Eur J Cardiothorac Surg 2006;30:278-284
© 2006 Elsevier Science NL
a Department of Anesthesia and Perioperative Care, University of California, San Francisco, CA, USA
b Institute of Hematology and Oncology, Triemli City Hospital Zurich, Switzerland
c Department of Cardiovascular Surgery, Triemli City Hospital Zurich, Switzerland
d Department of Psychosocial Medicine, University Hospital Zurich, Switzerland
e Institute of Anesthesiology and Intensive Care Medicine, Triemli City Hospital Zurich, Birmensdorferstrasse 497, CH-8063 Zurich, Switzerland
Received 25 December 2005; received in revised form 11 April 2006; accepted 8 May 2006.
* Corresponding author. Tel.: +41 44 466 2209; fax: +41 44 466 2743. (Email: Christoph.hofer{at}triemli.stzh.ch).
| Abstract |
|---|
|
|
|---|
Key Words: Activated clotting time (ACT) Aprotinin Kaolin Anticoagulation Measurement techniques Coagulation Cardiopulmonary bypass
| 1. Introduction |
|---|
|
|
|---|
Recently, a new ACT test, SaiACT (Sonoclot® Coagulation & Platelet Function Analyzer, Sienco Inc., Arvada, CO, USA), has been developed for the SONOCLOT analyzer, with specific claims to provide a heparin dose response substantially unaffected by the presence of aprotinin. This novel test is labeled aprotinin-insensitive by the manufacturer (SONOCLOT's aprotinin-insensitive ACT, SaiACT). The cuvette is manufactured with a blend of celite to initiate blood coagulation and a type of clay to neutralize aprotinin. Like other ACT machines, the SONOCLOT analyzer also incorporates a mechanical means to detect a fibrin clot. A resonant mechanical oscillator responds to viscoelastic changes that occur during clot formation. After adding a blood sample to the SaiACT cuvette and mixing the sample with the mineral reagents, the change in impedance to movement imposed by the developing clot is measured and an activated clotting time determined.
We have previously evaluated this new SaiACT technology in vitro in presence of clinically relevant concentrations of heparin, aprotinin, and hemodilution [5]. Addition of aprotinin to heparinized blood samples induced no significant changes of SaiACT measurements. By contrast, ACT values measured with an established kaolin-based ACT from HEMOCHRON (HkACT; Hemochron® 801, International Technidyne Corp., Edison, NJ, USA) increased significantly in heparinized samples if aprotinin was present, and in vitro hemodilution pronounced this effect.
The aim of the present study was to evaluate this new SaiACT from SONOCLOT in vivo in patients undergoing CPB in the presence of heparin and aprotinin, and to compare the SaiACT test with an established kaolin-based ACT from HEMOCHRON (HkACT) as well as plasma levels of heparin measured by anti-factor Xa activity (anti-Xa).
| 2. Materials and methods |
|---|
|
|
|---|
Anesthesia, heparin anticoagulation, CPB, protamine reversal, and transfusion therapy were all managed by standardized institutional protocols. Anesthesia was induced and maintained with propofol and fentanyl, and pancuronium bromide was given for neuromuscular blockade. Fluid management was done with lactated Ringer's solution (Laboratory Dr. Bichsel AG, Switzerland) and 6% hydroxy ethyl starch solution (HES 130/0.4, Voluven®, Fresenius Kabi, Bad Homburg, Germany). Anticoagulation for CPB was attained with IV porcine heparin (Liquemin®, Roche Pharma, Switzerland) 300 U kg1. Heparin management was guided by HkACT, and HkACT > 480 s was accepted as adequate anticoagulation for CPB. CPB was performed with a membrane oxygenator (Quadrox HMO1010, Maquet Cardiopulmonary AG, Hirlingen, Germany) under moderate hypothermia (2832 °C) at flows between 2.2 and 2.4 l min1 m2. Ten thousand units of heparin but no aprotinin was added to the standard priming volume (1800 ml) of the CPB circuit. After initiating the CPB, one bolus of aprotinin 2 Mio kIU (Trasylol®; Bayer Pharmaceuticals Corp., Germany) was administered directly to the CPB circuit (modified low-dose aprotinin regimen, also called pump-prime-only regimen [6]). To keep HkACT > 480 s, additional heparin in 5000 U increments were administered during CPB, if necessary. Anticoagulation was reversed after rewarming and separation from CPB with protamine up to a maximum dose of 1 mg per 100 U of total heparin dose administered.
Blood samples were taken from an unheparinized central venous line after removing five dead space volumes of blood at the following measurement time points: baseline, after induction of anesthesia but before skin incision (T0); before CPB, 3 min after the first (200 U kg1, T1) and 3 min after the second dose (100 U kg1, T2) of heparin; 5 min on CPB, before administration of aprotinin 2 Mio kIU to the CBP circuit (T3); 15, 30, and 60 min on CPB after administration of aprotinin (T4, T5, and T6); and after protamine infusion and re-transfusion of the remaining blood from the CPB circuit (T7).
ACT was measured with two different ACT analyzers each in duplicate, the SONOCLOT analyzer with the novel aprotinin-insensitive ACT test (SaiACT; Sonoclot® Coagulation & Platelet Function Analyzer, Sienco Inc.; normal range in whole blood 6293 s) and the HEMOCHRON analyzer with the standard kaolin-based ACT test (HkACT; Hemochron® 801, International Technidyne Corp.; normal range in whole blood 91151 s). For the SaiACT, 360 µl of freshly withdrawn blood was filled into the cuvette, mixed and analyzed. Immediately after, 2 ml of the same blood specimen was filled into a HkACT cuvette, mixed and analyzed. The performance of each machine was verified with recommended quality control tests according to the manufacturers. Results were recorded as mean of duplicate measurements for each of the devices. All measurements were performed by the same investigator to avoid inter-observer variability.
To measure laboratory blood coagulation, further blood samples were withdrawn at each time point in citrated tubes (final concentration of sodium citrate 0.109 mol l1; Vacuette® 9NC, Greiner Bio-One, Austria). The samples were immediately centrifuged (2500 x g for 20 min at 4 °C), and the supernatant (plasma) stored at 32 °C for later measurements. Blood coagulation parameters were measured on an automated STA-R® coagulation analyzer (Diagnostica Stago, Asnière, France). At each time point, the rate of factor Xa inhibition (anti-Xa, assessment of the heparin concentration) and antithrombin concentration (AT) was measured. Anti-Xa activity was determined by assessing the level of inhibition of the hydrolysis of a chromogenic substrate (by the factor Xa) in the presence of heparin-antithrombin complexes with the STA®-Rotachrom-Heparin test (Diagnostica Stago). This assay is specifically designed to reflect a directly proportional relationship between the rate of factor Xa inhibition and the heparin concentration. Some of the samples had to be diluted with normal pooled plasma prior to the analysis, in order to be on the linear part of the standard curve (0.100.70 U ml1). AT activity was determined with the use of the chromogenic STA®-Antithrombin-III test (Diagnostica Stago, normal range in whole blood = 80120%). All coagulation tests and quality controls (on normal and abnormal levels) were performed according to the manufacturer's instructions.
Statistical analysis was done using StatView® for Windows version 5.01® (SAS Institute Inc., Cary, NC, USA) and SPSS for Windows Release 12.0.2 (SPSS Inc., Chicago, IL, USA). Bland and Altman analysis was done to compare SaiACT with HkACT [7]. Bias is defined as mean of difference (SaiACT HkACT) and ±2 standard deviations (SD) reflect upper and lower limits of agreement. ANOVA for repeated measures with post hoc BonferroniDunn correction and a two-sided paired Student's t-test were performed to compare the coagulation parameters at different time points. Pearson's correlations with Fisher's z-transformation and Hotelling Williams test were used to assess the relationship between both SaiACT and HkACT with heparin levels (anti-Xa) before and after aprotinin administration as well as between AT and hematocrit. Test variability of duplicate measurements was calculated as percentage of the mean of duplicate measurements. Unless otherwise stated, data are presented as mean ± SD. P-values <0.05 were considered statistically significant.
| 3. Results |
|---|
|
|
|---|
|
|
|
Correlation of ACT measurements with anti-Xa activity (Table 2 ) was comparable for SaiACT and HkACT before patients received aprotinin (T0T3). After administration of aprotinin (T4T7), correlation remained unchanged for SaiACT, but worsened significantly for HkACT.
|
Test variability was comparable for both ACT measurement techniques: Overall test variability for SaiACT was 7.5 ± 7.4% and for HkACT 7.8 ± 11%. No significant difference was observed in heparinized blood samples before (T1T3) and after (T4T7) aprotinin administration (SaiACT: P = 0.438; HkACT: P = 0.087).
| 4. Discussion |
|---|
|
|
|---|
Heparin anticoagulation is used during cardiac surgery to prevent overt thrombosis of the extracorporeal circuit and to minimize excessive CPB-related activation of the hemostatic system. As there is substantial variability of heparin anticoagulant responsiveness, heparin administration is usually monitored by point-of-care instruments that measure ACT. The ACT, initially described as a manual technique by Hattersley [8] and introduced into cardiac surgery by Bull et al. [9] is the amount of time it takes to form a clot by contact activation of the coagulation cascade. ACT measurement may be performed using different coagulation activators like diatomaceous earth (celite), clay (kaolin), glass-beads, or a blend of these materials. Different activators have different characteristics and interactions, and even the same coagulation activator manufactured by different companies respond differently under similar conditions [3]. Results from different ACT tests cannot be used interchangeably. Different baseline for SaiACT has to be accounted for when used as alternative to HkACT. In this study, baseline SaiACT values were 1926% lower compared to HkACT, on average. This finding has to be considered using SaiACT to guide anticoagulation during CPB.
Although widely used, many factors patient, operator, and equipment may affect ACT measurements. The ACT is subject to bias from various interventions that are typical during cardiac surgery, particularly patient hypothermia [10], inadequacy of specimen warming [11], hemodilution [11], quantitative and qualitative platelet abnormalities [12], or aprotinin infusion [2,3]. The technique has been criticized because of its extreme variability and the weak correlation with plasma heparin concentrations during CPB [13]. Our data show a strong correlation between ACT measurements and heparin concentration before CPB. Initiating CPB increased both ACT readings similarly but heparin levels remained unchanged (Fig. 1A and B) worsening the correlation between ACT measurements and anti-Xa. Hemodilution by itself has been shown to be responsible for the prolongation of the ACT measurements during initiating CPB [11]. Furthermore, correlation of ACT measurements with anti-Xa activity was unchanged for SaiACT before and after aprotinin administration, but declined significantly for HkACT measurements after aprotinin administration.
The concentration of AT paralleled the course of the hematocrit therefore reflecting at least in part the degree of hemodilution by the CPB, as shown by Linden et al. [14] earlier. Interestingly, despite the decrease in AT to about 50% after initiating CPB in our study, the concentration of AT was still adequate for heparin to exert its potent anticoagulant effect. It is known that AT concentrations correlate with heparin's effect to inhibit coagulation: lower AT levels are associated with a decreased heparin dose response as measured by the ACT. However, in vitro data showed that only AT levels <30% are associated with impaired heparin action [15].
Aprotinin has been questioned lately for its overall safety in two large studies on patients undergoing CPB: Mangano et al. [16] and Karkouti et al. [17] reported dose-dependent serious renal, cardio- and cerebrovascular adverse events after aprotinin administration in this clinical setting. Interestingly, patients treated with aminocaproic acid or tranexamic acid showed similar reduction in blood loss compared to patients treated with aprotinin, but these patients had no renal, cardio- and cerebrovascular adverse events [16,17]. The question remains, why only aprotinin but not the other antifibrinolytic agents caused these serious adverse events. In both studies, patients received one of the anti-fibrinolytic agents at physician discretion, i.e., the studies were non-randomized. However, both studies included large numbers of patients and statistical procedures were used to adjust for known differences between the treatment groups. Another possible explanation for the adverse events seen in patients treated with aprotinin could have been that these patients were under-heparinized during CPB. Aprotinin is a non-specific serine protease inhibitor and may prolong ACT measurements to various degrees depending on the coagulation activator used. This drug is known to inhibit contact activation, preferentially celite mediated activation in vitro [18,19]. Kaolin-based ACT is less affected than celite-based ACT, most likely because kaolin binds aprotinin [20] and because kaolin more potently activates coagulation than celite [18]. However, kaolin-based ACT has also been shown to be prolonged significantly in the presence of aprotinin [3,4]. Overestimation of anticoagulation, i.e., falsely prolonged ACT implies a potential hazardous risk of subtherapeutic heparin anticoagulation and must be avoided during CPB. For example, in a recent investigation, Koster et al. [21] showed that heparin management with kaolin-based ACT resulted in lower heparin concentrations compared to a heparin concentration-based anticoagulation management during CPB. These lower heparin concentrations used in patients managed by kaolin-based ACTs were associated with increased hemostatic activation and inflammatory response.
Limited data exist that define the optimal ACT for initiation and maintenance of CPB. In a recent survey of the Society of Cardiovascular Anesthesiology and American Society of Extracorporeal Circulation, it was found that the target ACT used by 82% of responders was 400480 s or greater, with an additional 4.5% targeting an even higher ACT [10]. When the achievement of the usual target of >480 s was considered, an adequate anticoagulation during CPB after aprotinin administration, 96% of all ACT values were classified as therapeutic by HkACT, but this was only the case in 86% of all values if ACT was determined by SaiACT.
Each ACT measurement was performed in duplicate for both SaiACT and HkACT to evaluate test variability. Both methods were comparable and showed mean test variability between 7 and 8%. According to the manufacturers of both analyzers, coefficient of variation should not exceed 5% under control conditions. Nonetheless, published data on performance of ACT devices in control plasma and whole blood differ largely and were mostly below 10% [22,23].
Costs for the two ACT tests used in this study are comparable at our institution: we pay approximately three EUR for both ACT tests taking into account our current workload (we perform 700010,000 ACT tests per year) and rental agreements.
This study has some limitations. We were able to conclude that heparin management with SaiACT may result in an increased administration of heparin. However, we did not answer the question if higher heparin doses in patients treated with aprotinin are justified in terms of patient's outcomes. Further studies are needed, measuring coagulation activation as well as recording patient's outcome data in heparinized patients on CPB guided by SaiACT. Furthermore, according to our institutional protocol, our patients received aprotinin only in a modified half-dose regimen (pump-prime-only regimen, 2 Mio kIU in CPB circuit). For this regimen, anticipated aprotinin serum concentrations range between 150 and 250 kIU ml1 [24]. Additionally, different volumes of blood used to measure ACT with the SONOCLOT (360 µl) and the HEMOCHRON analyzer (2 ml) have to be considered. Both measurements may be affected to a varying degree by external conditions: Large volume ACT tests may be more influenced by temperature and the hypocoagulability seen in hemodiluted blood samples may depend on the sampling volume used [25].
In conclusion, the novel aprotinin-insensitive ACT test measured by the SONOCLOT analyzer (SaiACT) may be a valuable alternative to monitor heparin anticoagulation in presence of aprotinin compared to the kaolin-based ACT assessed by HEMOCHRON (HkACT). Values for SaiACT are lower than values for HkACT, and administration of aprotinin increases this difference significantly. Therefore, the use of SaiACT may result in an increased administration of heparin.
| Acknowledgments |
|---|
| Footnotes |
|---|
This work has been presented in part at the annual meeting of the European Society of Anaesthesiologists in Vienna, Austria, May 2005 and the American Society of Anesthesiologists in Atlanta, GA, October 2005. | References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. T. Ganter and C. K. Hofer Coagulation Monitoring: Current Techniques and Clinical Use of Viscoelastic Point-of-Care Coagulation Devices Anesth. Analg., May 1, 2008; 106(5): 1366 - 1375. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |