EJCTS Click here for details of sales representative
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Neil J. Howell
Bruce E. Keogh
Robert S. Bonser
Timothy R. Graham
Stephen J. Rooney
Ian C. Wilson
Domenico Pagano
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Howell, N. J.
Right arrow Articles by Pagano, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Howell, N. J.
Right arrow Articles by Pagano, D.
Related Collections
Right arrow Valve disease

Eur J Cardiothorac Surg 2006;30:10-14
© 2006 Elsevier Science NL

Patient-prosthesis mismatch does not affect survival following aortic valve replacement

Neil J. Howell, Bruce E. Keogh, Vivien Barnet, Robert S. Bonser, Timothy R. Graham, Stephen J. Rooney, Ian C. Wilson, Domenico Pagano *

Department of Cardiothoracic Surgery, University Hospital NHS Foundation Trust, Birmingham, UK

Received 28 November 2005; received in revised form 2 March 2006; accepted 14 March 2006.

* Corresponding author. Address: Department of Cardiothoracic Surgery, Queen Elizabeth Hospital, Birmingham B15 2TH, UK. Tel.: +44 121 627 2850; fax: +44 121 627 2895. (Email: domenico.pagano{at}uhb.nhs.uk).


    Abstract
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
Objective: Patient-prosthesis mismatch (PPM) has been reported to increase perioperative mortality and reduce postoperative survival in patients undergoing aortic valve replacement (AVR). We analysed the effect of PPM at values predicting severe mismatch on survival following AVR in our unit. Methods: Prospectively collected data on 1481 consecutive patients who had undergone AVR with or without coronary artery revascularisation between 1997 and 2005 were analysed. Projected in vitro valve effective orifice area (EOA) and geometric prosthesis internal orifice area (GOA) were evaluated and values were indexed to body surface area (cm2 m–2). PPM was defined as EOAi < 0.6 and/or GOAi < 1.1. Long-term survival data were obtained from the National Institute of Statistics. Results: One thousand four hundred and eighteen patients were identified. 67/1418 (4.7%) patients had GOAi < 1.1; 122/1418 (8.6%) had EOAi < 0.6 and 38 (2.6%) patients exhibited both forms of mismatch. One thousand two hundred and sixty-seven patients (89%) demonstrated no mismatch (reference group). There were 75 in-hospital deaths (overall mortality 5.3%) with no significant difference between the mismatch and the reference groups. Survival data were available for up to 8 years (median 36 months, IQR 6–60 months). There were 160 late deaths (13/143 PPM group vs 147/1198 reference group). The 5-year survival estimate was similar for both groups (83% PPM group; 81% reference group; p = 0.47). Cox-hazard analysis identified advanced age as the only predictor of reduced survival (age > 80, RR 2.13, 95% CI 1.38–4.586, p = 0.004). Conclusions: Severe patient-prosthesis mismatch was predicted in 4–10% of patients undergoing AVR but this did not affect in-hospital mortality or mid-term survival.

Key Words: Aortic valve disease • Aortic valve replacement


    1. Introduction
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
The concept of patient-prosthesis mismatch (PPM) was introduced to describe the condition of when the effective orifice of aortic valve prosthesis is less than that of a normal human valve. Different methods have been proposed to calculate PPM using either the effective orifice area (EOA) of the prosthesis [1] or the geometric orifice area (GOA) derived from the internal diameter of the prosthesis [2]. These values are then indexed to the patient's body surface area. Postoperative analysis by echocardiography has demonstrated an increased incidence of high transvalvular gradients in patients with EOAi < 0.6 cm2 m–2, which is associated with the internal valve diameter, but has failed to identify high gradients in patient groups classified with mismatch but with EOAi ≥ 0.6 cm2 m–2 [3]. It has been hypothesised that high residual transvalvular gradients due to PPM may affect left ventricular mass regression following aortic valve replacement and this can lead to an increased operative mortality, decreased long-term survival and reduced symptomatic benefit. There is, however, variable evidence to support this postulate. The purpose of this study was to analyse the effect of predicted PPM on in-hospital mortality and post-discharge survival in patients undergoing aortic valve replacement at our institution.


    2. Methods
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
This study was conducted at a single Cardiothoracic Surgical Centre. We reviewed data from the computerised cardiac surgical database, which holds clinical information on all the patients undergoing cardiac surgery at our unit since 1997. The data are acquired prospectively as part of the patients’ pathway and is based upon the Society of Cardiothoracic Surgeon of GB and Ireland minimal dataset with some customised additions. We analysed data on all consecutive patients aged >18 years undergoing first time or redo AVR (±coronary artery revascularisation) between January 1997 and March 2005. Patients requiring other valve procedures, those with pre-existing valve prosthesis other than aortic valve and those with active endocarditis were excluded from the study.

2.1 Definition of patient-prosthesis mismatch
The predicted functional effective orifice area was obtained from the manufacturers’ in vitro data and indexed to the patient body surface area (EOAi) as previously described and validated [3]. The geometric orifice area (GOAi) was calculated from the manufacturers’ geometric internal valve orifice diameter, a constant physical dimension, using the simple equation: area = {pi}r 2, and indexed to the patient body surface area [2].

For the purpose of this study, an EOAi < 0.6 cm2 m–2 and/or GOAi < 1.1 cm2 m–2 were used as markers of severe PPM [2,4]. Patients without PPM are referred to as the reference group. For the purpose of the survival analysis, the group with PPM by any method was compared with those with no PPM.

2.2 Study end points
The study end points were all-cause in-hospital mortality and late survival. Post-discharge survival data was obtained from the National Office of Statistics, UK.

2.3 Data analysis
The data were analysed using SPSS© Version 12.0.1 for Windows. The difference between groups, for preoperative and operative variables was tested by Fischer Exact test. Continuous data were compared by Student's t-test and are expressed as mean ± standard deviation. Backward conditional stepwise logistic regression analysis with a selection cut-off set at 0.05 using variables found to be significant in univariate analysis was performed to identify independent predictors of in-hospital mortality. Survival data was analysed by Kaplan–Meier life actuarial methods. Difference in survival between groups was tested by the log rank statistic. Cox proportional regression analysis was performed to identify independent predictors of late mortality.


    3. Results
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
The study population consists of 1418 patients and their baseline and operative characteristics are summarised in Table 1 .


View this table:
[in this window]
[in a new window]
 
Table 1. Patients’ demographics
 
3.1 In-hospital outcomes
The overall in-hospital mortality was 5.3% (75/1418 patients). Mortality for patients undergoing first time or redo isolated AVR was 3.8% (36/944) and for patients undergoing AVR + CABG was 8.2% (39/474) (p < 0.001). The EOAi method identified 122 patients with PPM (Fig. 1A), the GOAi method identified 67 patients with PPM (Fig. 1B), 38 patients reached both EOAi and GOAi criteria for PPM and 151 patients had PPM by either method (Table 2 ). Mismatch and reference patient groups were comparable for most risk factors, except for patients with EOAi < 0.6 cm2 m–2. This group was older and had a higher prevalence of female gender (Table 1). There was no significant difference for in-hospital mortality between patients with any form of PPM and the reference group (Table 2). At multivariate analysis, independent predictors of in-hospital mortality were advanced age (>80 years), female gender, diabetes, impaired left ventricular function, need for CABG and emergency surgery (Table 3 ).


Figure 1
View larger version (17K):
[in this window]
[in a new window]
 
Fig. 1. (A) Frequency distribution of patients’ effective orifice area index (EOAi); in dark EOAi < 0.6. (B) Frequency distribution of patients’ geometric orifice area index (GOAi); in dark GOAi < 1.1.

 

View this table:
[in this window]
[in a new window]
 
Table 2. In-hospital mortality by mismatch group
 

View this table:
[in this window]
[in a new window]
 
Table 3. Multivariate analysis of predictors for in-hospital mortality following AVR
 
There were 71 patients with poor LV (ejection fraction < 30%), of whom 6 (7%) reached the criteria for PPM. There was no significant difference in mortality between these two groups as there was one death in the PPM group (16.7%) and 11 deaths in the reference group (16.9%), p = 0.99.

3.2 Survival analysis
Survival data were available on all 1343 patients discharged from hospital. The median follow-up time was 3 years (IQR 18–44 months) (4029 patient-years follow-up). The 5-year survival estimate was similar for both groups (94% PPM group; 83.8% reference group; p = 0.37) (Fig. 2 ). Difference in survival between groups by log rank statistic was not significant, p = 0.61. Cox regression analysis identified only age > 80 years as the only independent predictor for reduced survival after hospital discharge (p = 0.007; RR 2.43, 95% CI 1.28–4.586).


Figure 2
View larger version (10K):
[in this window]
[in a new window]
 
Fig. 2. Survival plot of patients with patients-prosthesis mismatch (PPM) and those without PPM (reference group).

 
In the subgroup of patients with pure aortic stenosis there were 128 late deaths. There were 13 deaths in the PPM group and 115 in the reference group. The 5-year survival estimate was similar for both groups (91% PPM group; 83% reference group; p = 0.37). Difference in survival between groups by log rank statistic was not significant (p = 0.76).


    4. Discussion
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
Patient-prosthesis mismatch has been vaguely defined, often on the basis of small aortic prosthesis size, EOA, GOA and indexed values of EOA and GOA. In our study we defined PPM using values of EOAi and GOAi uniformly recognised to represent severe mismatch. In this study severe PPM occurred in 4–10% of patients undergoing AVR. This, however, did not translate into increased early in-hospital mortality or a decreased survival after discharge from hospital. Some reports have indicated that PPM can be associated with an increase in short-term mortality following AVR [1,5,6] and this phenomenon has been attributed to reduced cardiac performance following surgery. Most of these reports, however, have not been risk adjusted. Patients with a small aortic annulus are usually older, and they have more severe left ventricular hypertrophy with advanced diastolic dysfunction and coronary artery disease [7]. A recent study has identified a small increase (1%) in early postoperative mortality when using small aortic prostheses, but this finding was only evident after analysing a very large patient population [2]. Furthermore, the surgical alternative in these patients would have involved some form of aortic root enlargement, a procedure known to be associated with a significant increase in mortality [8].

It has been suggested that PPM is particularly harmful in patients with poor LV function preoperatively. Our study includes a small number of patients with poor LV function and is therefore not powerful enough to identify differences in outcome in this cohort.

Rao et al. [5] reported that severe PPM following AVR was associated with decreased long-term survival and hypothesised that this was due to the persistence of left ventricular hypertrophy in patients with small aortic prostheses. In that study, however, patients with PPM had an increase in valve-related mortality, whilst the overall survival was similar. This valve-related mortality may have included mechanisms of death probably unrelated to PPM such as embolic stroke, endocarditis, valve failure, and re-operation. In contrast a study by Hanayama et al. [4] has shown that patients with postoperative echocardiographic evidence of severe PPM (EOAi < 0.6 cm2 m–2) had long-term survival similar to patients without mismatch. Furthermore, left ventricular mass index was not different between the two groups in that study. A recent multicentre report including more than 13,000 patients [2] has confirmed that PPM does not affect intermediate and long-term survival. Left ventricular mass regression following AVR depends not only upon postoperative transvalvular gradients but also on the control of blood pressure, the use of angiotensin-converting enzyme inhibitors and the degree of preoperative hypertrophy and myocardial fibrosis. Studies that address LV hypertrophy regression do not usually include details of these confounding variables. More recent studies comparing stented to stentless aortic prostheses, which are hypothesised to be associated with lower postoperative transvalvular gradients, have failed to detect significant differences in hypertrophy regression and survival when using these prostheses [9].

In our study, advanced age at the time of surgery was the only adverse predictor of adverse long-term survival post AVR, as previously shown [5].

4.1 Study limitations
We have used in vitro manufactures’ EOA which may overestimate in vivo echocardiographic EOA [10] and we have no data on postoperative transvalvular gradients. However, controversy exists on the appropriate timing for the acquisition of in vivo echocardiographic data, as haemodynamic changes can occur up to 1 year post-AVR [11–13]. There is a poor correlation between PPM and postoperative gradients [4]. Furthermore, Doppler studies may underestimate EOA in bi-leaflet prostheses due to localised high velocity jets [14,15].

We could not address the impact of PPM on exercise capacity and quality of life following AVR in our study. The impact of PPM on the functional outcome following AVR is difficult to evaluate because of the confounding effects of concomitant cardiovascular and non-cardiovascular disease. Conflicting results are reported when exercise capacity is measured directly after AVR [16–19]. A recent study has shown that the majority of patients after AVR report improvements in functional quality of life and these are not affected by PPM [20]. Finally, the follow-up data obtained from the Office of National Statistics only provides knowledge of the survival status with no cause of death, this does not allow differentiation between cardiovascular and other cause mortality and we have neither follow-up data on patient functional status nor follow-up echocardiographic data on EOA or left ventricular mass regression.

In conclusion, our study indicates that severe PPM may occur in a small number of patients undergoing AVR, but its clinical significance may be less then previously hypothesised.


    Appendix A
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 
Conference discussion

Dr M. Thubrikar (Irvine, California): I have a basic question which I have pondered for several years and don’t seem to quite find an answer to. When we are looking at the size of the annulus and the size of the valve to try to match, the way we measure the annulus is the standard way, but my question relates to how do we know at what point in the cardiac cycle has the heart been arrested so the annulus of the valve now is either in a diastolic profile or in a systolic profile. That could easily make a difference of 2–3 mm.

Mr Howell : By necessity we size the aortic annulus in the arrested heart and therefore in conditions resembling diastole.

Dr Thubrikar : Is there any attempt to see if you have a valve size on the basis of angiogram or echo beforehand and then try to match the valve to that?

Mr Howell : It is not routinely done in our department. We adopt transoesopahegeal echocardiography to size the aortic annulus in patients in whom we plan to use an aortic homograft in order to defrost the right prosthesis.

Dr A. Wahba (Trondheim, Norway): There is one question that I would like you to answer, and that is the small group of patients with bad left ventricular function, was there a difference between those that had a patient-prosthesis mismatch and the other ones that didn’t? And was your study big enough, were there enough patients in the groups to see a difference, if there was one?

Mr Howell : This is a well-debated issue. In our study there were only 72 patients with poor LV function and therefore our data are not powered to answer this specific question.

Dr J. Tsai (Pingtung, Taiwan): I have a question. It puzzled me for years. The mismatch was found very serious in females from your study, I agree with you, and the disaster also in the females. Would you tell us the reason why?

Mr Howell : It often reported in the literature that smaller EOAs occur more commonly in female and in elderly patients, and this was the case in our study.


    Acknowledgments
 
In this study, statistical advice was provided by the University of Birmingham.


    Footnotes
 
{star} Presented at the joint 19th Annual Meeting of the European Association for Cardio-thoracic Surgery and the 13th Annual Meeting of the European Society of Thoracic Surgeons, Barcelona, Spain, September 25–28, 2005.


    References
 Top
 Abstract
 1. Introduction
 2. Methods
 3. Results
 4. Discussion
 Appendix A
 References
 

  1. Pibarot P, Honos GN, Durand LG, Dumesnil JG. The effect of prosthesis-patient mismatch on aortic bioprosthetic valve hemodynamic performance and patient clinical status. Can J Cardiol 1996;12:379-387.[Medline]
  2. Blackstone EH, Cosgrove DM, Jamieson WR, Birkmeyer NJ, Lemmer Jr. JH, Miller DC, Butchant E, Rizzoli G, Yacoub M, Chai A. Prosthesis size and long-term survival after aortic valve replacement. J Thorac Cardiovasc Surg 2003;126:783-796.[Abstract/Free Full Text]
  3. Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol 2000;36:1131-1141.[Abstract/Free Full Text]
  4. Hanayama N, Christakis GT, Mallidi HR, Joyner CD, Fremes SE, Morgan CD, Mitoff P, Goldman B. Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant. Ann Thorac Surg 2002;73:1822-1829.[Abstract/Free Full Text]
  5. Rao V, Jamieson WR, Ivanov J, Armstrong S, David TE. Prosthesis-patient mismatch affects survival after aortic valve replacement. Circulation 2000;102(19 Suppl. 3):III5-III9.
  6. He GW, Grunkemeier GL, Gately HL, Furnary AP, Starr A. Up to thirty-year survival after aortic valve replacement in the small aortic root. Ann Thorac Surg 1995;59:1056-1062.[Abstract/Free Full Text]
  7. David TE. Is prosthesis-patient mismatch a clinically relevant entity?. Circulation 2005;111:3186-3187.[Free Full Text]
  8. Sommers KE, David TE. Aortic valve replacement with patch enlargement of the aortic annulus. Ann Thorac Surg 1997;63:1608-1612.[Abstract/Free Full Text]
  9. Perez de Arenaza D, Lees B, Flather M, Nugara F, Husebye T, Jasinski M, Khan M, Henenin M, Gaer J, Guvendik L, Bochenek A, Wos S, Lie M, Van Nooten G, Pennell D, Pepper J. Randomized comparison of stentless versus stented valves for aortic stenosis: effects on left ventricular mass. Circulation 2005;112:2696-2702.[Abstract/Free Full Text]
  10. Dumesnil JG, Honos GN, Lemieux M, Beauchemin J. Validation and applications of indexed aortic valve prosthetic valve areas calculated by Doppler echocardiography. J Am Coll Cardiol 1990;16:637-643.[Abstract]
  11. Dumesnil JG, LeBlanc MH, Cartier PC, Metras J, Desaulniers D, Doyle DP, Lemieux M, Raymond G. Hemodynamic features of the freestyle aortic bioprosthesis compared with stented bioprosthesis. Ann Thorac Surg 1998;66(6 Suppl.):S130-S133.[CrossRef][Medline]
  12. Cartier PC, Dumesnil JG, Metras J, Desaulniers D, Doyle DP, Lemieux MD, Raymond G. Clinical and hemodynamic performance of the Freestyle aortic root bioprosthesis. Ann Thorac Surg 1999;67:345-349.[Abstract/Free Full Text]
  13. Westaby S, Huysmans HA, David TE. Stentless aortic bioprostheses: compelling data from the Second International Symposium. Ann Thorac Surg 1998;65:235-240.[Abstract/Free Full Text]
  14. Baumgartner H, Khan S, DeRobertis M, Czer L, Maurer G. Discrepancies between Doppler and catheter gradients in aortic prosthetic valves in vitro. A manifestation of localized gradients and pressure recovery. Circulation 1990;82:1467-1475.[Abstract/Free Full Text]
  15. Henneke KH, Pongratz G, Bachmann K. Limitations of Doppler echocardiography in the assessment of prosthetic valve hemodynamics. J Heart Valve Dis 1995;4:18-25.[Medline]
  16. Tatineni S, Barner HB, Pearson AC, Halbe D, Woodruff R, Labovitz AJ. Rest and exercise evaluation of St. Jude Medical and Medtronic Hall prostheses. Influence of primary lesion, valvular type, valvular size, and left ventricular function. Circulation 1989;80(3 Pt 1):I16-I23.
  17. Barner HB, Labovitz AJ, Fiore AC. Prosthetic valves for the small aortic root. J Card Surg 1994;9(2 Suppl.):154-157.[Medline]
  18. Pibarot P, Dumesnil JG, Jobin J, Lemieux M, Honos G, Durand LG. Usefulness of the indexed effective orifice area at rest in predicting an increase in gradient during maximum exercise in patients with a bioprosthesis in the aortic valve position. Am J Cardiol 1999;83:542-546.[CrossRef][Medline]
  19. Pibarot P, Dumesnil JG, Jobin J, Cartier P, Honos G, Durand LG. Hemodynamic and physical performance during maximal exercise in patients with an aortic bioprosthetic valve: comparison of stentless versus stented bioprostheses. J Am Coll Cardiol 1999;34:1609-1617.[Abstract/Free Full Text]
  20. Koch CG, Khandwala F, Estafanous FG, Loop FD, Blackstone EH. Impact of prosthesis-patient size on functional recovery after aortic valve replacement. Circulation 2005;111:3221-3229.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
HeartHome page
J.-L. Monin
Prosthesis-patient mismatch: myth or reality?
Heart, June 1, 2009; 95(11): 948 - 952.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. R. Moon, J. S. Lawton, N. Moazami, N. A. Munfakh, M. K. Pasque, and R. J. Damiano Jr.
POINT: Prosthesis-patient mismatch does not affect survival for patients greater than 70 years of age undergoing bioprosthetic aortic valve replacement.
J. Thorac. Cardiovasc. Surg., February 1, 2009; 137(2): 278 - 283.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
D. Mohty, J. G. Dumesnil, N. Echahidi, P. Mathieu, F. Dagenais, P. Voisine, and P. Pibarot
Impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: influence of age, obesity, and left ventricular dysfunction.
J. Am. Coll. Cardiol., January 6, 2009; 53(1): 39 - 47.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
M. Vicchio, A. Della Corte, L. S. De Santo, M. De Feo, G. Caianiello, M. Scardone, and M. Cotrufo
Prosthesis-Patient Mismatch in the Elderly: Survival, Ventricular Mass Regression, and Quality of Life
Ann. Thorac. Surg., December 1, 2008; 86(6): 1791 - 1797.
[Abstract] [Full Text] [PDF]


Home page
Asian Cardiovasc. Thorac. Ann.Home page
G. Casali, G. Luzi, M. Vicchio, P. Lilla della Monica, G. Minardi, and F. Musumeci
Echocardiographic Follow-Up after Implanting 17-mm Regent Mechanical Prostheses
Asian Cardiovasc Thorac Ann, June 1, 2008; 16(3): 208 - 211.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
S Bleiziffer, W B Eichinger, I Hettich, D Ruzicka, M Wottke, R Bauernschmitt, and R Lange
Impact of patient-prosthesis mismatch on exercise capacity in patients after bioprosthetic aortic valve replacement
Heart, May 1, 2008; 94(5): 637 - 641.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
P. Totaro and V. Argano
Patient prosthesis mismatch after mitral valve replacement: Myth or reality?
J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 697 - 701.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
P. Pibarot and J. G Dumesnil
Prevention of valve prosthesis--patient mismatch before aortic valve replacement: does it matter and is it feasible?
Heart, May 1, 2007; 93(5): 549 - 551.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
S. Bleiziffer, W. B Eichinger, I. Hettich, R. Guenzinger, D. Ruzicka, R. Bauernschmitt, and R. Lange
Prediction of valve prosthesis-patient mismatch prior to aortic valve replacement: which is the best method?
Heart, May 1, 2007; 93(5): 615 - 620.
[Abstract] [Full Text] [PDF]


Home page
Eur. J. Cardiothorac. Surg.Home page
S. H. Rahimtoola
Is severe valve prosthesis-patient mismatch (VP-PM) associated with a higher mortality?
Eur. J. Cardiothorac. Surg., July 1, 2006; 30(1): 1 - 1.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Neil J. Howell
Bruce E. Keogh
Robert S. Bonser
Timothy R. Graham
Stephen J. Rooney
Ian C. Wilson
Domenico Pagano
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Howell, N. J.
Right arrow Articles by Pagano, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Howell, N. J.
Right arrow Articles by Pagano, D.
Related Collections
Right arrow Valve disease


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS