| Abstract |
|---|
|
|
|---|
Key Words: Cell saver Interleukins Cardiac surgery Blood quality
| 1. Introduction |
|---|
|
|
|---|
Mechanical cell salvage devices are therefore increasingly used as an alternative method for intra-operative blood salvage. In a cell saver the shed wound blood is collected, washed and concentrated. The red blood cell concentrate is then retransfused in the patient.
Recently, the quality of this washed and concentrated blood was addressed in several studies [1,5,6]. It was found that processing of shed blood by a cell saver led to normalization of some, but not all, inflammatory markers in the processed blood. However, to our knowledge, these studies considered only one processing run of the cell saver. As a consequence it is unknown if the quality of processed cell saver blood is affected when larger quantities of blood are processed. We hypothesized that the quality of the washed blood decreases when, due to large blood loss, multiple runs of the cell saver are necessary. This has not been studied before and might have implications for the way cell savers are used during operations.
Therefore, in the present study we assessed the quality of the processed blood by measuring interleukin-6 (IL-6), leukocytes and free hemoglobin. After 1250 ml of wound blood was collected in the blood collection reservoir of the cell saver we measured the blood quality. After processing the collected blood we again measured the blood quality. We compared this to the quality of the collected and washed blood after a subsequent 1250 ml of wound blood was collected and processed.
| 2. Materials and methods |
|---|
|
|
|---|
2.2 Clinical management
Anesthesia was induced and maintained with propofol infusion, followed by 0.1 mg/kg pancuronium to facilitate intubation. Sufentanil (1–3 µg/kg) was administered in incremental doses. Before cannulation bovine lung heparin (300 IU/kg) was administered and supplemented as necessary to obtain activated clotting times (ACT, Hemochron, New Jersey, USA) in excess of 400 s. After cardiopulmonary bypass (CPB), heparin was neutralized by protamine in a 1:1 ratio.
The CPB-circuit consisted of roller pumps (Stöckert, München, Germany) and a hollow fiber oxygenator (Cobe, Lakewood, CO, USA) primed with 500 ml hydroxyethyl starch 10% (Fresenius, Bad Homburg, Germany) and 1000 ml lactated Ringer's solution. Myocardial protection consisted of cold crystalloid solution (Plegisol, Abbott laboratories, IL, USA).
Cell salvage was achieved by using the CATS system (continuous auto transfusion system, Fresenius, Bad Homburg, Germany). The cell saver device was installed identically for every patient and every run was according to the manufacturer's instructions. Washing conditions for all patients and runs were set at the automated, quality wash program, incorporated in the machine by the manufacturer. The blood collection reservoir of the cell saver was primed with 100 ml of normal saline with 30.000 IU heparin/l added. All shed wound blood during the operation including blood from the operative field during CPB was collected in the cell saver reservoir. Cardiotomy suction was not used. Thus, all cardiotomy blood was also collected in the cell saver reservoir, thereby increasing the quantity of blood to be processed. The residual blood from the heart–lung machine after CPB was also processed by the cell saver, but this blood was not analyzed in this study.
When 1250 ml of shed blood was collected in the blood collection reservoir of the cell saver this was processed and returned to the patient. If another 1250 ml of blood could be collected a second processing run was performed. The amount of 1250 ml was arbitrarily chosen as this is a volume at which a clinically relevant transfusable amount of blood can be acquired. Furthermore this amount reflects the actual amounts lost on average in CABG operations. It is also an economical balance between cost of an allogenic blood transfusion and the disposable centrifuge unit.
Blood samples were drawn from the blood collection reservoir of the cell saver device before each run and from the processed and washed blood after each run.
The following parameters were measured: white blood cell (9 x 109/l) and platelet counts (9 x 109/l), hemoglobin (mmol/l), free hemoglobin (g/l), hematocrit (%), and the concentration of the pro-inflammatory cytokine interleukin-6 (IL-6, µg/l). Samples for cytokines were collected in ethylenadiamine tetraacetic acid (EDTA). The samples were centrifuged immediately and the resultant plasma was stored at –80 °C until analysis. IL-6 was determined with a commercially available enzyme-linked immunosorbent assay (R&D Systems, Minneapolis, MN, USA) according to the manufacturer's instructions.
2.3 Statistics
Statistical analysis was performed with Wilcoxon signed ranks test as not all data had a normal distribution. A p value of less than 0.05 was considered significant. All the results are expressed as the mean ± standard deviation.
| 3. Results |
|---|
|
|
|---|
|
|
Processing of blood from the cell saver blood collection reservoir resulted, in both runs, in a significant increase in hemoglobin, hematocrit and a significant decrease of platelets. The hemoglobin concentration, free hemoglobin concentration and platelet counts were not different between the two processed runs (Table 2).
| 4. Discussion |
|---|
|
|
|---|
Our IL-6 concentrations in the salvaged wound blood were relatively low compared to other studies that investigated the IL-6 concentrations in salvaged wound blood. An explanation may be that in this study only salvaged wound blood was used whereas in other studies the residual heart–lung machine blood was also processed [5,6]. As in earlier studies we found a high standard deviation of IL-6 values between and in patients [1]. We did not measure the IL-6 concentration in the patients, as in this pilot study we were only interested in the blood quality aspect of the cell saver. The significant increase in IL-6 in the unprocessed blood between the two runs is most likely due to the ongoing inflammatory response during the course of the operation and CPB [3,7]. Hemodilution by priming of the cell saver reservoir is a less likely explanation for the lower IL-6 concentrations in the first blood collection reservoir. The priming volume was 100 ml and would thus have a dilution effect of 8%.
Free hemoglobin levels were not reduced. Approximately 0.5% of erythrocytes are being damaged during and after the washing procedure, resulting in new hemoglobin release [8].
Our study also demonstrates a concentrating effect of leucocytes after each run. This has been demonstrated previously [1,9] and might be explained by the centrifugation process itself. White blood cells, which are larger but not heavier than red blood cells, are spun to the rim of the washing chamber, the so-called buffy coat. With ongoing centrifugation and washing the buffy coat becomes so concentrated with debris that leukocytes pass with the red blood cells into the process bag. The formation of such a layer of debris has been demonstrated before [9,10].
The consistent performance of red blood cell saving can be explained by the operating process of the cell saver itself. A light sensor reacts on the red blood cell concentration and starts the pump to transport the processed blood to the collection bag. Thus one would, given time, always expect similar hemoglobin and hematocrit in subsequent runs.
The device used in this study is the CATS cell saver. This is a continuous auto-transfusion system, which means that the system runs continuously until the blood collection reservoir is empty. This is in contrast to a bowl system, which operates by using batches of blood from the blood collection reservoir. In the current study we used this continuous system as a bowl collection system for a clear separation of the amount of blood processed.
In retrospect in a few patients it would have been possible to perform a third processing run. It would have been interesting to see the effect of a further processing run on the blood quality. Furthermore, measurement of leukocyte activation in the processed blood could have shown if leukocytes are activated by the by the blood processing process, giving further information on processed blood quality.
In conclusion, our results suggest that with repeated runs blood quality is maintained. With repeated runs of the cell saver device leukocytes are retained. This is probably due to a concentration effect.
| Footnotes |
|---|
Poster presentation at ASA Annual Meeting, Chicago, Illinois, USA, October 14 2006.

Oral presentation at the combined Dutch Cardio-Thoracic society (NVT), Dutch anaesthesiologist (NVA) and Dutch intensivist (NVIC) meeting, Nieuwegein, Netherlands, November 16 2006.


Funding was received from the Department of Anesthesiology, University Medical Centre Groningen.