal heart rate was greater than 90 bpm, and the pump rate was set between 60 and 80 bpm. Hereby, asynchronous support was established. After stabilization of bypass flow and cardiac output, hemodynamic recordings of 30 s duration were made. To differentiate between the effects of systolic arrival (SA) and diastolic arrival (DA) of the pECLS pulse at the aortic valve, the associated beats were selected and averaged (at least > 5 beats of each type).
2.5.2 Synchronous protocol
As the T-PLS system supplied for investigation did not provide any means to synchronize its output with the heart beat, the reverse option was employed. A custom cardiac pacing loop was designed around the setup shown in Fig. 1 and implemented in LabView (National Instruments, Austin, TX). A trigger pulse was derived from the upward slope of the pump outflow signal and subjected to a selectable delay. The delayed signal was then fed to the trigger input of a cardiac pacing unit. Atrial pacing was employed to ensure homogenous cardiac activation. By this arrangement, heart rate was synchronized with pumping frequency.
The simultaneous ejection of the heart and the pump was defined as the zero delay setting. Consecutively, the pace pulse was delivered in delayed fashion covering the entire cardiac cycle in steps of 50 ms. With every run hemodynamics were allowed to stabilize before recordings were made.
The following hemodynamic indices were calculated. An estimate of external left ventricular work (EW) was calculated from (VlvED VlvES)(PlvPEAK PlvED). LV potential energy (PE) was estimated by VlvES(PlvPEAK PlvED). The total pressurevolume area (PVA) was calculated from PVA = EW + PE [13]. The ratio of EW over PVA was calculated as a measure of ventricular efficiency. From the synchronous data the percentage of coronary flow during diastolic (DCF, in %) was calculated. Statistical comparison was performed by application of the non-parametric Wilcoxon signed-ranks test (n = 10). Significance was assumed if p-values were lower than 0.05.
| 3. Results |
|---|
|
|
|---|
3.1 Pump system onoff
In Table 1
the hemodynamic effects of unsynchronized support have been summarized. With the pECLS system switched on, central venous pressure decreased and aortic pressure and coronary flow increased significantly: Pcv: 20%; Pasc: +30%; and Qcor: +30% (p
< .05). PlvPEAK increased from 79 ± 24 mmHg to 97 ± 27 mmHg (p
< .05). Peak aortic pressure was elevated beyond peak left ventricular pressure (PascPEAK 103 mmHg > PlvPEAK 97 mmHg). Fig. 2
shows a typical hemodynamic recording of asynchronous pECLS support. Aortic pressure becomes biphasic with the pECLS pressure pulse arriving in diastole (beats CE). Left ventricular pressure is higher when the pulse arrives at the aortic valve during systole (beats A and B). Both systolic and diastolic left ventricular volumes are higher during these beats as well. The changes in the coronary artery flow pattern correspond to those in aortic pressure. Progressively from beats A to E, systolic coronary flow is reduced while diastolic flow is augmented. Diastolic coronary flow peaks early in diastole, which is associated with the upward slope of the pECLS pressure pulse. Beats similar to type A were selected for analysis of systolic arrival. Similarly, D-type beats were averaged for diastolic arrival data. Fig. 3
illustrates both beat types and the corresponding left ventricular pressurevolume loops. With systolic arrival ventricular ejection is decreased and pressure development is higher. The ratio of external work over potential energy is greater with diastolic arrival.
|
|
|
|
|
In Table 3 , the grouped hemodynamic differences between the afterload extremes have been summarized (n = 5). At maximum afterload, marked increases in left ventricular volume (+35 ml) and PVA (57%) were measured. Concomitantly, the diastolic coronary fraction and the ventricular efficiency showed a strong decline: DCF decreased from 93% to 53% and EW/PVA was compromised (28% vs 45% at minimum load).
|
| 4. Discussion |
|---|
|
|
|---|
The synchronous evaluation of pECLS indicated a pronounced effect of pulse incidence on ventricular function. Spot on systolic arrival was unambiguously associated with the highest left ventricular afterload (Fig. 4). Systolic impairment clearly compromises ventricular efficiency and the oxygen demandsupply balance (Table 3). Diastolically phased pECLS featured decreased LV afterload and oxygen demand, increased diastolic coronary perfusion, and improved ventricular efficiency (Fig. 4, Table 3). Other investigators have found similar attributes under experimental conditions as well as in clinical practice [13,5,1720]. Corday et al. [17], showed that veno-arterial pulsatile partial bypass had greater efficacy in augmenting (diastolic) coronary flow than IABP alone. Moreover, their setup incorporated additional end-diastolic arterial suction as a means to actively reduce left ventricular afterload. However, they found that in experiments in dogs as well as in selected clinical cases, active afterload reduction by pulsatile ECLS was associated with elevated renal flow resistance and morbidity [18]. Synchronously counterpulsating pECLS as presented in this paper does not actively reduce afterload.
Asynchronous pECLS support produced an increase in both coronary and systemic perfusion pressure. The observed peak aortic pressure exceeded peak left ventricular pressure (Table 1). Apparently, the greater number of pECLS pulses arrived at the aortic valve during diastole. This could have been expected because diastole takes up more than 50% of the cardiac cycle period. If we consider the effect, well-aligned diastolic arrival pulses have (type D; Fig. 2), an enhanced aortic to ventricular pressure difference is noted (104 over 92 mmHg mean; Table 2). The DA data from the asynchronous protocol thus serve as a model of pulsatile ECLS, as if it were phased to the diastolic period. However, these analyses depend on specific beat selections from asynchronous recordings, and therefore only describe the acute, pseudo steady-state differences between SA and DA.
4.1 Oxygen demand and supply
Mechanical cardiac support and specifically myocardial preservation are aimed at creating favorable myocardial working conditions. To achieve this, either a reduction in work (demand) or an augmentation of coronary flow (supply), or both are mandatory [8,15,19,21]. The results of the present study showed no dependence of mean coronary flow on pulse incidence. The insensitivity of mean coronary flow to these changes in perfusion pressure has been found in other studies and is associated with intact coronary autoregulation [17,19,20]. Nonetheless, diastolic coronary perfusion did change dramatically with varying delays of synchronized ejection (Fig. 2, Table 3). The perfusion of subendocardial layers is known to depend strongly on diastolic flow [15,22]. To satisfy metabolic demands during partial bypass, the heart entirely depends on coronary flow reserve, which normally is limited to 2030%. This suggests that in pathological situations little headroom is left to fulfill myocardial oxygen requirements [8].
Left ventricular myocardial work was significantly affected by systolic arrival of the pECLS pulse (Tables 2 and 3, Figs. 3 and 4). Pressurevolume area has been identified as an appropriate measure of myocardial oxygen consumption [23]. Avoidance of systolic impairment provided a reduction of 57% in oxygen consumption (Table 3). This capability to unload the myocardium while maintaining coronary flow and increasing subendocardial perfusion proves that synchronously counterpulsating pECLS is an interesting cardiac assist option when regarding myocardial working conditions. Failure to establish proper myocardial oxygen balance during ECLS may lengthen the therapeutic time path or may even lead to expansion of the ischemic/stunned myocardial mass [14,15]. Pappas et al. [14] studied the use of intra-aortic balloon pumping during cardiopulmonary bypass in both low-risk and high-risk patients. They found an immediate postoperative improvement of LV ejection fraction for the CPB plus IABP group. In the CPB only group, however, postoperative low output syndrome occurred and inotropic support and diuretics were necessary. Metabolic evaluations showed significantly higher myocardial lactate production in the CPB only group. Apparently, IABP support during CPB improved the myocardial oxygen balance, which resulted in better outcome.
4.2 Alternative to standard ECLS?
The implantation and management of both ECLS and IABP systems make the combined therapy invasive, complex (logistically), and costly. Synchronously counterpulsating pECLS may combine the attributes of IABP with those of extracorporeal partial bypass in a single system. The augmentation of cardiac output and aortic pressure and the simultaneous reduction of LV afterload suggest that myocardial oxygen consumption and peripheral perfusion pressure can be decoupled by phasing pECLS ejection into diastole (Tables 2 and 3, Figs. 3 and 4). In contrast, with non-pulsatile ECLS left ventricular afterload is directly related to bypass flow, which should fit the perfusion demand of all systemic organs [1,2,5,14]. This finding implies that phased ejection pECLS has an advantage over continuous flow ECLS. The potential benefit of synchronously counterpulsating pECLS over continuous flow ECLS plus IABP in terms of economy and simplicity remains to be investigated, preferably in a comparative (pre-) clinical trial.
The pump system used in this investigation was not equipped for synchronous support. Further technological developments in this direction should include a sync input (ECG), a selectable delay option, and dedicated monitoring of aortic pressure. The initial setup and calibration may be facilitated by catheterization or 2D-echo. The adjustment procedure of the pump ejection delay would be similar to that with ordinary intra-aortic balloon pumping [20,24]. Due to the inertia inherent in the current pump design beat-to-beat adaptation of ejection timing may not be feasible; IABP features faster pumping technology that may cope better with rapid adjustments. In most clinical applications timing optimization over several cardiac beats would probably suffice and should be aimed at avoiding spot on systolic impairment (Fig. 4). An important requirement for stable synchronization is that the propagation delay of the pump pulse from the femoral artery to the aortic arch remains constant over time. During our acute measurements this delay did not vary. With aortic cannulation this propagation delay and its variance would be negligible.
4.3 Clinical context
Recognizing the potential of phased ejection pECLS, it is important to identify where the advantages of this support modality fit in the therapeutic timeline. The patient suffering from reversible cardiogenic shock will be considered as a model in the following discussion [2,3,8]. In the early stages of treatment veno-arterial bypass flow will be nearly 100% of the total cardiac output, and the minimal volume loading of the heart will ensure low absolute myocardial oxygen requirement allowing for stabilization of basal metabolism. However, in compromised areas of the myocardium the oxygen balance remains crucial, as pressure work still requires sufficient oxygen delivery (Figs. 3 and 4) [13]. Stunned myocardium would be the most vulnerable in this stage of support [8,22]. As soon as global systemic organ function has improved up to an acceptable level, cardiopulmonary activity should be increased as a first step in rehabilitation [25]. This second stage involves lower assist ratios and consequent increased myocardial loading. Synchronously counterpulsating pECLS can be particularly helpful to optimize myocardial working conditions and controlling them independently of global perfusion (Tables 2 and 3, Fig. 4). The presence of working myocardium and increased coronary perfusion make this stage apt for the launch of adjunctive therapies [22]. When further progress is made towards weaning off ECLS, hemodynamic challenges can be employed to study cardiac and overall circulatory functional stability. Phased pECLS offers an alternative to specifically evaluate cardiac function by a simple switch over to the asynchronous mode or even timed systolic impairment for brief trial periods. Eventually, actual weaning can take place, and the patient can be treated off mechanical support or bridged to transplantation or an assist device.
4.4 Study limitations
Left ventricular function with synchronized pECLS was studied in five animals only. Solid statistical evidence could therefore not be produced, but these results did, however, prove consistent with the rest of the data and peer findings [5,14,17,18].
In the present study, we did not evaluate aortic arterial cannulation. In clinical practice, ECLS is applied with thoracic cannulation but mostly in the immediate postoperative setting. For long-term support and minimal invasive cases, peripheral cannulation is employed more widely. The avoidance of systolic impairment is more critical during long-term ECLS, which is associated with partial bypass.
Because the curves as presented in Fig. 4 would merely shift in a cyclic fashion with an arbitrary propagation delay, the results of synchronized operation are considered to remain applicable in case of aortic cannulation.
The extracorporeal circuit setup we used had a limited capability in terms of pump rate (max. 80 bpm) and flow (1.9 ± 0.3 l/min). This was due to the small size of the cannulae and the limited filling timeboth appear to be inherent to the pulsatile ECLS concept. The drainage of venous blood was essentially passive and depended on the gravitational effect and the elastic recoil of the TPLS pumping chamber only. In this study, we chose to straightforwardly evaluate the properties of the TPLS system as a unit. In clinical practice, drainage may be enhanced by the use of kinetic or vacuum assisted drainage techniques. With sufficient venous drainage, the use of ß-blocker could have been dispensed with in our protocol, because the pump would not have limited its rate in order to avoid suction.
A representative model of cardiac failure was not included in this study's design, as our focus was on the mechanical interaction of pECLS with the left ventricle. The results show that bypass flow ratio is a key determinant of how much impact pulse arrival has on ventricular function. A full range of possible bypass situations was covered in this investigation, which proved useful in determining in which stage of clinical application synchronization would pay off. We believe that the basic findings of this study can be extrapolated to pathological conditions and associated assist modes. However, the therapeutic effect of synchronized counterpulsation remains to be investigated in a long-term, failing myocardium setting.
| 5. Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |