Eur J Cardiothorac Surg 2006;29:175-180
© 2006 Elsevier Science NL
Residual air in the venous cannula increases cerebral embolization at the onset of cardiopulmonary bypass
Rosendo A. Rodriguez
a
,
b
,
*
,
Fraser Rubens
a
,
Dean Belway
c
,
Howard J. Nathan
b
a Department of Surgery, Division of Cardiac Surgery, University of Ottawa Heart Institute, Ottawa, Ont., Canada
b Department of Anesthesiology, Cardiac Division, University of Ottawa Heart Institute, Ottawa, Ont., Canada
c Department of Cardiovascular Perfusion, University of Ottawa Heart Institute, Ottawa, Ont., Canada
Received 18 May 2005;
received in revised form 21 October 2005;
accepted 3 November 2005.
* Corresponding author. Address: Department of Surgery, Division of Cardiac Surgery, Room H-340, 40 Ruskin Street, University of Ottawa Heart Institute, Ottawa, Ont., Canada K1Y 4W7. Tel.: +1 613 761 4263; fax: +1 613 761 4392. (Email: Rrodriguez{at}Ottawaheart.ca).
 |
Abstract
|
|---|
Objective: When the right atrium (RA) cannula is connected to the venous return line of the cardiopulmonary bypass (CPB) circuit, air is often introduced. Air in the venous cannula may increase cerebral air embolization at initiation of CPB despite the arterial line filter. We measured the volume of air present in the venous cannula after cannulation of the RA. Transcranial Doppler quantified emboli as high-intensity transient-signals (HITS) in both middle-cerebral arteries (MCA) at the beginning of CPB. Methods: After RA cannulation, the air column in the venous line was measured and the total volume calculated using the known lumen diameter. CPB onset was defined as the instant when the CPB machine started moving the patient's blood from the RA into the venous reservoir. Starting from CPB onset, HITS were counted: (a) until completion of the first minute on CPB (1-min count) and (b) until aortic cross clamping (pre-clamping count). Results: We studied 135 patients during coronary artery bypass surgery operated on by 10 cardiac surgeons. HITS during onset of CPB were detected in 95% of patients. Median counts were 10 HITS (25th, 75th percentiles: 3, 26) at 1-min and 21 HITS (8, 51) during pre-clamping. A significant correlation was found between the volume of air in the venous cannula and the HITS counts (r
= 0.524, p
< 0.0001). Absence of retained air was associated with lower HITS counts [3 HITS (1, 11)] compared with any amount of air [13 HITS (4, 29), p
= 0.002)]. The volume of air in the venous cannula, the MCA mean blood flow velocity and the pre-clamping time were the only independent predictors of the pre-clamping HITS counts (p
< 0.001). Conclusion: Air in the venous cannula can result in HITS in the MCA. Minimizing the volume of air introduced into the venous cannula after cannulation of the RA can decrease cerebral air embolization at the beginning of CPB.
Key Words: Cerebral embolization Cardiopulmonary bypass Transcranial Doppler
 |
1. Introduction
|
|---|
Systemic microembolization is considered a potential mechanism of injury to the brain [1] and other organs such as the kidney [2] during cardiopulmonary bypass (CPB). The composition of the majority of these emboli is gaseous [3], but a smaller fraction is solid particles derived from lipid, clots, thrombi, and atherosclerotic plaques [3,4]. Emboli can be detected in the cerebral circulation as high-intensity transient signals (HITS) using transcranial Doppler (TCD) [5]. Previous studies [1,69] have reported that between 6 and 9% of the total number of emboli detected during coronary artery bypass (CABG) surgery occur during the onset of CPB. These embolic signals are assumed to represent air bubbles delivered into the systemic circulation as a result of air retained in the components of the extracorporeal circuit [7,8,10]. Taylor et al. [11], however, have shown that air inadvertently introduced into the venous cannula due to non-occlusive purse strings or caval snares may result in HITS during CPB. During cannulation of the right atrium (RA), air often remains in the venous cannula when the cannula is connected to the venous return line of the CPB circuit. Air is not completely eliminated by the arterial line filter [12,13], and so the presence of air in the venous cannula increases the risk of cerebral air embolization during initiation of CPB. Although the role of cerebral air embolization in causing neurological damage is still debatable [14,15], it seems good practice to minimize the risk of air being introduced into the cerebral circulation. In this investigation, we measured the volume of air retained in the venous cannula after cannulation of the RA and TCD was used to quantify HITS in the cerebral circulation on initiation of CPB. We hypothesized that the volume of air retained in the venous cannula before CPB would be correlated with the counts of HITS in the middle cerebral arteries (MCA) during initiation of CPB.
 |
2. Materials and methods
|
|---|
After IRB approval, patients undergoing CABG surgery under CPB participating in a clinical trial assessing the neuroprotective effects of hypothermia during CPB [16] were studied. Patients were randomized to two different nasopharyngeal temperatures (34 °C or 37 °C) under CPB. Cannulation of the ascending aorta was achieved by using a short (2022 Fr Medtronic, Minneapolis, MN, USA) or long-tip cannula (size: 24 Fr, Sarns Flex, Ann Arbor, MI, USA) according to the surgeon's preference. The RA was cannulated through the atrial appendage using a two-stage venous cannula (32/40 or 36/46 Fr; Medtronic® Minneapolis, MN, USA). The venous cannula was attached by a connector to a 0.5 in. diameter polyvinyl chloride (PVC) tubing (length: 238 cm) connected to the venous inlet of the reservoir. Surgeons were free to practice their individual technique of cannulation and were not aware of the TCD signals at the time of surgery.
2.1 Extracorporeal circuit
The characteristics of our extracorporeal circuit and standard CPB protocol have been described in detail elsewhere [17]. The circuit was flushed with medical grade filtered CO2, primed with 1400 ml of Ringers lactate (5000 IU of heparin), and the solution was continuously re-circulated through a 5-µm filter to remove any impurities and air bubbles before cannulation. CPB was achieved by the use of gravity venous drainage.
2.2 Measurement of air in the venous cannula
After the venous cannula and its connections were secured, the venous line was inspected for the presence of air. The surgeon held the venous line from the connector in order to maintain the end of the tubing above the level of the heart. This maneuver facilitated the displacement of the air column to the upper wall of the venous line close to the venous cannula. Subsequently, the air column was measured using a scale and the total volume calculated using the known lumen diameter of the tubing line according to the following formula: air volume =
r
2
L, where
(pi) is 3.1416, r
= radius, and L
= length of air column in centimeters.
2.3 Transcranial Doppler
After induction of anesthesia, 2 MHz pulsed-wave Doppler probes (sample volume: 10 ml) were secured on the temporal area bilaterally for monitoring HITS and flow velocities in the right and left MCA. All Doppler signals were reviewed off-line by an experienced ultrasonographer (RAR). The recording Doppler parameters, equipment, criteria for differentiating between HITS and non-embolic signals as well as the degree of inter-observer agreement are described in detail elsewhere [17]. A 9-dB intensity threshold was used during Doppler signal recording. The onset of CPB was defined as the instant when the CPB machine started moving the patient's blood from the RA into the venous reservoir and this time was marked on the TCD record. Starting from the onset of CPB, HITS were counted: (a) until the completion of the first minute of CPB (1-min count) and (b) until the application of the aortic cross clamp (pre-clamping count). Interventions or manipulation of the aorta were kept at minimum during this period.
2.4 Statistical analysis
HITS counts are presented as medians (25th, 75th percentiles) and their range. Spearman rank correlation was used to investigate the association between HITS counts and air volume or between HITS counts and the time from the onset of CPB to aortic cross clamping. A linear regression analysis was used to determine the best fitting curve. The counts of HITS, air volumes, and time to aortic clamping were stratified by surgeon, type of aortic cannula, and body temperature. Their differences were assessed by non-parametric MannWhitney or KruskalWallis tests. The critical value of statistical significance was p
< 0.05. A multivariate linear regression analysis was performed using the count of HITS as dependent variable and incorporating significant univariate descriptors. The nasopharyngeal temperature was kept in the regression model as the design variable. Analyses were performed using SPSS software (version 13.0).
 |
3. Results
|
|---|
One hundred and thirty-five consecutive patients who underwent CABG surgery under CPB by 10 cardiac surgeons were studied. Table 1
shows demographics, CPB characteristics, and number of patients according to participating surgeons and body temperature. The median volume of air observed in the venous cannula before CPB was 1.14 ml (0.5 1, 2.40 ml) and the median time between the onset of CPB and the application of the aortic cross clamp was 4 min (2, 8 min). HITS in the MCA in the first minute after initiation of CPB were detected in 95% of our patients. The median count of HITS per patient from onset of CPB to aortic cross clamping was 21 HITS (8, 51 HITS) (range: 0221). Sixty-three percent of these HITS occurred within the first minute of CPB [median: 10 HITS (3, 26 HITS) (range: 0101)].
3.1 HITS counts during the first minute of CPB
There was a significant correlation between the volume of air measured in the venous line and the count of HITS per patient during the first minute of CPB (r
= 0.524, p
< 0.0001). In this relationship, larger volumes of air introduced in the venous line were associated with higher counts of HITS. This association is illustrated in Fig. 1
for all patients. In 13% of the patients there was no visible air retained in the venous line. Patients who had no visible air had lower counts of HITS compared with those where air was observed [no retained air, median HITS count: 3 HITS (1, 11 HITS); retained air, median HITS count: 13 HITS (4, 29 HITS), p
= 0.002)]. Significant differences were found among participating surgeons in both the volume of air retained in the venous line (p
< 0.0001) and the count of HITS during the first minute of CPB (p
= 0.045). Fig. 2
shows the volume of retained air and the counts of HITS stratified by surgeon. In addition, the mean blood flow velocity in the MCA measured at the end of the first minute of CPB was significantly associated with the count of HITS (r
= 0.364, p
< 0.001). Patients with blood flow velocities equal to or greater than 40 cm/s [n
= 48 (range: 4085 cm/s)] had higher counts of HITS within the first minute of CPB [median count: 25 HITS (8, 38 HITS)] compared with those with lower flow velocities [n
= 87, median count: 7 HITS (2, 21 HITS), p
< 0.0001]. There were no significant differences in the count of HITS due to the differences in nasopharyngeal temperature [37 °C, median count: 12 HITS (3, 29 HITS); 34 °C, median count: 10 HITS (3, 25 HITS); p
= 0.721] or length of the tip of the aortic cannula [short tip, median HITS count: 12 HITS (3, 26 HITS); long tip, median HITS count: 7 HITS (2, 26 HITS); p
= 0.314].

View larger version (20K):
[in this window]
[in a new window]
|
Fig. 1. Estimated linear regression (solid line) and its 95% confidence interval (interrupted lines) of the relationship between HITS counts per patient measured during the first minute of CPB (vertical axis) and the volume (in milliliter) of air (horizontal axis) identified in the venous cannula after cannulation of the right atrium. Larger volumes of air introduced into the venous cannula before CPB were associated with significantly higher counts of HITS (r
2
= 0.27). The regression function of the mathematical model corresponds to the following equation: HITS = 11.2 + 3.1 (air volume in milliliter).
|
|

View larger version (41K):
[in this window]
[in a new window]
|
Fig. 2. Distribution of the median HITS counts (upper panel) measured within the first minute of CPB and median air volumes (ml, lower panel) estimated in the venous cannula according to participating surgeons (capital letters). The median HITS counts and air volumes were higher for surgeons C, F and G, but lowest for A, D, and I.
|
|
3.2 HITS counts during pre-clamping
There was a significant correlation between the volume of air introduced into the venous cannula and the pre-clamping count of HITS (r
= 0.507, p
< 0.0001). Several factors were associated with differences in the count of HITS during this measurement. These included surgeon (p
< 0.0001), duration between the onset of CPB and aortic cross clamping (r
= 0.451, p
< 0.001), MCA mean flow velocity (r
= 0.342, p
< 0.0001), and length of the tip of the aortic cannula (p
= 0.01). Differences in nasopharyngeal temperature at initiation of CPB did not influence the pre-clamping count of HITS [37 °C, median count: 20 HITS (9, 51 HITS), 34 °C, median count: 22 HITS (8, 50 HITS), p
= 0.963]. Although higher counts of HITS during this period were observed in patients whose aorta was cannulated with a short-tip cannula relative to the long-tip cannula [short cannula, median HITS count: 28 HITS (8, 55 HITS), long cannula, median HITS count: 12 HITS (4, 30 HITS), p
= 0.01)], surgeons who cannulated the aorta with the short-tip cannula had longer durations of the pre-clamping period than those who used the long-tip cannula (p
= 0.003). The median pre-clamping duration in patients with a short-tip cannula was 5 min (2, 8 min); while in those with a long-tip cannula was only 2 min (2, 4 min).
3.3 Multivariate linear regression analysis
Due to the fact that several variables including the volume of air in the venous cannula, MCA blood flow velocities, length of the aortic cannula, and the duration of the pre-clamping period became significant during the univariate analysis, their relationship as independent predictors on the pre-clamping count of HITS was investigated using a multivariate regression analysis. Table 2
summarizes these results. The volume of air in the venous cannula, the MCA mean blood flow velocity at the end of the first minute of CPB, and the duration between CPB onset and the time of aortic cross clamping were the only independent predictors of the pre-clamping count of HITS (p
< 0.0001). The length of the tip of the aortic cannula did not have any significant effect on the pre-clamping counts of HITS (p
= 0.37).
 |
4. Discussion
|
|---|
Although modern techniques of CPB and better circuit designs have minimized the fatal cases of massive air embolism described in earlier studies [18], recent investigations [12,13,17] suggest that systemic air microembolization derived from extracorporeal sources still represents a common problem during conventional CPB. Several clinical studies [1,6,8,9] have documented large amounts of cerebral microemboli during initiation of CPB. This was assumed to be the result of air retained within the CPB circuit [3,8,10,12], but the precise mechanisms contributing to this phenomenon were not clearly identified.
In the present study, the onset of CPB was associated with varying numbers of HITS in the cerebral circulation in 95% of patients undergoing CABG surgery. The counts of HITS in the MCA during initiation of CPB were correlated with the volume of air introduced into the venous cannula at the time of connecting the RA cannula to the venous return line of the CPB circuit (see Fig. 1). These findings suggest that individual techniques of de-airing the venous cannula and its connections have a significant impact on the counts of HITS in the MCA at the onset of CPB (see Fig. 2). The present results are in agreement with a previous in vitro study [12], which demonstrated that air introduced into the venous line always resulted in increased embolic activity distal to the arterial filter. Moreover, we found that careful de-airing of the venous cannula and its connections decreased the number of HITS delivered to the brain during initiation of CPB. Patients with no visible air in the venous line had lower counts of HITS at the beginning of CPB compared with those where air was seen. The fact that a few HITS occurred in the absence of visible air in the venous line suggests that other sources might contribute to the incidence of cerebral emboli during the onset of CPB. One of the most likely sources is the presence of residual air entrapped in the extracorporeal circuit due to failure to completely de-air the arterial line filter (ALF), tubing connectors, or ports before CPB [3,10].
Several factors present during the onset of CPB may influence the speed of air bubbles traveling from the venous return line into the systemic circulation. These include the speed of venous return (7,12), low volume level in the reservoir (3,10,17), short time for achieving full flow CPB (3,10) and the physical characteristics of the venous reservoir (12,17). The internal design of the extracorporeal components may also influence the ability of the circuit to remove air bubbles [12,17]. Once air has been introduced into the venous line, venous reservoirs with a long central conduit that extends the venous inlet to the bottom of the reservoir in close approximation to the outflow may facilitate the delivery of air bubbles into the systemic circulation particularly under conditions of low reservoir volume such as those occurring at the onset of CPB [17]. In this type of circuit, air bubbles are less likely to float to the top of the reservoir and may be siphoned directly into the outflow. In addition, a sudden increase in the pump flow rate for achieving full-flow CPB shortens the circulating time of blood between the aortic cannula and the carotid arteries, and increases the speed of air bubbles [3,7,13,17] increasing the likelihood that they will reach the brain before being reabsorbed. While pericardial suction and venting have been considered potential sources for the introduction of air to the venous circuit, we assume that the contribution of pericardial suction and aortic venting on the count of HITS at the beginning of CPB was minimal. In our investigation, pericardial suction was collected into a separate cardiotomy reservoir. Due to the low volume of suction collected prior to and at the onset of CPB, our perfusionists rarely returned the cardiotomy volume into the systemic circulation prior to aortic cross clamping. In addition, active venting was usually deferred until after the first dose of cardioplegia was delivered.
Our study demonstrated that the majority of HITS (63%) detected between the beginning of CPB and aortic cross clamping occurred within the first minute of CPB. The volume of air introduced in the venous cannula, the MCA mean blood flow velocities at the end of the first minute of CPB, and the duration of the period between the onset of CPB and application of the aortic cross clamp were independent predictors of the pre-clamping count of HITS. Our finding that higher flow velocities in the MCA were associated with higher counts of HITS supports the assumption that increases in cerebral perfusion in the presence of systemic embolization result in a greater incidence of cerebral emboli [9]. The results of experimental studies suggest that increased cerebral blood flow is the most likely mechanism for the higher counts of cerebral emboli found in the brains of hypercarbic animals after systemic embolization under CPB when compared with those under hypocarbia [19].
In the present study, nasopharyngeal temperature and the length of the tip of the aortic cannula were not independent predictors of the pre-clamping count of HITS. Previous investigators [17,20] have reported that the long-tip aortic cannula delivers fewer emboli to the cerebral vasculature during CPB compared to the short-tip cannula. Despite the different count of HITS during the pre-clamping period between the two types of aortic cannulae, this relationship was obscured by the stronger association between the HITS counts and the pre-clamping times and the small sample size in the group with the long-tip cannula.
The clinical relevance of cerebral air embolization on neurologic and cognitive outcome is debatable [14,15]. Studies that have documented large quantities of gaseous microemboli during CPB have reported varying incidences of postoperative clinical effects, ranging from non-neurological sequelae [8] or mild cognitive dysfunction [14] to seizures, coma, and death [21]. Although experimental studies have shown that cerebral blood flow or metabolism remains unaffected after air embolization has occurred during hypothermic CPB [22], cerebral air embolization is a concern due to the potentially damaging effect of air bubbles on the cerebral vascular endothelium [23] and the resultant disruption of the bloodbrain barrier [21,22]. In addition, experimental studies indicate that air bubbles in blood may promote complement activation [24], leukocyte aggregation [3,22], increased platelet adherence [25], and fibrin deposition in the micro-vasculature [24] that may predispose to solid emboli.
It has proven difficult to establish a doseresponse relationship between the count of HITS during CPB and quantitative measures of cognitive outcome. While some previous studies have found correlation between HITS counts and behavioral outcomes [1,14,26,27], another study failed to find a correlation [15]. Part of the problem relates to the limitation of current Doppler systems in differentiating embolus size and composition [28]. This may be important as previous studies indicate that the composition and size of emboli may have a major impact on the severity of brain injury [29].
In summary, we investigated the effect of different volumes of air retained in the venous cannula on cerebral embolization during the onset of CPB. We found that larger volumes of air introduced into the venous cannula before CPB increase the incidence of cerebral microemboli during the beginning of CPB. Differences in individual techniques of de-airing the venous cannula and its connections have a significant impact on both the volume of air retained in the venous cannula and cerebral embolization during initiation of CPB. Although the clinical relevance of cerebral air embolization in causing neurological damage is unclear, best surgical practice should attempt to reduce the amount of air delivered to the cerebral circulation during CPB. Cerebral embolization, which frequently occurs during the onset of CPB, can be minimized by carefully de-airing the venous cannula when the RA cannula is connected to the venous return line of the CPB circuit.
 |
Acknowledgments
|
|---|
The authors are indebted to all staffs, cardiac surgeons, residents, anesthetists, perfusionists, nurses, and technicians at the University of Ottawa Heart Institute for their participation and co-operation in this study. We thank the participation of Carlos Daniel Rodriguez during transcranial Doppler analysis.
 |
References
|
|---|
- Clark RE, Brillman J, Davis DA, Lovell MR, Price TRP, Magovern GJ. Microemboli during coronary artery bypass grafting. Genesis and effect on outcome. J Thorac Cardiovasc Surg 1995;109:249-258.[Abstract/Free Full Text]
- Sreeram GM, Grocott HP, White WD, Newman MF, Stafford-Smith M. Transcranial Doppler emboli count predict rise in creatinine after coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth 2004;18:548-551.[CrossRef][Medline]
- Butler BD, Kurusz M. Embolic events. In: Gravlee GP, Davis RF, Kurusz M, Utley JR, editors. Cardiopulmonary bypass. Principles and practice. Philadelphia, PA: Lippincott; 2000. pp. 320-341.
- Brown WR, Moddy DM, Challa VR, Stump DA. Histologic studies of brain microemboli in humans and dogs after cardiopulmonary bypass. Echocardiography 1996;13:559-566.[Medline]
- Edmonds Jr. HL, Rodriguez RA, Audenaert SM, Austin III EH, Pollock Jr. SB, Ganzel BL. The role of neuromonitoring in cardiovascular surgery. J Cardiothorac Vasc Anesth 1996;10:15-23.[CrossRef][Medline]
- Braekken SK, Russell D, Brucher R, Abdelnoor M, Svennevig JL. Cerebral microembolic signals during cardiopulmonary bypass surgery. Frequency, time of occurrence and association with patient and surgical characteristics. Stroke 1997;28:1988-1992.[Abstract/Free Full Text]
- Fearn SJ, Pole R, Burgess M, Ray SG, Hooper TL, McCollum CN. Cerebral embolisation during modern cardiopulmonary bypass. Eur J Cardiothorac Surg 2001;20:1163-1167.[Abstract/Free Full Text]
- Van der Linden J, Casimir-Ahn H. When do cerebral emboli appear during open heart operations? A transcranial Doppler study. Ann Thorac Surg 1991;51:237-241.[Abstract]
- Stump DA, Rogers AT, Hammon JW, Newman SP. Cerebral emboli and cognitive outcome after cardiac surgery. J Cardiothorac Vasc Anesth 1996;10:113-119.[CrossRef][Medline]
- Merkle F, Boettcher W, Schulz F, Kopitz M, Koster A, Henning E, Hetzer R. Reduction of microemboli count in the priming fluid of cardiopulmonary bypass circuits. J Extra Corpor Technol 2003;35:133-138.[Medline]
- Taylor RL, Borger MA, Weisel RD, Fedorko L, Feindel CM. Cerebral microemboli during cardiopulmonary bypass: increased emboli during perfusionist interventions. Ann Thorac Surg 1999;68:89-93.[Abstract/Free Full Text]
- Jones TJ, Deal DD, Vernon JC, Blackburn N, Stump DA. How effective are cardiopulmonary bypass circuits at removing gaseous microemboli?. J Extra Corpor Technol 2002;34:34-39.[Medline]
- Georgiadis D, Hempel A, Baumgartner RW, Zerkowski HR. Doppler microembolic signals during cardiac surgery: comparison between arterial line and middle cerebral artery. J Thorac Cardiovasc Surg 2003;126:1638-1639.[Free Full Text]
- Borger MA, Peniston CM, Weisel RD, Vasiliou M, Green RE, Feindel CM. Neuropsychological impairment after coronary bypass surgery: effects of gaseous microemboli during perfusionist interventions. J Thorac Cardiovasc Surg 2001;121:743-749.[Abstract/Free Full Text]
- Neville MJ, Butterworth J, James RL, Hammon JW, Stump DA. Similar neurobehavioral outcome after valve or coronary artery operations despite differing carotid embolic counts. J Thorac Cardiovasc Surg 2001;121:125-136.[CrossRef][Medline]
- Nathan HJ, Parlea L, Dupuis JY, Hendry P, Williams KA, Rubens FD, Wells GA. Safety of deliberate intraoperative and postoperative hypothermia for patients undergoing coronary artery surgery: a randomized trial. J Thorac Cardiovasc Surg 2004;127:1270-1275.[Abstract/Free Full Text]
- Rodriguez RA, Williams KA, Babaev A, Rubens F, Nathan HJ. Effect of perfusionist technique on cerebral embolization during cardiopulmonary bypass. Perfusion 2005;20:3-10.[Abstract/Free Full Text]
- Pickard LR. Venous air embolism. Ann Thorac Surg 1982;33:102-103.[Medline]
- Plochl W, Cook DJ. Quantification and distribution of cerebral emboli during cardiopulmonary bypass in the swine. The impact of
. Anesthesiology 1999;90:183-190.[CrossRef][Medline]
- Borger MA, Taylor RL, Weisel RD, Kulkarni G, Benaroia M, Rao V, Cohen G, Fedorko L, Feindel CM. Decreased cerebral emboli during distal aortic arch cannulation: a randomized clinical trial. J Thorac Cardiovasc Surg 1999;118:740-745.[Abstract/Free Full Text]
- Mitchell S, Gorman D. The pathophysiology of cerebral gas embolism. J Extra Corpor Technol 2002;34:18-23.[Medline]
- Johnston WE, Stump DA, DeWitt DS, Viten-Johansen J, O'Steen K, James RL, Prough DS. Significance of gaseous microemboli in the cerebral circulation during cardiopulmonary bypass in dogs. Circulation 1993;88(Pt 2):319-329.[Free Full Text]
- Kapoor T, Gutierrez G. Air embolism as a cause of the systemic inflammatory response: a case report. Crit Care 2003;7:R98-R100.[CrossRef][Medline]
- Tamiya T, Yamasaki M, Maeo Y, Yamashiro T, Ogoshi S, Fujimoto S. Complement activation in cardiopulmonary bypass, with special reference to anaphylotoxin production in membrane and bubble oxygenators. Ann Thorac Surg 1988;46:47-57.[Abstract]
- Malmgren R, Thorsen T, Lie RT, Holmsen H. Micro bubble-induced serotonin secretion in human platelets. Thromb Haemost 1991;65:399-402.[Medline]
- Pugsley W, Klinger L, Paschalis C, Treasure T, Harrison M, Newman S. The impact of microemboli during cardiopulmonary bypass on neuropsychological functioning. Stroke 1994;25:1393-1399.[Abstract]
- Barbut D, Lo YW, Gold JP, Trifiletti RR, Yao FS, Hager DN, Hinton RB, Isom OW. Impact of embolization during coronary artery bypass grafting on outcome and length of stay. Ann Thorac Surg 1997;63:998-1002.[Abstract/Free Full Text]
- Rodriguez RA, Giachino A, Hosking M, Nathan HJ. Transcranial Doppler characteristics of different embolic materials during in-vivo testing. J Neuroimaging 2002;12:259-266.[CrossRef][Medline]
- Gaunt ME, Martin PJ, Smith JL, Rimmer T, Chrerryman G, Ratliff DA, Bell PR, Naylor AR. Clinical relevance of intraoperative embolization detected by transcranial Doppler ultrasonography during carotid endarterectomy: a prospective study of 100 patients. Br J Surg 1994;81:1435-1439.[Medline]
This article has been cited by other articles:

|
 |

|
 |
 
R. Dittrich and E. B. Ringelstein
Occurrence and Clinical Impact of Microembolic Signals During or After Cardiosurgical Procedures
Stroke,
February 1, 2008;
39(2):
503 - 511.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. F. Corno
Systemic venous drainage: can we help Newton?
Eur. J. Cardiothorac. Surg.,
June 1, 2007;
31(6):
1044 - 1051.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. A Rodriguez and D. Belway
Comparison of two different extracorporeal circuits on cerebral embolization during cardiopulmonary bypass in children
Perfusion,
September 1, 2006;
21(5):
247 - 253.
[Abstract]
[PDF]
|
 |
|